COMPARISON OF BACTEC VERSUS CONVENTIONAL METHOD FOR CULTURE OF STERILE BODY FLUIDS IN A TERTIARY CARE HOSPITAL
HTML Full TextCOMPARISON OF BACTEC VERSUS CONVENTIONAL METHOD FOR CULTURE OF STERILE BODY FLUIDS IN A TERTIARY CARE HOSPITAL
Simranjit Kaur, Vishal Sharma * and Deepak Arora
Department of Microbiology, G. G. S. MCH, Faridkot, Punjab, India.
ABSTRACT: Aim & Background: Body fluids are sent to clinical microbiology laboratory for culture to find the etiological agent causing the infection. This study was conducted to compare the culture results of sterile body fluids processed simultaneously by both conventional method and BACTEC blood culture system. Methods: Sterile body fluids except blood were included in the study. A total of 61 body fluid samples were received and simultaneously processed by both the culture methods-conventional and BACTEC. Results: In this study, overall culture positivity for BACTEC was 85.25% as compared to 59.02% by conventional system. According to the results of our study, the overall culture positivity was increased by 26.23% by BACTEC method than the conventional method. BACTEC blood culture system detected more pathogenic isolates (52.45%) than the routine conventional culture method (21.31%) and it was found to be statistically significant (p value=<0.005). The most common pathogenic micro-organisms detected by conventional and BACTEC methods in this study were Gram-negative followed by Gram-positive. The mean time to detection of pathogenic isolates by BACTEC and conventional method was 1.19(+/- 0.39) days and 2.00(+/- 1.00) days and the difference was statistically significant (p value=0.013). Conclusion: The study recommends the use of automated blood culture system for culture of sterile body fluids as it was found to improve the yield of isolates with reduced time to detection. However, cost may be a limiting factor in resource constrained settings.
Keywords: Conventional culture method, BACTEC, Sterile body fluids, Infection
INTRODUCTION: Sterile body fluid infections can be caused by both Gram-positive and Gram-negative bacteria. Gram-negative bacteria like Escherichia coli, Klebsiella pneumoniae, Acinetobacter baumanii and Pseudomonas aeruginosa are the frequently isolated pathogens followed by Gram-positive bacteria like Staphylococcus aureus and Enterococcus 1-4. Presence of microorganisms in normally sterile body sites causes life-threatening infections 2, 5.
For early and accurate diagnosis of those infections, sterile body fluids are required to be sent to microbiology laboratory for culture 2, 5. The conventional method used for culture of sterile body fluids, is culture on a solid medium with or without an enrichment broth. In addition, various procedures, such as filtration or centrifugation, are required to concentrate organisms within the specimen.
However, there are many challenges for recovery of microorganisms by conventional culture methods: Low number of microorganisms; low volume of sample e.g: CSF and synovial fluids; initiation of antibiotic therapy prior to collection of sample which lead to the false negative results; fastidious organisms may be missed if proper media or supplements are not incorporated into the culture method. Conventional methods of culture are now being replaced by automated culture systems like BACTEC (Becton Dickinson Diagnostic Instrument) because of reduced time for culture and ease of laboratory work using machines and higher isolation rate against conventional methods. The main objective of this study was to compare the culture results of sterile body fluids processed simultaneously by both conventional method and BACTEC blood culture system.
MATERIALS AND METHODS: This study was conducted after taking permission from the Institutional Ethics Committee. The IEC approval number of this study was No. BFUHS/2K22p-TH/7232. A total of 61 sterile body fluids except blood were included in this study.
