APPLICATION OF QbD APPROACH FOR ROBUST ANALYTICAL METHOD DEVELOPMENT OF SUNITINIB MALATE BY UHPLC
HTML Full TextAPPLICATION OF QbD APPROACH FOR ROBUST ANALYTICAL METHOD DEVELOPMENT OF SUNITINIB MALATE BY UHPLC
M. Sunilkumar, S. B. Haritha, P. Naresh, V. Chandran and Bharani Pandilla *
Department of Pharmaceutical Analysis, C. L. Baid Metha college of Pharmacy, Thoraipakkam, Chennai, Tamil Nadu, India.
ABSTRACT: Objective: Sunitinib malate is an oral, potent multi-targeted TKI that displays antitumor and anti-angiogenic actions. The core focus of the present research work is to establish an innovative QbD-based analytical method for the estimation of Sunitinib malate in solid dosage form by ultra-high performance liquid chromatography (UHPLC). Materials and Methods: The change in independent variables, wavelength, flow rate and Acetonitrile concentration with response to retention time, area and tailing factor were studied using Box-Behnken design. Chromatographic separation was accomplished by using XBridge shield RP18 (50mm x 3mm, 2.5μm) column, flow rate was adjusted to 0.7mL/min. at a temperature of 25ºC. The detection was carried out at 388nm with a run time of 5 min. Results and Discussion: The retention time of Sunitinib malate was found to be 2.391 min. The method is robust, accurate and precise, as depicted by the statistical data of analysis. The correlation coefficient was found to be 0.999. The solution was found to be stable for 24 hours at 250C. Hence the current research work was found to be economical and rapid and can be used for routine laboratory analysis for quality control.
Keywords: Sunitinib, QbD, Method development, UHPLC, ICH guidelines
INTRODUCTION: Sunitinib malate is chemically N-[2-(diethylamino) ethyl] – 5 - [(Z) - (5 – fluoro -1, 2dihydro- 2-oxo-3H-indol-3-ylidine) methyl]-2, 4-dimethyl-1H pyrrole-3-carboxamidean. It is an oral, potent, multi-targeted TKI that displays antitumor and antiangiogenic actions 1. It has been reported to inhibit VEGFR-1, VEGFR-2, VEGFR-3, KIT (stem-cell factor [SCF] receptor), PDGFR-a, and PDGFR-b in both cellular and biochemical evaluations.
All these receptors are involved in the angiogenesis of well-differentiated primitive neuro ectodermal tumors (PNETs). In-vitro, sunitinib malate was described to persuade apoptosis of endothelial cells of the human umbilical vein. Furthermore, it prohibits FMS-like tyrosine kinase 3, the neurotrophic factor receptor of the glial cell line, and the colony-stimulating factor-1 (CSF-1R) receptor.
Mechanism of sunitinib malate action in endothelial cells expressing the receptors of vascular endothelial growth factor (VEGFs) was reported to be the binding of VEGFs to VEGFR results in the dimerization of VEGFR and activation of VEGFR intracellular kinase domain, which requires the presence of ATP and penetration of sunitinib malate into the cell cytoplasm and its competition with ATP for the VEGFR ATP binding pocket. In the presence of sunitinib malate, activated VEGFR can no longer activate its intercellular kinase domain, further inhibiting downstream cell signalling 2, 3.
FIG. 1: STRUCTURE OF SUNITINIB MALATE
The main goal of the design of experiments is to ensure the quality of the finished products. QbD ensures combination of both the process and the product knowledge obtained during development which can deliver its planned performance consistently. Development measurements are based on the analytical target profile (ATP) and critical quality attribute (CQA) to assess the performance characteristics. The QbD approach overcomes the disadvantages of traditional development methods, such as time delay in studying one parameter variation one time, the number of runs required and further optimization. The study is performed based upon Box-Behnken design. The design consists of set of points lying at the midpoint of each edge and the replicated center point of the multi-dimensional cube. In order to evaluate the robustness of the strategy, the Box-Behnken Design (BBD), which requires fewer runs than a central composite design in cases of three or four variables, was used in the experimental design and statistical analysis of the data, which was acquired using Design-Expert software (version 12.0.3.1). The literature survey revealed various analytical techniques like UV spectrophotometry, HPTLC, HPLC and LC -MS/MS 4-21 were reported. No QbD-based UHPLC method for Sunitinib malate was developed and validated. An attempt was made to develop a simple, rapid, accurate, economical, and QbD-based technique for the dosage form of Sunitinib malate capsules.
