METHOD DEVELOPMENT AND VALIDATION OF ATENOLOL USING TWO HPLC SYSTEMS
HTML Full TextMETHOD DEVELOPMENT AND VALIDATION OF ATENOLOL USING TWO HPLC SYSTEMS
Nidhal Sher Mohammed
School of Chemistry, Faculty of Science, University of Zakho, Duhok, Kurdistan, Iraq.
ABSTRACT: A reverse phase high performance liquid chromatography (RP-HPLC) method was developed and validated for the determination of Atenolol (ATN) in bulk and pharmaceutical formulation. The method was optimized selecting chromatographic conditions of 60: 40 acetonitrile: water, Inertsil_ column(ODS-3 250 mm × 4.6 mm 5 ?m), 20 ?L injection volume, flow rate of 1 mL/min at ambient temperature (30 °C), and 276 nm. Using two HPLC systems of first HPLC system coupled with PDA detector and the second HPLC system coupled with UV detector showed no big difference in the method results. The method was validated giving good precision (RSD% < 1), acceptable linearity (R2 > 0.998), and low LOD and LOQ (0.5 and 1.5 ?g/mL, respectively) on both systems. Successful application on pharmaceutical dosage tablet form gave high recovery of 97.7%. The proposed method is economic, simple, and rapid and hence can be employed for routine analysis in quality control laboratories.
Keywords: |
Atenolol; Hypertension; RP- HPLC; Optimization; Method validation
INTRODUCTION: Atenolol (ATN) is chemically described as 2-{4-[2-Hydroxy-3-(isopropylamino) propoxy] phenyl} acetamide (Fig. 1). It is white powder and water soluble. Atenolol is a selective β1 receptor antagonist, a drug belonging to the group of beta blockers class of drugs used primarily in cardiovascular diseases. Atenolol is used to treat angina (chest pain) and hypertension (high blood pressure). It is also used to treat or prevent heart attack or heart damage after a heart attack 1, 2. Atenolol is available in oral dosage forms viz. tablet, syrups and capsules. In the tablet dosage this drug is commonly available in three different strengths i.e. 25, 50 and 100 mg. Atenolol is also available as an antihypertensive treatment in the form of compound preparations with diuretics 3.
FIG. 1: CHEMICAL STRUCTURE OF ATENOLOL
In reviewing the literature, various analytical methods were found for the determination of atenolol whether alone or in its combination with other drugs in pharmaceutical preparations including spectrophotometry 4-6 kinetics 7, 8 Titrimetric 9 GC 10 and HPLC 11-15.
The aim of the present study was to develop of isocratic HPLC method for the analysis of ATN employing two HPLC systems of first HPLC system coupled with PDA detector and the second HPLC system coupled with UV detector. A further objective was to compare the sensitivity of a photo diode array detector (PDA) and UV/Vis detector for the determination of atenolol.
The analytical method was validated according to international conference on harmonization (ICH) guidance for validation of analytical procedure by examining the precision of the HPLC instruments used in this study, the linearity of the calibration curve and calculating the limit of detection (LOD) and the limit of quantification (LOQ) 16.
MATERIALS AND METHOD:
HPLC systems: Two HPLC systems were used during this study;
- First HPLC system (A) comprised a Flexar auto sampler -no Peltier, Flexar Binary pump, Inertsil® (ODS-3 250 mm x 4.6 mm 5 µm siloxan), and a Flexar /S275/S200EP Photo Diode Array detector (PDA) coupled with chromera® version 3.4.0.5712 software. Flexar column oven was connected with a system to control the column temperature.
- Second HPLC system (B) comprised a Flexar auto sampler -no Peltier, Flexar Quaternary pump, Inertsil® (ODS-3 250 mm x 4.6 mm 5 µm siloxan), and a Flexar UV/Vis detector coupled with chromera® version 3.4.0.5712 software.
Chemicals: Pharmaceutically pure sample of atenolol drug was obtained from Awamedica company (Erbil city - Kurdistan region of Iraq). Commercial tablet of ATN (100 mg) was brought from the local drug market. Acetonitrile and water were HPLC grade from merck KGaA, 64271 Darmstadt, Germany.
Preparation the mobile phase: The mobile phase of HPLC analysis was prepared from organic solvent of acetonitrile and water (v:v%). The mobile phase was degassed using an ultrasonic bath (100W, 80 KHz, Elmasonic P- Germany.).
Stock and working solutions: Stock solution of 1000 µg/mL of atenolol was prepared by dissolving 0.1 g of sample in 100 mL of acetonitrile. This stock standard solution was stored in a refrigerator and used to prepare the working solutions at different concentrations.
Sample solution preparation: Ten tablets were crushed to fine powder. An accurately weighed portion of the powder (equivalent to 20µg/mL) was taken and dissolved in 100mL solvent of acetonitrile.