Gram stain was done immediately followed by direct plating of sample onto conventional media –Blood agar, MacConkey agar, Chocolate agar and the media was incubated at 37°C for 18-24 hours as per standard protocol. Out of the remaining sample, half of the sterile body fluid was inoculated into aerobic BACTEC culture bottle (Becton Dickinson Diagnostic Instrument Systems) and remaining half was inoculated into conventional enrichment broth (Brain-heart infusion broth). Conventional broth was incubated and processed as per standard protocol. Subcultures from the broth were done at the end of 24 hour, 72 hour and 7th day of incubation. If no growth was detected at the end of 7 days of incubation, it was reported negative. The BACTEC bottle was incubated in BACTEC 9120 system till signal for growth was detected. When positive signal was observed, the bottles were unloaded from instrument and Gram stain and cultures on Blood agar, MacConkey agar, Chocolate agar were performed as per the standard microbiological protocols. Identification of the micro-organisms was done by observing the colony morphology, gram staining and biochemical reactions as per standard microbiological protocol 6. BACTEC bottles giving no signal were reported negative after 5 days of incubation.
The isolation rate of pathogenic microorganisms, the time to culture positivity, i.e., time taken from incubation to growth detection for conventional and signal positive for BACTEC 9120 was noted and mean time to detection was calculated for both methods. Kirby-Bauer’s disc diffusion method was used for antibiotic susceptibility testing and results were interpreted according to Clinical and Laboratory Standards Institute (CLSI) guidelines 6. The antibiotics used for Gram-negative organisms were (concentration /disk in microgram)-Cefotaxime (30µg), Ceftriaxone (30µg), Amikacin (30µg), Ciprofloxacin (5µg), Ceftazidime (30µg), Cefepime (30µg), Meropenem (10µg), Piperacillin-Tazobactam (100/10µg): The antibiotics used for Gram-positive organisms were (concentration / disk in microgram) Ampicillin (10µg), Erythromycin (15µg), Cefoxitin (30µg), Ciprofloxacin (5µg), Amikacin (30µg), Linezolid (30µg), High-Level Gentamicin (120 µg) ATCC Control for Gram-negative micro-organisms: ATCC E. coli 25922 ATCC Control for Gram-positive micro-organisms: ATCC S. aureus 25923
Vancomycin Screen Agar Test: Method was considered for interpretation of susceptibility / resistance to vancomycin in Staphylococcal isolates as per CLSI guidelines.
Statistical Analysis: The data pertaining to sociodemographic and other clinical variables was entered in the form of data matrix in Microsoft “Excel” and analysed using Statistical Package for Social Sciences (SPSS).Detailed data of the patients was recorded in the proformas. Appropriate test of significance was applied using the statistical package for social sciences. At any point of time, p value of 0.05 was used as a threshold for significance.
RESULTS: The most common age group in this study belonged to 41-60 years (33%), followed by 21-40 years (28%) and </=20 years (26%). The least common age group was 61-80 years (13%). Out of 61 body fluid samples, majority were ascitic fluid (32.8%), pleural fluid (32.8%), followed by CSF (29.5%). The least were of pericardial fluid (3.3%) and synovial fluid (1.6%) as shown in Fig. 1. Overall 52 specimens were detected culture positive by BACTEC as compared to 36 culture positive specimens by conventional method. Exact McNemar’s test was applied and according to this evaluation, the difference between the methods in respect to overall culture positive specimens was found to be statistically significant ( p value<0.005) as shown in Table 1.
FIG. 1: SAMPLE-WISE DISTRIBUTION OF DIFFERENT BODY FLUIDS
TABLE 1: COMPARISON OF CULTURE METHODS ACCORDING TO OVERALL CULTURE POSITIVE /NEGATIVE
Method | BACTEC positive | BACTEC negative | Total |
Conventional
Positive |
36 | 0 | 36 |
Conventional Negative | 16 | 9 | 25 |
Total | 52 | 9 | 61 |
Conventional Positive/BACTEC Positive=culture positive p value <0.005*
Pathogenic Isolates and Contaminants: BACTEC method detected 32 pathogenic isolates whereas the conventional culture method detected only 13 pathogenic isolates. McNemar-Bowker’s test was applied and according to this evaluation, the difference between the methods in respect to pathogenic microorganisms was found to be statistically significant (p value=<0.005) as shown in Table 2.