MATERIALS AND METHODS: A simple, rapid, economical, QbD based method development was developed and validated as per ICH guidelines for Sunitinib malate capsule dosage form using UHPLC method.
Chromatographic Conditions: A UHPLC-developed chromatographic separation is achieved using the XBridge Shield RP18 column (50mm× 3 mm×2.5μm) column operating at a 0.7 mL/min. flow rate. The volume of injection was changed to 10μl at a temperature of 25 ° C with a run time of 5 min. Detection was carried out at 388 nm using the UV detector. Mobile phase separation was achieved by taking Ammonium acetate buffer: Acetonitrile (55:45v/v). The dimethyl sulfoxide was used to solubilize sunitinib.
Quantitative Estimation:
Preparation of Buffer Solution: 0.7mM Ammonium acetate buffer was prepared by dissolving 0.7791 g of Ammonium acetate in 1000 mL water. The solution was filtered through0.45μ nylon filter.
Preparation of Mobile Phase: Mobile phase is prepared by adding buffer and acetonitrile in the ratio of 55:45. The solution was mixed and degassed.
Preparation of Standard Stock Solution: 25mg of Sunitinib malate transferred in to 20Ml volumetric flask and added 10mL of Dimethyl sulfoxide (DMSO) dissolved and makeup with mobile phase (1250 µg/mL).
Standard Preparation Solution: Pipette out 2 ml of standard stock preparation into volumetric flask and makeup with mobile phase (125µg/mL) and mix well.
Preparation of Sample Solutions: 20 capsules were accurately weighed and average weight was calculated. Approximately139.18 mg (25 mg of Sunitinib malate) capsule powder was weighed and transferred into 20 mL volumetric flask.
About 10 mL of DMSO was added and sonicated. Use the mobile phase to make up the volume and mix well. 2 mL was pipetted from the above solution and the volume was made up by the mobile phase. The solution has been filtered with a 0.45 micron nylon syringe filter in the vial.
QbD Technique: The optimized chromatographic technique was developed by applying QbD based Box- Behnken design. Three independent factors were taken. The impact of change in these independent factors (wavelength, flow rate and ACN concentration) were compared with the response to dependent factors (retention time, area and tailing factor) were studied.
The Design-Expert software (version 12.0.3.1) was utilized for the experimental design and statistical analysis of data. The Box-Behnken Design (BBD) was chosen due to its efficiency in requiring fewer runs compared to a central composite design, especially in scenarios involving three or four variables, ensuring a thorough evaluation of the method's robustness.
The variables Wavelength (A), flow rate (B), and Acetonitrile concentration (C) were treated as independent factors, while Retention time (RT) (R1, min), Area (R2), and Tailing factor (R3) were designated as response variables. A 2-factor, 3-level Box-Behnken Design (BBD) was established, encompassing 15 experimental runs. The significance of the design was assessed through statistical parameters, including the ANOVA method and Good fit evaluation. Method parameter optimization was carried out using the response surface method, and the results of experimental runs were incorporated back into the software. The coding factor and 3D surface responses were considered and analyzed in the process.
Further the developed method was validated according to ICH guidelines for specificity, system suitability, linearity, precision, accuracy and robustness.
Method Validation:
Specificity: Specificity is the ability to check unequivocally of the analyte in the presence of components which be impurities, degradeant and matrix. 10µl of blank solution was injected into chromatograph and checked.
Accuracy: It is the state of being correct. The closeness of agreement between the true values which is accepted the value found. Based on the comparison to reference standard method.
Accuracy of Sunitinib malate was achieved by injecting three replicate injections at the concentration of 50%, 100% and 150% respectively. The percentage Recovery for each level should be between 98.00 to 102.00%
Precision: Precision of an analytical method is the degree of agreement among individual test results when the procedure is applied repeatedly to multiple sampling of a homogenous sample. Precision of analytical method is usually expressed as the standard deviation and relative standard deviation. The precision of the instrument was determined by assaying the samples six of times and relative standard deviation was calculated. Inject 10μl of the blank solution and the standard solution five times and calculate the %RSD for the area of six replicate injections.