Development and optimization of the HPLC method: To construct a basic background for developing an HPLC method with fast analysis for the determination of ATN, an isocratic programming was employed by analyzing solution of 10 µg/mL at 60%, (v/v) of ACN. The chromatographic conditions were set using a 20 μL injection volume, flow rate of 1 mL/min, column oven at ambient temperature (30 °C) and UV detection at a wavelength of 276 nm.
Method Validation:
Assessment the precision for HPLC systems: The precision of the two HPLC systems was evaluated by repeated injections (n=5) of concentration of 5 μg/mL ATN calculating the relative standard deviation (RSD %) of the peak area of ATN.
Linearity assessment: Set of ATN standards in the range of 2-10 μg/mL was prepared and injected into two HPLC systems and the detector response (peak area) plotted against the nominated concentration to generate the calibration curve. Using Excel® software, from the regression line, the coefficient of determination (?2) was obtained to statistically assess the linear relationship.
Limit of detection and limit of quantification: The limit of detection (LOD) and limit of quantification (LOQ) of ATN in solution were calculated statistically based on the data from the calibration curve 17. The calibration curve at low levels of ATN was used to statistically estimate the ATN peak area, which was significantly different from the blank peak area. The equation used to estimate this peak area was:
LOD peak area= yB + 3 sB
Where yB is the blank peak area and sB is the standard deviation of the blank peak area. In practice, the value of yB and sB can be estimated from the calibration graph at the lower range by using values from the regression analysis of variance. Therefore, the value of yB was replaced by the intercept value of the regression equation of the calibration graph of ATN at the lower range of 2-10 µg/mL. Similarly, the term sB was replaced by the residual standard deviation, which is the square root of the error mean square. Both terms were estimated using the regression function in Excel® software. The statistical data consist of the intercept, slope and standard deviation of the regression line (SD), after calculating the LOD and the LOQ peak area depending on the above information of the regression line, the LOD and the LOQ were determined according to the following equations:
LOD Peak area = intercept + 3 SD
LOD= (LOD Peak area – intercept) / slope
LOQ Peak area = intercept + 10 SD
LOQ= (LOQ Peak area – intercept) / slope
Application of the method to dosage form: The accuracy of proposed method was ascertained on the basis of average recovery % to determine ATN in its pharmaceutical preparation as tablets (equivalent to 20µg/mL) by triplicate injections under the same chromatographic conditions of the proposed method. The recovery was calculated according to the below:
% Recovery = (Recovered amount x 100)/Injected amount.
RESULTS AND DISCUSSION:
Development and optimization of method: Selecting best HPLC system through optimizing the chromatographic conditions is initial and important step prior to determination of atenolol in its commercial pharmaceutical dosages.
In this study, RP-HPLC method was developed and optimized using two HPLC systems connected with Inertsil®column(ODS-3 250 mm x 4.6 mm, 5 µm ciloxan). First HPLC system is coupled with PDA detector and the second HPLC system coupled with UV detector. Commonly, the UV detectors used for HPLC are single wavelength detectors for quantitative analysis. Whilst for more qualitative and quantitative information about the sample, photo diode-array detectors (PDA) may be employed to measure the absorbance at multi-wavelengths simultaneously.
The HPLC chromatograms in Fig. 2 show the analysis of 10 µg/mL solution of ATN using two HPLC systems under isocratic chromatographic conditions.
FIG. 2: CHROMATOGRAMS OF ATN PEAK USING TWO HPLC SYSTEMS: A (PDA DETECTOR) AND B (UV/VIS DETECTOR)
As shown in Fig. 2 both chromatograms indicate a good determination of ATN peak at 60% strength of the mobile phase, although chromatogram B showed broadening, peak of ATN was appeared at approximately 1.5 min retention time under the same chromatographic conditions for both systems. A possible explanation for broadening might be to large injection volume, molecular weight of drug and build-up of contamination on column inlet, residual silanol interaction and column deterioration 18.
The best chromatographic parameters for the proposed method showed no difference between both HPLC systems as summarized in Table 1.
Method validation: Validation of the analytical method prior to determining ATN in dosage sample was done by examining such parameters of precision, linearity, and LOD and LOQ using two HPLC systems as follows:
Assessment of the precision for HPLC systems: Validation of the method was performed through examining the precision. The two HPLC systems achieved good precision reporting RSD% values of the peak area of ATN less than 1% as shown in Table 2.