TABLE 2: COMPARISON BETWEEN METHODS ACCORDING TO CONSIDERING THE MICROORGANISMS AS PATHOGENIC/ CONTAMINANT
Method | BACTEC Pathogenic | BACTEC Contaminant | BACTEC Negative | Total |
Conventional Pathogenic | 13 | 0 | 0 | 13 |
Conventional Contaminant | 12 | 11 | 0 | 23 |
Conventional Negative | 7 | 9 | 9 | 25 |
Total | 32 | 20 | 9 | 61 |
p value<0.005.
Out of 32 pathogenic isolates detected by BACTEC, the most common microorganism was Escherichia coli followed by Klebsiella pneumoniae, Methicillin-resistant Staphylococcus aureus (MRSA), Pseudomonas aeruginosa, Methicillin-susceptible Staphylococcus aureus (MSSA), Methicillin-resistant Coagulase-Negative Staphylococci (MRCoNS), Citrobacter freundii, Enterococcus and Acinetobacter baumanii. Out of 13 pathogenic isolates detected by the conventional method, the most common microorganism was MRSA followed by Escherichia coli, Klebsiella pneumoniae, Pseudomonas aeruginosa, Enterococcus, Citrobacter freundii and Acinetobacter baumanii as shown in Table 3.
TABLE 3: DISTRIBUTION OF PATHOGENIC MICROORGANISMS ISOLATED BY BOTH THE CULTURE METHODS
Microorganism | Growth in conventional method | Growth in BACTEC system |
MRSA | 4(30.7%) | 5(15.6%) |
MSSA | 0 | 3(9.4%) |
MRCoNS | 0 | 3(9.4%) |
Enterococcus | 1(7.7%) | 1(3%) |
Escherichia coli | 2(15.4%) | 7(22%) |
Klebsiella pneumoniae | 2(15.4%) | 6(18.8%) |
Citrobacter freundii | 1(7.7%) | 2(6.3%) |
Pseudomonas aeruginosa | 2(15.4%) | 4(12.5%) |
Acinetobacter baumanii | 1(7.7%) | 1(3%) |
Total | 13 | 32 |
The comparison of time to detection of pathogenic isolates by both the methods is shown in Table 4. The mean time to detection of pathogenic isolates was 1.19 (+/- 0.39) days by BACTEC as compared to 2.00 (+/-1.00) days by conventional method. Independent-Samples T test was applied and according to this evaluation, the difference between the methods in respect to mean time to detection of pathogenic isolates was found to be statistically significant (p value=0.013) Table 4.
TABLE 4: COMPARISON OF TIME TO DETECTION OF GROWTH OF PATHOGENIC ORGANISMS
Serial no. | Method | Day 1 | Day 2 | Day 4 | Total no. of pathogenic isolates |
1. | Conventional | 4(30.8%) | 7(53.8%) | 2(15.4%) | 13 |
2. | BACTEC | 26(81.2%) | 6(18.8%) | 0 | 32 |
DISCUSSION: Body fluids like pleural fluid, peritoneal fluid, CSF, synovial fluid and pericardial fluid are usually sterile. There are certain common pathogenic bacteria like Escherichia coli, Klebsiella species, Staphylococcus aureus, Neisseria meningitidis, Non fermenting Gram-negative bacilli, Pseudomonas, Acinetobacter species, which invade and infect the sterile body fluids leading to morbidity and life-threatening infections 2, 5, 7. Hence infections of sterile body fluids are a medical emergency and need an early diagnosis and effective treatment 7.
In the present study, overall culture positivity for BACTEC was 85.25% as compared to 59.02% by conventional system. The recovery rate with the BACTEC culture method was higher than with conventional culture methods and the difference was statistically significant (p value<0.005). Similar results of higher positivity by BACTEC culture method were obtained by many authors in their studies on sterile body fluids 8-12.