The RSD for the area of six injections results should not be more than 2%.
Ruggedness: Ruggedness is also called as intermediate precision. The method performed on different day by using different make column and different analyst.
The standard solutions were injected for five times and measured the area for all 5 injections in UHPLC.
Robustness: The robustness of the method was analysed by changing experimental, chromatographic condition. Altering in flow rate (0.7±1mL/min), changes in column oven temperature (40±5°C), Changes mobile phase buffer pH (3.5±0.2), changes in mobile phase composition and changes in wavelength allowable limits from actual chromatographic conditions.
Linearity: Linearity is the ability of the method to obtain test results that are directly proportional to the analyte concentration within a given range. The linearity of the method was performed by preparing the concentration range of 62.5-187.5µg/mL for Sunitinib malate, from standard stock solution. Calibration curves were constructed by plotting concentration versus area of sunitinib malate. Correlation coefficient should be not less than 0.999.
RESULTS AND DISCUSSION: The horizon of the present work is to develop a new QbD based optimization of the chromatographic condition using RP-UHPLC method and to validate the method using ICH guidelines. The mobile phase comprising of mixture of ammonium acetate buffer: Acetonitrile (55:45, v/v) was chosen as an ideal mobile phase, since it gave a good resolution and peak shapes with perfect optimization. The detection was carried out at 388 nm. The flow rate was optimized at 0.7ml/min. X Bridge Shield RP-18 (50mm x 3mm, 2.5μm) column was used for the analysis. A mixture of dimethyl sulfoxide and mobile phase was used as diluent. The retention time of Sunitinib malate was found to be 2.391 min. respectively with a run time of 5 min. theoretical plate for Sunitinib malate were 7971.38 respectively. Chromatogram is given in Fig. 2.
FIG. 2: CHROMATOGRAM OF SUNITINIB MALATE
QbD Approach: Three independent factors (wavelength, flow rate and ACN concentration) were taken based on the observation during the trials and from various optimization studies. The experiment was carried out based on the qualitative responses of the co- factors (retention time, area and tailing factor) and observed responses were noted and summarized in Table 1.
TABLE 1: EXPERIMENTAL DESIGN FOR INDEPENDENT VARIABLES AND CO-VARIANTS, EXPERIMENTAL AND PREDICTED VALUES
S. no. | Wavelength (A) | Flow rate (B) | ACN(C) | Rt(R1) | Area(R2) | Tailing factor (R3) | Desirability |
1 | 388 | 0.7 | 45 | 2.391 | 6093 | 1.486 | 1.000 |
2 | 388 | 1 | 30 | 2.109 | 6123 | 1.297 | 1.000 |
3 | 420 | 0.4 | 45 | 3.297 | 5893 | 1.623 | 1.000 |
4 | 388 | 0.8 | 45 | 2.393 | 6093 | 1.486 | 1.000 |
5 | 420 | 0.7 | 30 | 2.602 | 5723 | 1.456 | 1.000 |
6 | 389 | 0.7 | 45 | 2.396 | 6093 | 1.486 | 1.000 |
7 | 388 | 0.4 | 30 | 3.412 | 6234 | 1.617 | 1.000 |
8 | 420 | 0.7 | 60 | 2.005 | 5784 | 1.477 | 1.000 |
9 | 356 | 0.7 | 60 | 2.007 | 5234 | 1.469 | 1.000 |
10 | 388 | 0.4 | 60 | 3.122 | 6342 | 1.614 | 1.000 |
11 | 388 | 1 | 60 | 1.253 | 5934 | 1.287 | 1.000 |
12 | 356 | 0.4 | 45 | 3.402 | 4999 | 1.609 | 1.000 |
13 | 356 | 1 | 45 | 1.954 | 5096 | 1.291 | 1.000 |
14 | 420 | 1 | 45 | 1.957 | 5892 | 1.287 | 1.000 |
15 | 356 | 0.7 | 30 | 2.611 | 5523 | 1.478 | 1.000 |
Among various models’ quadratic model was selected. The procedure was evaluated for the effect of individual factors and their covariant responses Table 2 (ANOVA Analysis).