TABLE 1: CHROMATOGRAPHIC CONDITIONS OF PROPOSED ATN METHOD USING TWO HPLC SYSTEMS
Parameter | HPLC System A | HPLC System B |
Column | Inertsil® (ODS-3 250 mm x 4.6 mm 5 µm siloxan ) | Inertsil® (ODS-3 250 mm x 4.6 mm 5 µm siloxan ) |
Detector | PDA | UV |
Wavelength detection | 276 nm | 276 nm |
Mobile phase | 60 % CAN | 60% ACN |
Flow rate | 1 mL/min | 1 mL/min |
Chromatographic run | 5 minutes | 5 minutes |
ATN retention time | ~ 1.5 minutes | ~ 1.5 minutes |
Injection volume | 20 µL | 20 µL |
Column temperature | 30 °C | 30 °C |
TABLE 2: RSD% VALUES OF FIVE REPLICATE INJECTIONS OF 5 µg/mL OF ATN ON TWO SYSTEMS
Injection No. | Peak area of HPLC system A | Peak area of HPLC system B | |
1 | 9396 | 343201 | |
2 | 9406 | 343795 | |
3 | 9476 | 342956 | |
4 | 9398 | 339675 | |
5 | 9337 | 346665 | |
Mean | 9403 | 343258 | |
STDEV | 49.41 | 2491.4 | |
%RSD | 0.52 | 0.72 |
Linearity of the calibration curve: As shown in Fig. 3, the results illustrate a good linearity between the peak area and the concentrations of the standard solutions of ATN on both HPLC systems giving coefficient of determination (R2) > 0.998 over the selected range of concentrations.
FIG 3: CALIBRATION CURVES OF ATN STANDARDS ON TWO HPLC SYSTEMS: A (PDA DETECTOR) AND B (UV/VIS DETECTOR)
LOD and LOQ determination: LOD and LOQ were calculated by calibration curve at low level of the concentrations. The results in Table 3 show close values of LOD and LOQ of both systems.
TABLE 3: LOD AND LOQ VALUES OBTAINED FROM THE CALIBRATION CURVE OF LOWER CONCENTRATIONS (2-10 µg/mL) OF ATN OF BOTH SYSTEMS
Parameter | HPLC system A | HPLC system B |
Slope | 2181.45 | 70500.1 |
Intercept | 116.3 | 1471.2 |
SD | 313.62 | 10830.9 |
LOD peak area | 1057.17 | 33964.02 |
LOQ peak area | 3252.55 | 109780.62 |
LOD | 0.4 µg/mL | 0.5 µg/mL |
LOQ | 1.4 µg/mL | 1.5 µg/mL |
In the present method, two HPLC systems were operated HPLC coupled with photo diode array detector and HPLC coupled with UV/Vis detector. It can be concluded that the two HPLC systems offered acceptable precision, linearity and LOD and LOQ values despite the difference in the specification of the PDA and UV/Vis detectors. However, the PDA detector has the ability to monitor multi wavelength and present entire spectra of all peaks during chromatogram run and monitor the peak purity as well 19. The HPLC system coupled with PDA detector was selected for application the present method on pharmaceutical formulation sample.
Recovery: The validity and reliability of proposed method was assessed by recovery study. Determination of the ATN drug in its pharmaceutical dosage gave recovery equal to 97.7% with RSD% value of 0.49 indicating that the proposed method is suitable and reliable for the determination of ATN in pharmaceutical dosage forms.
TABLE 4: THE PERCENT RECOVERY (N= 3) OF 20 µg/mL ATN DOSAGE SAMPLE
Peak area of recovered amount
(conc. 20 µg/mL) |
Peak area of injected amount
(conc. 20 µg/mL) |
Recovery% |
Mean of Recovery |
RSD% |
|
Injection 1 | 42137 |
42879 |
98.26 |
97.70 |
0.49 |
Injection 2 | 41792 | 98.46 | |||
Injection 3 | 41756 | 97.38 |
CONCLUSION: Using two HPLC systems (HPLC- PDA and HPLC-UV), an isocratic RP- HPLC method for the determination of atenolol was developed and validated. Optimizing the method using an Inertsil (ODS3 C18, 4.6x250 mm, 5µ) column offered reliable chromatography conditions of 20 µL injection volume, 1 mL/min flow rate, 30 °C column oven temperature and retention time of approximately 1.5 min at a run time of 5 minutes on both systems and 60% ACN of mobile phase strength with 276 nm wavelength detection.
Validation the developed method using both HPLC systems offered acceptable precision (RSD %< 1), good linearity (R2 > 0.998) and low values of LOD (~ 0.5µg/mL) and LOQ (~ 1.5 µg/mL).
It can be concluded that the method shows no difference between both HPLC systems. HPLC system coupled with PDA detector was used to apply the present method on pharmaceutical formulation sample giving high recovery (97.7%).
ACKNOWLEDGMENT: The author would like to thank Mr. Ahmed J. Mohammed for his help during this study.
