COMPARATIVE PERFORMANCE OF LIQUID CHROMATOGRAPHY AND SPECTROPHOTOMETRY FOR QUANTITATIVE ANALYSIS OF PARACETAMOL IN TABLETS FORMULATIONS
HTML Full TextCOMPARATIVE PERFORMANCE OF LIQUID CHROMATOGRAPHY AND SPECTROPHOTOMETRY FOR QUANTITATIVE ANALYSIS OF PARACETAMOL IN TABLETS FORMULATIONS
Mahmoud A. Mansour *, Badran Alyahya, Abdullatif Al-Dulaimi, Ibrahim Farh, Waleed Alhussaini, Amani Kurdi, Sultan Alqahtani, Wesam S. Abdel-Razaq, Imadul Islam and Abdullkaree M. Albekairy
King Abdullah International Medical Research King Abdullah International Medical Research Center, Ministry of National Guard Health Affairs, Riyadh, Saudi Arabia.
ABSTRACT: Several techniques have been developed and validated for analyzing paracetamol in pharmaceutical formulations, including Ultra High-Performance Liquid Chromatography (UHPLC). This study employed a validated method for analyzing three commercially available paracetamol products. The chromatographic separation utilized a mobile phase consisting of a 70:30 v/v ratio of methanol to water. The separation process was carried out using a C18 Rapid Resolution column (4.6 x 100 mm, 3.5 µm, Agilent HPLC Column). The maximum wavelength (λmax) for paracetamol estimation was determined at 243 nm. The separation process was achieved using an injection volume of 20 μl and a run time of 3.0 minutes. The method exhibited linearity within the concentration range of 10-150 μg/ml for Paracetamol. The validation of the method followed the International Conference on Harmonization (ICH) guidelines, covering assessments of specificity, selectivity, linearity, accuracy, and precision, as well as the lower limit of quantification (LLOQ), and lower limit of detection (LLOD). The LLOQ and LLOD for paracetamol were established at 250 ng/ml and 100 ng/ml, respectively. The assay percentage for all pharmaceutical products containing paracetamol met the acceptable criteria range of 90-110%. These methodologies have proven effective for the accurate quantitative analysis of paracetamol in pharmaceutical formulations.
Keywords: Paracetamol, Liquid Chromatography, Spectrophotometry, Quantitative Analysis, Validation, % Assay, UHPLC
INTRODUCTION: Paracetamol, also known as acetaminophen or 4-acetamidophenol, is a widely used analgesic and antipyretic.
In 1878, an American chemist named Harmon Northrop Morse synthesized paracetamol by reducing p-nitrophenol with tin in glacial acetic acid.
However, the therapeutic potential of paracetamol wasn’t recognized and widely used until much later in the 20th century. Unlike NSAIDs, paracetamol does not have significant anti-inflammatory properties. Paracetamol, available over-the-counter, is the primary alternative for individuals with sensitivity to aspirin (acetylsalicylic acid) 1. It is often combined with decongestants, antihistamines, cough suppressants, or expectorants for treating cold or flu 2.
The literature describes several methods and techniques for determining paracetamol in various pharmaceutical preparations. These methods include titrimetric 3, 4, high-performance liquid chromatography (HPLC) and reversed-phase high-performance liquid chromatography (RP-HPLC) 5, 6, 7, high-performance thin layer chromatography (HPTLC) 8, 9, voltammetry 10, 11, 12, 13, electrochemical analysis 14, and spectrophotometry 15, 16, all of which have been reported and validated. Nonetheless, the spectrophotometric techniques employed in determining paracetamol can often lack sensitivity, and selectivity, and may pose toxicity risks 17, 18, 19, 20.
Reliable, robust, and safe analytical techniques are pivotal for the analysis and detection of paracetamol, both for quality control in pharmaceutical formulations and subsequently for medical monitoring in biological fluids. Therefore, the ability to accurately quantify paracetamol in different settings is essential. The present research aims to evaluate the sensitivity and specificity of the Ultra High Performance Liquid Chromatography (UHPLC) technique for the quantitation of paracetamol in both pure and pharmaceutical dosage forms to establish the most sensitive methods.