TABLE 2: STATISTICAL PARAMETERS BY ANOVA ANALYSIS FOR THE RESPONSE
Parameter | SS (Sum of Squares) | Df | MS (Mean Square) | F -value | P-value | Model F value | Model P value |
Retention time (R1) | 0.2461 | 3 | 0.0820 | 7.57 | 0.0263 | 55.77 | 0.0002 |
Area (R2) | 1.429 | 3 | 4.762 | 14.80 | 0.0064 | 7.80 | 0.0179 |
Tailing factor (R3) | 0.2065 | 3 | 0.0688 | 3.76 | 0.0939 | 94.03 | 0.0001 |
The high F-values implies that the models are significant and there is only a 0.01% chance that F-values could occur due to noise. p-values (0.0002, 0.0179, 0.0001) less than 0.0500 indicate model terms are significant. If the value is greater than 0.0500 that indicates the model terms are not significant. The suggested quadratic equation for all the individual responses in terms of coded factors are shown as follows:
Retention time = +2.47 - 0.0144A - 0.7386B – 0.2934C + 0.0270AB + 0.0017AC-0.1415BC + 0.0083A2 + 0.1803B2 -0.1702C2
Area = +6096.14 + 304.95A– 52.20B- 38.63C-24.50AB+87.50AC– 74.25BC – 609.31A2 – 16.61B2–+78.94C2
Tailing factor = +1.47 - 0.0004A - 0.1596B- 0.0001C +0.0005AB -0.0032AC-0.0005BC - 0.0086A2 - 0.0009B2 – 0.0001C2+0.0000ABC – 0.0000 A2 B – 0.0000 A2 C – 0.0000 AB2 + 0.0000AC2 + 0.0000B2C + 0.0000BC2 +0.0000A3 +0.0000B3+0.0000C3
The variables R1, R2, and R3 and quadratic terms with positive sign indicates synergistic effect and the negative sign indicates antagonistic effect in polynomial equation. The independent variables provide varying factor from low to high. When two variables interact their responses were recorded as AB, AC, BC. It is observed that when there is interaction between A and B independent variables R1 and R3 have synergistic effect where as R2 shows antagonistic effect. When there is interaction between A and C independent variables R1 and R2 have synergistic effect where as R3 shows antagonistic effect. All the dependent variables have antagonistic effect when there is interaction between B and C independent variables. These effects were further embellished by using 3D surface response and their contour plots given by the software explains the interactive relationship of two factors on the response by keeping the third factor constant. The 3D surface plots of the interaction effect were shown in Fig. 3, 4.
FIG. 3: 3D- RESPONSE OF RETENTION TIME (RT) (R1), AREA (R3), TAILING FACTOR (R3)
The optimized method was determined by setting the maximum and minimum values for the independent variables and entering their response variables as summarized in the Table 3. The numerical optimization suggests the desired method solution.
The desired method parameters were found to be wavelength at 388 nm and flow rate at 0.7 mL/min. with ACN concentration of 45% at desirability of 1.000. The graphical optimization of desirability was shown in the Fig. 4 (desirability)
TABLE 3: INDEPENDENT VARIABLES AND THEIR CORRESPONDING LEVELS
Variable | Symbol | -α | -1 | 0 | +1 | +α |
Wavelength | A | 356 | 376 | 388 | 400 | 420 |
Flow rate | B | 0.4 | 0.6 | 0.7 | 0.8 | 1 |
ACN | C | 30 | 40 | 45 | 50 | 60 |
FIG. 4: DESIRABILITY CHART
The adjusted and predicted R-squared concludes that the applied statistical model effectively predicts the response. The 3D surface shows that R2, P-value, F-value are within the limits which shows that the method is significant. Hence, the obtained optimized method by Box-Behnken design and optimized chromatographic conditions which was further validated as per ICH guidelines.
Validation Parameters:
System Suitability: System suitability parameters like tailing factor, retention time, theoretical plates per unit, resolution factor are determined by injecting 5 replicates of Sunitinib malate. %RSD was found to be less than 2% indicating good performance of the system.
Specificity: Specifity studies shows that there is no interfering peak at retention time of the analyte peak. Calibration curve shows that the method is linear and greater correlation exists between concentration and peak area.
The %RSD is 0.74, Rt was 2.391 min, Peak area was 6023.448. The developed method was vadilated and the results obtained were found to be within the limits.