REFERENCES:
- Pai NR and Patil SS: Development and validation of liquid chromatographic method for atenolol and its related substance. Der Pharmacia Sinica 2013; 4 (2): 76–84.
- Naikini P, Akula A, Ajitha A and Rao VUM: RP-HPLC method development and validation for the simultaneous estimation of amlodipine and atenolol in bulk and tablet dosage forms. International Journal of Pharmacy and Pharmaceutical Sciences 2014; 6(1): 390-394.
- Chaudhari V, Hussian S, and Ubale M: A newer validated and stability indicating HPLC method for the estimation of Atenolol and Hydrochlorothiazide in bulk drug and dosage form. International Journal of Chemical Studies 2013; 1(4): 93–101.
- Lalitha G, Salomi P and Ravindra RK: Development of an analytical method and its validation for the analysis of Atenolol in tablet dosage form by UV-spectrophotometry. International Journal of Pharmacy and Pharmaceutical Sciences 2013; 5 (2): 197–199.
- Attia KM, Elabasawy NMA and Abolmagd E: Simultaneous spectrophotometric determination of amlodipine andatenolol in pharmaceutical preparations using chemometric tech-niques. Analytical chemistry an indian journal 2016; 16(5): 205–210.
- Lalitha KV, Kiranjyothi R, and Padma B: UV spectrophotometric method development and validation for the determination of Atenolol and Losartan Potassium by Q-analysis. International Bulletin of Drug Research 2013; 3(4): 54-62.
- Sikora A,Sroka W, Siódmiak T and Marszall M: Kinetic resolution of (R,S)-atenolol with the use of lipases in various organic solvent Current Organic Synthesis 2017; 14(1).
- Agarwal R and Fadnis A: Kinetic spectrophotometric determination of Atenolol in perchloric acid medium. International Journal of Pharmacy and Pharmaceutical Sciences 2012; 4(2): 350-352.
- Basavaiah K, Chandrashekar U and Nagegowda P: Titrimetric, spectrophotometric and kinetic methods for the assay of Atenolol using bromate-bromide and methyl orange. Journail of Serbian Chemical Socity 2006; 71 (5): 553–563.
- Yilmaz B and Arslan S: Determination of Atenolol in Human urine by Gas Chromatography-Mass Spectrometry method. Journal of chromatographic science 2011; 49(5): 365-9.
- Tengli AR and Gurupadayya BM: Method Development and Validation of Tablet Dosage form Containing Losartan, Atenolol and Hydrochlorthiazide Using Internal Standard by RP-HPLC. Journal of Chromatography & Separation Techniques 2013; 4(5): 1-5.
- Srivani A, Hazari S and Malichetti S: Development and validation of RP-HPLC method for simultaneous estimation of atenolol and amlodipine in bulk and tablet dosage form. International Journal of Pharmacy 2014; 4(1): 265-269.
- Belal F, Sharaf El-Din M, Aly F, Hefnawy El-Awady M: Stability-indicating HPLC Method for the Determination of Atenolol in Pharmaceutical Preparations. Journal Chromatography Separation Techniques 2013; 4 (1):1–7.
- Abdallah I, Ibrahim A, Ibrahim N, Rizk M and Tawakkol S: Simultaneous Determination of Atenolol and Nifedipine by Using Spectrophotometric Method with Multivariate Calibration and HPLC Method Implementing “Design of Experiment”. Pharmaceutica Analytica Acta 2015; 6(6): 1–9.
- Anderson JFF, Gerlin MGG, Sversut RA, Oliveira LCS, Singh AK, Amaral MS and Kassab NM: Development and Validation of an Isocratic HPLC Method for Simultaneous Determination of Quaternary Mixtures of Antihypertensive Drugs in Pharmaceutical Formulations. Acta Chromatographica 2016: 1-16.
- International Conference on Harmonization, Q1A9 (R2) Stability Testing of New Drug and Product, step 5version, 2003.
- Miller JN and Miller JC: Statistics and Chemometrics for Analytical Chemistry, Pearson Prentice Hall, Harlow, UK, 2005.
- Williams S: Ghost peaks in reversed-phase gradient HPLC. Journal of Chromatography A 2004; 1052 (1–2):1–11.
- Snyder LR, Kirkland JJ, and Glajch JL: Practical HPLC method development, 2nd ed. New York: John Wiley & Sons., 1997.
How to cite this article:
Mohammed NS: Method development and validation of atenolol using two HPLC systems. Int J Pharm Sci Res 2017; 8(6): 2706-11.doi: 10.13040/IJPSR.0975-8232.8(6).2706-11.
All © 2013 are reserved by International Journal of Pharmaceutical Sciences and Research. This Journal licensed under a Creative Commons Attribution-NonCommercial-ShareAlike 3.0 Unported License.