MATERIALSAND METHODS:
Chemicals: Paracetamol standard powder (Paracetamol, Molekula®, UK) was used to prepare the standard stock solution and serial dilutions. Methanol CHROMASOLV™, for HPLC, with a purity of ≥ 99.9%, was obtained from Honeywell (USA).Ultra-pure water system (Milli-QÒ Integral 5 (ZRXQ005T0) Millipore®) (USA), was used as HPLC grade water.
UHPLC Analysis: An Agilent 1290 Infinity model UHPLC equipped with a Variable Wavelength Detector (VWD) was used. Data acquisition, recording, and chromatographic integration were performed using the Open labEZ Chromsoftware (version A.01.05), for method development and validation of paracetamol. Isocratic elution was performed using an aqueous mobile phase consisting of a mixture of methanol and water (70:30, v/v). The mobile phase was filtered through a 0.45 µm pore size filter, followed by sonication and degassing in an ultrasonic bath (CPX5800H-E, Bransonic®). As Aljeboree et al.21 outlined, a standard paracetamol solution was prepared by dissolving 1000 mg of the powder in Milli-Q water and gently shaken to ensure complete dissolution, yielding stock solution of 1.00 mg/ml. The stock standard solution of paracetamol (1.00 mg/ml) was diluted with water to prepare calibration standard solutions ranging from 10 to 150 μg/ml, up to a total volume of 100 ml. The standard solution was then transferred into a 2 ml glass vial (Thermo Fisher Scientific Inc.) and injected into the UHPLC system 21.
According to the method described by Saeed et al. 20 for the preparation of pharmaceutical samples for assay tests of tablet formulations, the average weight of ten tablets from each of the following products was determined: Panadol®, Panadol Cold & Flu®, and Panadol Extra. Each set of tablets was then finely crushed into a powder, and an amount equivalent to the average weight of the tablet was transferred to a 500 ml amber volumetric flask, diluted with water, and gently shaken to ensure the homogeneity of the solution. To standardize the paracetamol concentration across all samples, the assay test solutions were prepared to a concentration of 1.00 mg/ml.
These solutions were filtered immediately using a syringe (BD 10 ml, Luer-Lok™ Tip) fitted with a filter (Nalgene™ syringe filter 25 mm (0.45 µm) PES, Thermo Fisher) into a 10 ml glass tube. Next, 10 ml of the sample was pipetted using the (Micropipette 1000 μl, Eppendorf Research® plus) into a 100 ml amber volumetric flask, diluted to the mark with water, and gently shaken resulting in a final concentration of 100.0 μg/ml. Three replicate solutions were prepared for analysis, labeled and 20 µl of each sample was injected into the UHPLC system 20.
Reverse-Phase Ultra-High-Performance Liquid Chromatography (RP-UHPLC) Method Development, Optimization, and Validation: The RP-HPLC was carried out using ZORBAX Eclipse Plus C18 Rapid Resolution Agilent HPLC Column (4.6 x 100 mm, 3.5 µm) at room temperature. The detection wavelength was set at λmax of 243 nm using anisocratic mobile phase consisting of a mixture of methanol and water in a 70:30 ratio(v/v) delivered at a flow rate of 1.00 ml/min, following the method described by Youssef et al 22.
The chromatographic separation was performed with an injection volume of 20 μl at 25 °C with a 3.0-minuterun time. By choosing a shorter column packed with smaller particles, analysis time can be significantly reduced while maintaining column efficiency and resolution. Previously, the smallest commonly used porous particles commonly used were either 3.5 or 3 µm in diameter. In the present study, a C18 column was used as it can easily resolve paracetamol in under 3 minutes with high efficiency.
Method Validation: The proposed UHPLC method has been validated in compliance with the International Conference on Harmonization (ICH) guidelines, encompassing specificity, selectivity, linearity, accuracy, precision, limit of quantification (LOQ), limit of detection (LOD), and robustness 23.
Specificity/ Selectivity: Selectivity refers to the capability of an analytical method to differentiate and measure the analyte in the presence of potential interfering substances in the sample. The selectivity assessment showed no significant response attributable to interfering ingredients at the retention time of the analyte, which was 1.657 min for paracetamol. This method demonstrated high specificity for detecting the paracetamol analyte, despite other components in the formulation.
Linearity and Range: According to the ICH guidelines, the linearity of the analytical method is defined by its ability to obtain results that are directly proportional to the analyte concentration within a given range. Various aliquots of the standard solution from the stock solution were prepared and analyzed to determine the linearity of the used method. The drug showed linearity within the range of 10-150μg/ml with a correlation coefficient (r2) ≥ 0.997712.