Linearity: Chromatograms were acquired when standard solutions with linearity in the concentration range of 62.5-187.5µg/mL were injected. Linearity of concentration and peak area was plotted. R2 was found to be 0.9991. From the calibration graph regression parameters were calculated and linearity graph was shown in Fig. 5.
FIG. 5: LINEARITY OF SUNITINIB
Accuracy: The accuracy was analysed at 50-150%. The chromatograms were recorded and %RSD was found to be within the limits. The accuracy data was tabulated in Table 4. The accuracy of Sunitinib was found to be 99.16 to 99.65% respectively which is within the acceptance limit as per ICH guidelines
TABLE 4: ACCURACY RESULTS
Recovery Level | Sample | Mean Peak Area | % Recovery | Mean Recovery (%) | %RSD |
50% | 1 | 3028.192 | 99.78 | 99.65 | 0.13 |
2 | 3050.525 | 99.65 | |||
3 | 3078.419 | 99.52 | |||
100% | 1 | 6206.661 | 99.65 | 99.47
|
0.22
|
2 | 6185.477 | 99.23 | |||
3 | 6213.884 | 99.54 | |||
150% | 1 | 9358.491 | 99.65 | 99.16 | 0.22
|
2 | 9364.791 | 99.54 | |||
3 | 9419.931 | 99.23 |
NOTE: Three samples at three different concentration were analysed; SD: Standard Deviation, n=3
Precision: The precision of the analytical method was determined by assaying sufficient number of samples and relative standard deviation was calculated.
The results obtained were tabulated in Table 5. The %RSD of Sunitinib was found to be 0.33.
TABLE 5: PRECISION RESULTS
S. no. | Peak Retention Time
(min.) |
Peak area counts |
1. | 2.391 | 6075.336 |
2. | 2.391 | 6044.421 |
3. | 2.391 | 6064.390 |
4. | 2.392 | 6080.911 |
5. | 2.392 | 6040.565 |
6. | 2.391 | 6037.234 |
MEAN | 2.391 | 6057.143 |
STD DEV | 0.0005 | 18.87 |
%RSD | 0.02 | 0.33 |
%RSD: Relative Standard deviation, [n=6]
Ruggedness: The standard solutions were injected for five times and measured the area for all five injections in UHPLC. Results obtained were tabulated in Table 6.
TABLE 6: RUGGEDNESS
S. no. | Peak Retention Time
(min.) |
Peak area counts |
Injection1 | 2.391 | 6016.362 |
Injection2 | 2.391 | 6022.849 |
Injection3 | 2.392 | 6039.673 |
Injection4 | 2.392 | 6068.921 |
Injection5 | 2.391 | 6034.285 |
MEAN | 2.3914 | 6036.418 |
STD DEV | 0.0005 | 20.3589 |
%RSD | 0.0229 | 0.33 |
SD: Standard deviation, n=5
Robustness: Robustness was performed by changing experimental, chromatographic conditions. The results were tabulated in Table 7.
TABLE 7: ROBUSTNESS RESULTS FOR SUNITINIB MALATE
Drug name | Parameter | Chromatographic condition | |
Sunitinib Malate |
Flow rate change | RT (min.) | Area Counts |
0.6mL/min | 2.485 | 6156.326 | |
0.7mL/min | 2.391 | 6085.354 | |
0.8mL/min | 2.289 | 6056.239 | |
Wavelength change±2% | RT (min.) | Area Counts | |
386nm | 2.392 | 6181.546 | |
388nm | 2.391 | 6074.523 | |
390nm | 2.391 | 6065.236 |
SD: Standard deviation, n=6
Assay: Test result is showing that the test method is precise. The percentage assay of Sunitinib malate is found to be 100.14%. Results are within the limits. The results are summarized in the Table 8.
TABLE 8: ASSAY
Label claim | Average weight | Area Counts | Percentage assay |
25mg | 139.18 mg | 6051.247 | 100.14 |
All the parameters were validated and the validation reports were summarised in Table 9.