The results of this study showed that overall 36(59.02%) specimens were culture positive by both the methods, 16(26.23%) were positive by BACTEC only and 9(14.75%) were sterile by both the methods. Conventional culture did not detect bacteria in any instance not detected by BACTEC. Similar findings have been reported by Cetin et al., and Mengeloglu et al 9, 10. According to the results of our study, the overall culture positivity was increased by 26.23% by BACTEC method than the conventional method. In this study, pathogenic microorganisms were isolated from 13(21.31%) specimens by both culture systems; however, for 19 specimens (31.14%), growth was detected only with the BACTEC system. BACTEC blood culture system detected more pathogenic isolates (52.45%) than the routine conventional culture method (21.31%) and it was found to be statistically significant. In other studies conducted by various authors, similar findings of increased isolation of pathogenic microorganisms by BACTEC as compared to conventional culture were seen 8-13. In this study, the most frequently isolated microorganisms with the use of both BACTEC cultures and conventional methods were Gram-negative bacteria followed by Gram-positive bacteria. Similar findings of predominance of Gram-negative bacteria as compared to Gram-positive bacteria have been reported by various authors 2, 8. However, Cetin et al., reported that the most frequently isolated microorganisms recovered were Gram-positive cocci followed by Gram-negative bacilli 9.
The time to culture positivity, i.e., time taken from incubation to growth detection for conventional and signal positive for BACTEC 9120 was noted. Mean time to detection (MTTD) is the average of the time to culture positivity for all the isolates in a particular category 12. Using BACTEC, majority (100%) of the pathogenic isolates were detected on/before day 2 whereas, during the same time, only eleven (84.6%) pathogenic isolates were detected by conventional method. This is similar to the studies conducted by other authors who have reported majority of their isolates being detected within 48 hours by automated systems as compared to the conventional culture methods 8, 12. Similar reduced time to detection by BACTEC method was found in previous studies 9, 10, 12, 13. Some authors have reported a mean detection time of 19-24 hours by BACTEC and 5-7 days by conventional method 14-16.
Since, BACTEC is a continuous monitoring system, and instrument takes readings every 10 minutes, detection occurs earlier and can be reported in hours. However, conventional method, by its very nature of processing, cannot be reported positive before 24 hours 12, 17. There are many factors which may contribute to increased isolation by BACTEC. The resins, present in the BACTEC bottle and the dilutional effect of the liquid in the bottles decrease the inhibitory effects of antimicrobial substances 5, 18. BACTEC plus media are reported to effectively remove many antimicrobials. They use ion exchange and non-ionic adsorbent resins to remove antimicrobials thereby enhancing the recovery of micro-organisms.
In terms of speed, sensitivity and shortened turnaround time, BACTEC automated culture system helps in earlier presumptive reporting of the pathogen as compared to conventional culture methods 13. This would significantly affect the change of empirical antibiotic therapy given to the patient, eventually aiding in therapy and the final clinical outcome 13.
However, cost may be a limiting factor for BACTEC in resource poor laboratories 11, 17. Another concern is the increased risk of needle stick injury for technical staff as a result of the need to use syringes to inoculate sterile body fluids into BACTEC vials. These disadvantages should be balanced against the clinical advantage of better isolation rate of pathogens from infected body fluids 18.
In the present study, the most common isolate among Gram-positive cocci was MRSA and these strains showed maximum resistance to cefoxitin (100%), followed by ciprofloxacin (60%), amikacin (40%), erythromycin (40%). MSSA showed maximum resistance to erythromycin (100%), followed by ciprofloxacin (33%), amikacin (33%). MRCoNS showed maximum resistance to erythromycin (100%) and cefoxitin (100%), followed by ciprofloxacin (67%), amikacin (67%). None of the above isolate was resistant to vancomycin and linezolid in the present study.