Accuracy and Precision: The intraday accuracy and precision study involved preparing a set of calibration standard curve samples and 12 control samples (QCs) at each concentration [low (25.00 µg/ml), medium (75.00 µg/ml), and high (125.00 µg/ml)] and analyzed within a single day. Additionally, another set of calibration standard curve samples and six QC samples of each concentration (low (25.00 µg/ml), medium (75.00 µg/ml), and high (150.00 µg/ml)) were prepared and analyzed over three days. The mean, % accuracy, SD, and % CV were calculated for both the intraday and interday validation.
Lower Limit of Quantification (LLOQ) and Lower Limit of Detection (LLOD): The LLOD refers to the smallest amount of an analyte detectable in the sample, while the LLOQ is the minimum amount of analyte that can be quantitatively measured with acceptable precision and accuracy. Both LLOQ and LLOD were determined using data from the lowest concentration standard during accuracy and precision runs, as previously described, and by measuring signal/noise ratios (S/N). The mean of these S/N ratios is then utilized to calculate the theoretical LLOD (S/N ≥ 3), and the theoretical LLOQ (S/N ≥ 10).
Application of the Proposed Method for Pharmaceutical Formulation: The assay test for the various strengths of paracetamol tablets was designed to ascertain the precise quantity of the active ingredient. The analysis of each batch involved the preparation of six calibration standards, six QCs, and six assay samples, all of which were intended for injection into the RP-UHPLC method.
RESULTS: This paper outlines the development and validation of a simple, sensitive, precise, and accurate UHPLC method for quantifying paracetamol in pharmaceutical formulations. Several mobile phases and columns were initially tested to achieve an optimal chromatogram. The selected column and the mobile phase were evaluated for selectivity, sensitivity, and acceptable chromatographic parameters of the resulting peaks in terms of sharpness, symmetry, and tailing factor. Fig. 1 displays a UHPLC chromatogram of varying concentrations of paracetamol from the lowest concentrations (10 µg/ml) to the highest concentrations (150 µg/ml).
The UHPLC chromatography results showed excellent linearity with a determination coefficient (r2) ≥ 0.997712 and correlation coefficient (r ≥ 0.998844) within the range of 10-150.0µg/ml. The accuracy for both intraday and interday was between 10-110%, and the % CV Error was< 10%. The LLOD was established at 100ng/ml and LLOQ Fig. 2.
FIG. 1: UHPLC CHROMATOGRAM INDICATING THE LINEARITY OF THE METHOD
FIG. 2: THE CALIBRATION CURVE OF PARACETAMOL (10 - 150.00 µG/ML) IN WATER
Fig. 2 shows the UHPLC chromatogram indicating the linearity of the method. In addition, the method was applied to a study with high sensitivity. All the validation parameters were performed according to ICH guidelines. The validation data confirmed that the UHPLC method in aqueous solutions is sensitive, accurate, and precise, with the intraday and interday precision assessment (% CV) less than 2% (accepted range ± 10%). The intraday and interday % accuracy of paracetamol ranged from 98%-100% which was in the accepted range of (90-110%), in Table 1.
TABLE 1: SUMMARY OF THE ACCURACY AND PRECISION RESULTS OF THE PROPOSED UHPLC METHOD
| Proposed methods | Paracetamol Concentration (µg/ml) | Mean | % Accuracy | SD | % CV |
| Intraday Validation (Accuracy & Precision) | 25.00 | 25.1018 | 100.4073 | 0.1052 | 0.4190 |
| 75.00 | 76.2811 | 101.7081 | 0.1362 | 0.1785 | |
| 125.00 | 123.7828 | 99.0262 | 0.1791 | 0.1447 | |
| Interday Validation (Accuracy & Precision) | 25.00 | 25.0946 | 100.3849 | 0.0830 | 0.3308 |
| 75.00 | 76.2956 | 101.7274 | 0.1603 | 0.2101 | |
| 125.00 | 123.6797 | 98.9438 | 0.1889 | 0.1527 |
The chromatographic technique demonstrated that the LLOQ and the LLOD were0.625 μg/ml and 0.156 μg/ml respectively Fig. 3A, B & 4. Three pharmaceutical formulations of paracetamol tablets, Panadol, Panadol (Cold & flu) and Panadol Extra were analyzed by the UHPLC method. The peak area of the sample solution was measured and the amount of paracetamol present in the tablet formulation was determined through the calibration curve. The assay results are summarized in Table 2. The percentage content of the active ingredient in the three paracetamol tablets displayed acceptable values of 99.01%-100.83%, which comply with the ICH acceptance criteria of (90%-110%).