TABLE 9: VALIDATION PARAMETERS
S. no. | Parameters | Acceptance Criteria | Results |
1 | System suitability | Theoretical plate | 7971.38 |
Tailing Factor | 1.2 | ||
2 | Specificity | No interference was observed between placebo and blank with principal peak | Blank-Nil, Placebo-Nil
Standard-2.39, Sample-2.392 |
3 | Accuracy | 98-102% | 99.53% |
4 | Precision | RSDNMT 2% | 0.33% |
5 | Linearity | Correlation coefficient NLT 0.999 | 0.999 |
6 | Solution stability | RSD NMT 2% | Stable for 24 hours |
7 | Assay | 90-110% | 100.14% |
8 | Robustness | RSDNMT 2% | Complies |
CONCLUSION: The optimized method was suitable, linear, precise accurate and robust for the estimation of sunitinib malate in capsule dosage form. The developed method is rapid, extremely small flow rate with relatively short run time.
All these factors enable rapid quantification and estimation of the sunitinib malate in capsule pharmaceutical formulation without any excipient interference. It can therefore be concluded that the developed method is more economical and practical. It can be applied for the analysis of the sunitinib malate in research and quality control laboratories.
ACKNOWLWDGEMENT: The authors are thankful to Synthiya Research Lab Pvt Ltd., Puducherry, India, for providing standards and all facilities throughout the research work. The authors sincerely show gratitude to managemet of C.L. Baid Metha College of Pharmacy, Chennai, for the constant encouragement during the research work carried out.
CONFLICT OF INTEREST: The authors have no conflicts of interest in this work.
REFERENCES:
- Wilkes GM: Targeted therapy: Attacking cancer with molecular and immunological targeted agents. Asia Pac J Oncol Nurs 2018; 5(2): 137-55
- Jannin A, Penel N, Ladsous M, Vantyghem MC and Do Cao C: Tyrosine kinase inhibitors and immune checkpoint inhibitors-induced thyroid disorders. Crit Rev Oncol Hematol 2019; 141: 23-35.
- Kerbel R and Folkman J: Clinical translation of angiogenesis inhibitors. Nat Rev Cancer 2002; 2(10): 727-39.
- Kavitha J, Saidevaraj AB and Lakshmi KS: UV spectrophotometric estimation of sunitinib malate in pharmaceutical dosage form. Int J Pharm Pharm Sci 2016; 8(2): 99-03.
- Pourbasheer E, Malekpour L, Azari Z, Masand VH and Ganjali MR: Magnetic solid phase extraction of Sunitinib malate in urine samples assisted with mixed hemimicelle and spectrophotometric detection. Sci Rep 2023; 13(1): 3361.
- Monireh H, Goudarzi M, Ghaffari S, Attar H and Mazlagha M: Development and validation of a HPTLC method for analysis of sunitinib malate. Braz J Pharm Sci 2016; 52(4): 595-01.
- Jiang W, Zhao T, Zhen X, Jin C, Li H and Ha J: Rapid determination of 9 tyrosine kinase inhibitors for the treatment of hepatocellular carcinoma in human plasma by QuEChERS-UPLC-MS/MS. Front Pharmacol 2022; 13: 920436.
- Kasturi R, Naidu CG, Kumar CN and Shankar EB: Quality by Design based pareto charts responses Evaluation approach for a Validated Stability-indicating RP-HPLC assay Method for sunitinib and its process related impurities in Oral Dosage Forms. Research Journal of Pharmacy and Technology 2023; 16(4): 1999-05.
- Guo ZX, Wu YE, Shi HY, van den Anker J, Liang P, Zheng Y, Zhao XW, Feng R and Zhao W: A liquid chromatography-tandem mass spectrometry method for simultaneous quantification of thirty-nine tyrosine kinase inhibitors in human plasma. J Pharm Biomed Anal 2023; 224: 115159.
- Nouriyeh MH, Mogaddam MR, Nemati M, Farajzadeh MA, Abbasalizadeh A and Hojghan AS: Development of dispersive solid phase extraction based on magnetic metal organic framework for the extraction of sunitinib in biological samples and its determination by high performance liquid chromatography-tandem mass spectrometry. J Chromatogr B Biomed Appl 2024; 1239: 124109.
- Hirasawa T, Kikuchi M, Takasaki S, Kumondai M, Sato Y, Sato T, Imoto E, Hayakawa Y, Maekawa M and Mano N: High throughput LC/ESI-MS/MS method for simultaneous analysis of 20 oral molecular-targeted anticancer drugs and the active metabolite of sunitinib in human plasma. Heliyon 2023; 9(6): 16926.