Enterococcus showed maximum resistance (100%) to ampicillin, erythromycin, ciprofloxacin but was 100% sensitive to vancomycin, linezolid and high-level gentamicin. Most common isolate among Gram-negative bacilli was Escherichia coli which showed maximum resistance to cefotaxime (86%), followed by ceftriaxone (57%), ciprofloxacin (43%), amikacin (29%), piperacillin-tazobactam (29%). Klebsiella pneumoniae showed maximum resistance to cefotaxime (83%), followed by ceftriaxone (50%), ciprofloxacin (33%), amikacin (33%), piperacillin-tazobactam (17%). Pseudomonas aeruginosa, showed maximum resistance to ceftazidime (50%), cefepime (50%), ciprofloxacin (50%), followed by amikacin (25%) and piperacillin tazobactam (25%).
Citrobacter freundii showed 50% resistance to cefotaxime, ceftriaxone but was 100% sensitive to ciprofloxacin, amikacin and piperacillin-tazobactam. Only one strain of Acinetobacter baumanii was isolated which showed 100% resistance to ceftazidime, cefepime, ciprofloxacin, amikacin and piperacillin-tazobactam. All the Gram-negative isolates in the present study showed 100% susceptibility to meropenem.
Increased resistance to antimicrobials was observed which corroborates with other studies by different authors 19, 20. High level of resistance to the first line drugs in the present study is a worrisome problem and it warrants routine and regular surveillance of the antibiogram.
CONCLUSION: Body fluids may be infected by both Gram-positive and Gram-negative bacteria. Conventional culture methods may not detect the causative organism from the sterile body fluids and also these are time-consuming and laborious. Automated systems significantly shorten the time to diagnosis, thus allowing rapid diagnosis and early administration of appropriate treatment. The study recommends the use of automated blood culture system for culture of sterile body fluids. Regular monitoring and surveillance of prevalence of organisms causing infection of body fluids is required for formulating an antibiotic and infection control policy so as to guide the clinicians in choosing appropriate antibiotics before a culture report is available thus preventing the development of antimicrobial resistance.
ACKNOWLEDGEMENTS: Nil
CONFLICTS OF INTEREST: Nil.
REFERENCES:
- Madigubba H, Deepashree R, Monika, Gopichand P and Sastry AS: Bacteriological profile and antimicrobial susceptibility pattern insterile body fluid specimens from a tertiary care hospital, South India. J Curr Res Sci Med 2020; 6(2): 96-101.
- Shrestha LB, Bhattarai NR and Khanal B: Bacteriological profile and antimicrobial susceptibility pattern among isolates obtained from body fluids. J Nepal Health Res Counc 2019; 17(2): 173-177.
- Sodani S and Hawaldar R: Bacteriological profile and antibiotic sensitivity pattern in various body fluids-A retrospective study. Indian J Microbiol Res 2020; 7(1): 51-58.
- Sharma B, Kasana D and Ganbhir S: Bacterial profile, their antibiogram and a light on emerging multi drug resistant microorganisms from sterile body fluids in a northern tertiary care hospital in India. J Bacteriol Mycol Open Access 2018; 6(4): 249‒252.
- Hughes JG, Vetter EA, Patel R, Schleck CD, Harmsen S, Turgeant LT and Cockerill FR: 3rd. Culture with BACTEC Peds Plus/F bottle compared with conventional methods for detection of bacteria in synovial fluid. J Clin Microbiol 2001; 39(12): 4468-71.
- CLSI. Performance Standards for Antimicrobial Susceptibility Testing. 32nd ed. CLSI supplement M100. Clinical and Laboratory Standards Institute 2022.
- Harshika YK, Shobha MKR, Patil AB and Smita NR: A study on bacteriological profile and antimicrobial resistance pattern from various body fluids of patients attending the tertiary care hospital, KIMS, Hubli. Indian J Microbiol Res 2018; 5(4): 530-534.