FIG. 3A: UHPLC CHROMATOGRAM OF 30 ΜG/ML SELECTED FROM CALIBRATION CURVE OF VARYING CONCENTRATIONS OF PARACETAMOL
FIG. 3B: THE LOWER LIMIT OF QUANTIFICATION (LLOQ) OF PARACETAMOL WITH THE SAME Y AXIS
TABLE 2: PARACETAMOL ASSAY TESTS WERE ANALYZED BY THE UHPLC
| S. no. | Name of Product | Strength
(mg) |
Test product concentration (µg/ml) | Average (of six samples) | %Assay | STDV | %RSD |
| 1 | Panadol | 500 | 100 | 99.0157 | 99.0157 | 0.12165 | 0.12286 |
| 2 | Panadol (Cold+ flu) | 500 | 100 | 100.8360 | 100.8360 | 0.11545 | 0.11449 |
| 3 | Panadol Extra | 500 | 100 | 99.54483 | 99.54483 | 0.13107 | 0.13167 |
The chromatogram showed the separation of paracetamol (500 mg tablets) Panadol, Panadol (Cold+ flu) and Panadol Extra both made to an injectable dilution of 100μg/ml in Fig. 5, 6 & 7.
FIG. 4: THE LOWER LIMIT OF DETECTION (LLOD) OF PARACETAMOL
FIG. 5: THE CHROMATOGRAM OF PANADOL EXTRA 500 MG TABLETS
FIG. 6: THE CHROMATOGRAM OF PANADOL 500 MG TABLETS
FIG. 7: THE CHROMATOGRAM OF PANADOL COLD AND FLU 500 MG TABLETS
DISCUSSION: The results also demonstrated high sensitivity, allowing for drug detection at the nano-gram level and recoveries between 99.01%-100.83%, which confirmed the effectiveness of the UHPLC method in quantifying paracetamol in different pharmaceutical formulations. The validated analytical techniques in the present study ensured accuracy, precision, linearity, and sensitivity compared with the previously reported HPLC methods. The present study was carried out using different analysis methods that were validated according to ICH guidelines. The validated analytical methods enable accurate and sensitive quantification of paracetamol. The UHPLC method was more accurate, sensitive, precise, linear, and. These analytical methods are applicable in any in-vitro experiments such as quality control and stability tests.
CONCLUSION: This study highlights the value of UHPLC analysis using affordable and readily available solvents commonly found in any quality control laboratory. This rapid method with a 3 min runtime, combined with reduced solvent consumption, offers substantial economic benefits.
ACKNOWLEDGEMENT: We express our gratitude and appreciation for the help and support received in the process of this research article finalization. I also extend my heartfelt thanks to KAIMRC and department of drug analysis for their unwavering support.
Data Availability Statement: Data will be made available on request.
Credit Authorship Contribution Statement: Badran Alyahya, Abdullatif Al-Dulaimi and Wesam Abdel-Razaq: contributed to the methodology and writing the manuscript. Ibrahim Farh, Waleed Alhussaini, Amani Kurdi: Supervised all practical work, and reviewed the manuscript. Abdulkareem AlBekairy, Imadul Islam, Sultan Alqahtani and Mahmoud Mansour: Suggested the research point and experimental design.
Declaration of Competing Interest: The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.
Funding: The authors disclosed receipt of the following financial support for the research, authorship, and/or publication of this article: This work was supported by a grant from the King Abdullah International Research Center, National Guard Health Affairs, Riyadh, Saudi Arabia (grant no. SP22R/218/09). The funding agency had no role in the study design, data collection and analysis, decision to publish, or preparation of the manuscript.
CONFLICTS OF INTEREST: Nil
REFERENCES:
- de’Angelis GL, Vincenzi F, Fornaroli F, Buonvicino D and Chiarugi A: New perspectives for optimizing fever and pain management in pediatrics: evidence supporting therapeutic awareness in clinical practice. Italian Journal of Pediatrics 2025; 51: 255. DOI: 10.1186/s13052-025-02107-3.