- Tazeen A: UV-Vısıble spectroscopıc method development and validation of sunitinib malate ın bulk and formulatıon. WJPPS 2017; 6(5): 730-65.
- Blanchet B, Saboureau C, Benichou A, Billemont B, Taieb F, Ropert S, Dauphin A, Goldwasser F and Tod M: Development, and validation of an HPLC-UV-visible method for sunitinib quantification in human plasma. Clin Chim Acta 2009; 404(2): 134-39.
- Padervand M, Ghaffari S, Attar H and Nejad M: Reverse phase HPLC determination of sunitinib malate using UV detector, its isomerisation study, method development and validation. J Anal Chem 2017; 72(5): 567-74.
- Etienne-Grimaldi M, Renée N, Izzedine H and Milano G: A routine feasible HPLC analysis for the anti-angiogenic tyrosine kinase inhibitor, sunitinib, and its main metabolite, SU12662, in plasma. J Chromatogr B Analyt Technol Biomed Life Sci 2009; 877(29): 3757-61.
- Gurjar M, Mehta P, Sharma J, Patil S, Kulkarni P, Patil A, Nookala M, Joshi A and Gota V: An HPLC method for simultaneous quantification of sunitinib and its active metabolite, SU12662, using hydrophilic interaction chromatography principle. Bioanalysis 2020; 12(2): 75-85.
- Mınkın P, Zhao M, Chen Z, Ouwerkerk J, Gelderblom H and Baker S: Quantification of sunitinib in human plasma by high-performance liquid chromatography-tandem mass spectrometry. J Chromatogr B Analyt Technol Biomed Life Sci 2008; 874(1- 2): 84-88.
- Musijowski J, Piórkowska E and Rudzki P: Determination of sunitinib in human plasma using liquid chromatography coupled with mass spectrometry. J Sep Sci 2014; 37(19): 2652-58.
- Zhou Q and Gallo J: Quantification of sunitinib in mouse plasma, brain tumor and normal brain using liquid chromatography–electrospray ionization-tandem mass spectrometry and pharmacokinetic application. J Pharm Biomed Anal 2010; 51(4): 958-64.
- De Bruijn P, Sleijfer S, Lam M, Mathijssen R, Wiemer E and Loos W: Bioanalytical method for the quantification of sunitinib and its n-desethyl metabolite SU12662 in human plasma by ultra performance liquid chromatography/tandem triple-quadrupole mass spectrometry. JPBA 2010; 51(4): 934-41.
- Rodamer M, Elsinghorst P, Kinzig M, Gütschow M and Sorgel F: Development and validation of a liquid chromatography/tandem mass spectrometry procedure for the quantification of sunitinib (SU11248) and its active metabolite, N-desethyl sunitinib (SU12662), in human plasma: Application to an explorative study. J Chromatogr B Analyt Technol Biomed Life Sci 2011; 879(11-12): 695-06.
- Gotze L, Hegele A, Metzelder S, Renz H and Nockher W: Development and clinical application of a LC-MS/MS method for simultaneous determination of various tyrosine kinase inhibitors in human plasma. Clin Chim Acta 2012; 413(1-2): 143-49.
How to cite this article:
Sunilkumar M, Haritha SB, Naresh P, Chandran V and Pandilla B: Application of QBd approach for robust analytical method development of sunitinib malate by UHPLC. Int J Pharm Sci & Res 2025; 16(5): 1351-59. doi: 10.13040/IJPSR.0975-8232.16(5).1351-59.
All © 2025 are reserved by International Journal of Pharmaceutical Sciences and Research. This Journal licensed under a Creative Commons Attribution-NonCommercial-ShareAlike 3.0 Unported License.
Article Information
21
1351-1359
851 KB
24
English
IJPSR
M. Sunilkumar, S. B. Haritha, P. Naresh, V. Chandran and Bharani Pandilla *
Department of Pharmaceutical Analysis, C. L. Baid Metha college of Pharmacy, Thoraipakkam, Chennai, Tamil Nadu, India.
bharaniashok@gmail.com
19 December 2024
31 December 2024
05 January 2025
10.13040/IJPSR.0975-8232.16(5).1351-59
01 May 2025