- Deb A, Mudshingkar S, Dohe V and Bharadwaj R: Bacteriology of body fluids with anevaluation of enrichment technique to increase culture positivity. J Evolution Med and Dental Sci 2014; 3(72): 15230−15238.
- Cetin ES, Kaya S, Demirci M and Aridogan BC: Comparison of the BACTEC blood culture system versus conventional methods for culture of normally sterile body fluids. Adv Ther 2007; 24(6): 1271-7.
- Mengeloglu Z, Tas T, Bucak O, Kocoglu E and Kucukbayrak A: Comparison of classical methods versus BACTEC blood culture system for culture ofnormally sterile body fluids. Russian Open Med J 2015; 4(4): e0401.
- Akcam FZ, Yayli G, Uskun E, Kaya O and Demir C: Evaluation of BACTEC microbial detection system for culturing miscellaneous sterile body fluids. Res Microbiol 2006; 157(5): 433-6.
- Surase PV, Nataraj G, Pattamadai K, Mehta PR, Pazare AR, Agarwal MC and Nanavati RN: An appropriately performed conventional blood culture can facilitate choice of therapy in resource constrained settings-comparison with BACTEC 9050. J Postgrad Med 2016; 62(4): 228-234.
- Udayan U and Dias M: Evaluation of BACTEC blood culture system for culture of normally sterile body fluids. Indian J Crit Care Med 2014; 18(12): 829-830.
- Kaur A, Soodan PS and Singh VA: Comparative evaluation of conventional blood culture with BACTEC 9050 for bacterial isolates in clinically suspected cases of fever of unknown origin. IOSR J Dent Med Sci 2014; 13: 17-21.
- Gopi A, Kaushik A and Harindranath D: Evaluation of time to positivity in detection of micro-organism using BACTEC 9050 in a tertiary care hospital over a period of one year. J Evol Med Dent Sci 2014; 3: 13999-14005.
- Gopi A, Ravikumar KL, Ambarish MG, Shenoy S, Shree SK, Ashwini KV, Arpita S and Debjani C: Time to positivity of microorganisms with BACTEC 9050: An 18-month study among children of 28 days to 60 months in a South Indian tertiary hospital. Int J Microbiol Res 2011; 2(1): 12-17.
- Nutt L, Orth H, Goodway J and Wasserman E: Superior detection of pathogens in synovial fluid by the BACTEC 9240 Peds Plus/F system compared to the conventional agar-based culture method. South Afr J Epidemiol Infect 2010; 25(4): 11-14.
- Sorlin P, Mansoor I, Dagyaran C and Struelens MJ: Comparison of resin-containing BACTEC plus Aerobic/F medium with conventional methods for culture of normally sterile body fluids. J Med Microbiol 2000; 49(9): 787-791.
- Sheikhbahaei S, Abdollahi A, Hafezi-Nejad N and Zare E: Patterns of antimicrobial resistance in the causative organisms of spontaneous bacterial peritonitis: a single-centre, six-year experience of 1981 samples. Int J Hepatol 2014; 2014: 917856.
- Shume T, Tesfa T, Mekonnen S, Asmerom H, Tebeje F and Weldegebreal F: Aerobic bacterial profile and their antibiotic susceptibility patterns of sterile body fluids among patients at Hiwot Fana Specialized University Hospital, Harar, Eastern Ethiopia. Infect Drug Resist 2022; 15: 581-593.
How to cite this article:
Kaur S, Sharma V and Arora D: Comparison of BACTEC versus conventional method for culture of sterile body fluids in a Tertiary Care Hospital. Int J Pharm Sci & Res 2023; 14(11): 5404-09. doi: 10.13040/IJPSR.0975-8232.14(11).5404-09.
All © 2023 are reserved by International Journal of Pharmaceutical Sciences and Research. This Journal licensed under a Creative Commons Attribution-NonCommercial-ShareAlike 3.0 Unported License.