- Paul IM and Walson PD: Acetaminophen and ibuprofen in the treatment of pediatric fever: a narrative review. Current Medical Research and Opinion 2021; 37: 1363-1375. DOI: 10.1080/03007995.2021.1928617.
- dos Santos WT, Gimenes DT, de Almeida EG, Eiras SdP, Albuquerque YD and Richter EM: Simple flow injection amperometric system for simultaneous determination of dipyrone and paracetamol in pharmaceutical formulations. Journal of the Brazilian Chemical Society 2009; 20: 1249-1255. DOI: 10.1590/s0103-50532009000700008.
- Alam P, Shakeel F, Ali A, Alqarni MH, Foudah AI, Aljarba TM, Alkholifi FK, Alshehri S, Ghoneim MM and Ali A: Simultaneous determination of caffeine and paracetamol in commercial formulations using greener normal-phase and reversed-phase HPTLC methods: A contrast of validation parameters. Molecules 2022; 27: 405. DOI: 10.3390/molecules27020405.
- Vu Dang H, Truong Thi Thu H, Dong Thi Ha L and Nguyen Mai H: RP‐HPLC and UV spectrophotometric analysis of paracetamol, ibuprofen, and caffeine in solid pharmaceutical dosage forms by derivative, Fourier, and wavelet transforms: A comparison study. Journal of analytical methods in chemistry 2020; 2020: 8107571. DOI: 10.1155/2020/8107571.
- Wahba IA, Hassan SA, Fayed AS and El-Mosallamy SS: Impurity profiling of paracetamol toxic impurities in pharmaceutical combination with ibuprofen and chlorzoxazone using HPLC and TLC densitometric methods. BMC Chemistry 2025; 19: 110. DOI: 10.1186/s13065-025-01466-6.
- Pereira FJ, Rodríguez-Cordero A, López R, Robles LC and Aller AJ: Development and validation of an RP-HPLC-PDA method for determination of paracetamol, caffeine and tramadol hydrochloride in pharmaceutical formulations. Pharmaceuticals 2021; 14: 466. DOI: 10.3390/ph14050466.
- Paul K, Gowda JB and Shankar S: A Simple and Rapid RP-HPLC Method for simultaneous estimation of Paracetamol and Pamabrom in Combinational Dosage Form. Research J of Pharmacy and Technology 2022; 15: 2047-2052. DOI: 10.52711/0974-360x.2022.00338.
- Gaikwad SS, Kalkate SD, Bankar AA and Bansode AS: Quality by design optimization and validation of a HPTLC-MS method for simultaneous estimation of paracetamol and prochlorperazine from bulk and formulation. International Journal of Mass Spectrometry 2024; 495: 117171. DOI: 10.1016/j.ijms.2023.117171.
- Kalambate PK, Dhanjai, Sinha A, Li Y, Shen Y and Huang Y: An electrochemical sensor for ifosfamide, acetaminophen, domperidone, and sumatriptan based on self-assembled MXene/MWCNT/chitosan nanocomposite thin film. Microchimica Acta 2020; 187: 402. DOI: 10.1007/s00604-020-04366-9.
- Cortés K, Triviño JJ and Arancibia V: Simultaneous voltammetric determination of acetylsalicylic acid, caffeine and paracetamol in pharmaceutical formulations using screen‐printed carbon electrode. Electroanalysis 2023; 35: 202200484. DOI: 10.1002/elan.202200484.
- Hajmalek S, Jahani S and Foroughi MM: Simultaneous voltammetric determination of tramadol and paracetamol exploiting glassy carbon electrode modified with FeNi3 nanoalloy in biological and pharmaceutical media. Chemistry Select 2021; 6: 8797-8808. DOI: 10.1002/slct.202102341.
- Mugwiza E, Hahirwa I and Umumararungu T: A UV Spectrophotometric method for simultaneous determination of Ibuprofen and Paracetamol concentration in suspensions. Rwanda Medical Journal 2023; 80: 49-56. DOI: 10.4314/rmj.v80i3.7
- Wang H, Zhang S, Li S and Qu J: Electrochemical sensor based on palladium-reduced graphene oxide modified with gold nanoparticles for simultaneous determination of acetaminophen and 4-aminophenol. Talanta 2018; 178: 188-194. DOI: 10.1016/j.talanta.2017.09.021.
- Abbas RF, Allawi AG, Abdulhassan NM and Mahmoud NH: Spectrophotometric Determination of Paracetamol using a Newly Synthesized Chromogenic Reagent 4-[(2-amino-1, 3thiazol-4-yl) amino] nitro benzene. Egyptian Journal of Chemistry 2020; 63: 4681-4693. DOI: 10.21608/ejchem.2020.17136.2053.
- Mabaya TM, Kakumba JM, Makola MM, Puati AM, Kialengila DM, Mbay JMT and Kindenge JM: Development and validation of a UV/visible spectrophotometric method for simultaneous assay of paracetamol and ibuprofen. Orapuh Journal 2025; 6: 1202-1202. DOI: 10.4314/orapj.v6i1.2.
- Younis MS and Othman NS: Indirect Spectrophotometric Assay of Paracetamol in Pharmaceutical Preparations. International Journal of Enhanced Research in Medicines & Dental Care 2018; 5: 23-29. DOI: 10.1063/5.0000318.
- Rahman MAA, Elghobashy MR, Zaazaa HE and El-Mosallamy SS: Novel analytical method based on chemometric models applied to UV–Vis spectrophotometric data for simultaneous determination of Etoricoxib and Paracetamol in presence of Paracetamol impurities. BMC Chemistry 2023; 17: 176. DOI: 10.1186/s13065-023-01095-x.
- Kelani KM, Fekry RA, Fayez YM and Hassan SA,: Advanced chemometric methods for simultaneous quantitation of caffeine, codeine, paracetamol, and p-aminophenol in their quaternary mixture, Scientific Reports 2024; 14: 2085. DOI: 10.1038/s41598-024-52450-4.
- Thanoon ES, Spectrophotometric determination of paracetamol using diazotization coupling reaction. Rafidain Journal of Science 2019; 28: 76-83. DOI: 10.33899/rjs.2019.159979.
- Aljeboree AM, Jasim LS, Abdulrazzak FH, Sahib IJ, Radia ND and Alkaim AF: Validated spectrophotometric approach for determination of paracetamol in pure and pharmaceutical dosage forms using oxidative coupling reaction. In. AIP Conference Proceedings. AIP Publishing LLC 2022; 2386: 030004.
- Dinç-Zor Ş, Dönmez ÖA and Bozdoğan AE: Application of Chemometrics-assisted HPLC-DAD Strategies for Simultaneous Determination of Paracetamol, Pseudoephedrine HCl, Dextromethorphan HBr, Doxylamine Succinate and Saccharin in Syrup Formulation, Current Pharmaceutical Analysis 2021; 17: 1043-1050. DOI: 10.2174/1573412916999200729173449.
- Burgess C and McDowall R: Quo vadis analytical procedure development and validation?, 2022. DOI: 10.56530/spectroscopy.mf9790d5.
How to cite this article:
Mansour MA, Alyahya B, Al-Dulaimi A, Farh I, Alhussaini W, Kurdi A, Alqahtani S, Abdel-Razaq WS, Islam I and Albekairy AM: Comparative performance of liquid chromatography and spectrophotometry for quantitative analysis of paracetamol in tablets formulations. Int J Pharm Sci & Res 2026; 17(3): 1025-32. doi: 10.13040/IJPSR.0975-8232.17(3).1025-32.
All © 2026 are reserved by International Journal of Pharmaceutical Sciences and Research. This Journal licensed under a Creative Commons Attribution-NonCommercial-ShareAlike 3.0 Unported License.
Article Information
29
1025-1032
753 KB
3
English
IJPSR
Mahmoud A. Mansour *, Badran Alyahya, Abdullatif Al-Dulaimi, Ibrahim Farh, Waleed Alhussaini, Amani Kurdi, Sultan Alqahtani, Wesam S. Abdel-Razaq, Imadul Islam and Abdullkaree M. Albekairy
King Abdullah International Medical Research King Abdullah International Medical Research Center, Ministry of National Guard Health Affairs, Riyadh, Saudi Arabia.
mansoura@ksau-hs.edu.sa
12 October 2025
28 October 2025
02 November 2025
10.13040/IJPSR.0975-8232.17(3).1025-32
01 March 2026