DEVELOPMENT AND VALIDATION OF A HPLC-UV METHOD FOR SIMULTANEOUS DETERMINATION OF MOXIFLOXACIN HYDROCHLORIDE AND KETOROLAC TROMETHAMINE IN OCULAR FORMULATION
HTML Full TextReceived on 24 March, 2014; received in revised form, 19 June, 2014; accepted, 29 June, 2014; published 01 July, 2014
DEVELOPMENT AND VALIDATION OF A HPLC-UV METHOD FOR SIMULTANEOUS DETERMINATION OF MOXIFLOXACIN HYDROCHLORIDE AND KETOROLAC TROMETHAMINE IN OCULAR FORMULATION
Tapas Majumder, Sarbojit Kundu*, Subrata Kumar Ray and Prasanta Kumar Barat
Central Drugs Laboratory, (Govt. of India),3, Kyd Street, Kolkata-700 016, India
ABSTRACT: A simple, rapid, and sensitive high-performance liquid chromatographic method with UV detection has been developed and validated according to the ICH guidelines for the quantitation of Moxifloxacin Hydrochloride (MOX), Ketorolac Tromethamine (KET) in pharmaceutical dosage form. Chromatographic separation was carried out in a Zorbax eclipse plus, C18 column (250 mm × 4.6 mm; 5 µm particle size) with simple mobile phase composition of 10 mM Potassium dihydrogen phosphate buffer with Triethylamine (pH 3.14) and acetonitrile (40:60, v/v) at a flow rate of 0.5 mL min-1 where detector was set at 302 nm with a total run time of 8 mins. The method was linear over the concentration range of 40-100, µg mL-1 with a correlation coefficient of 0.9891 and 0.994. Limit of quantifications (LOQ) of 13.3, 26.3 and limit of detections (LOD) 4.4, 8.7 µg mL-1 for MOX, and KET respectively. Accuracy and precision values of both within-run and between-run obtained from six different sets of three quality control (QC) samples analyzed in separate occasions for both the analytes ranged from 98.13% to 99.75% and 0.95% to 2.15%, respectively. Extraction recovery of analytes in pharmaceutical formulation from 97.82% to 98.68%. The developed and validated method was successfully applied to quantitative determination of MOX and KET in pharmaceutical formulation.
Keywords: |
High performance liquid chromatography, UV Spectrophptometry, Moxifloxacin Hydrochloride & Ketorolac Tromethamine.
INTRODUCTION: Moxifloxacin hydrochloride (MOX), 1-Cyclopropyl-6-fluoro-1, 4-dihydro-8-methoxy-7-[(4aS,7aS)-octahydro-6H-pyrrolo [3, 4-b]pyridin-6-yl]-4-oxo-3 quinolinecarboxylic acid hydrochloride (Figure 1), is a synthetic fourth-generation broad-spectrum fluoroquinolone antibiotic. It acts by inhibiting DNA gyrase, a type II topoisomerase and topoisomerase IV, which are involved in DNA replication and metabolism 1.
Ketorolac tromethamine (KET) (±)-5-benzoyl-2,3-dihydro-1H-pyrrolizine-1-carboxylic acid, a compound with 2-amino-2-(hydroxymethyl)-1,3-propanediol, a pyrrolizine carboxylic acid derivative (Figure 2), a nonsteroidal anti-inflammatory drug, is indicated for short-term management of moderate to severe pain and shows a high incidence of side effects like gastric bleeding1. The primary mechanism of action responsible for ketorolac's anti-inflammatory, antipyretic and analgesic effects is the inhibition of prostaglandin synthesis by competitive blocking of the enzyme cyclooxygenase (COX). The combination of Ketorolac Tromethamine with Moxifloxacin Hydrochloride is extensively used for the treatment of postoperative inflammation and infection following cataract surgery 1.
Various methods like UV spectrophotometry 2, 3, 4 estimation in biological fluids by HPLC 5, 6 HPTLC 8 UV spectrophotometry 9 methods for deterination of MOX with other drug are reported in literature.
Ketorolac Tromethamine was determined using HPTLC 7, 8 HPLC 10, 14 methods. However, a Few analytical methods were reported for the simultaneous determination of Moxifloxacin Hydrochloride and Ketorolac Tromethamine in a mixture, namely, rapid liquid chromatography–electrospray ionization mass spectrometry (LC-MS) 1 HPLC using a diode array detector 13, RP-HPLC 14 and HPTLC 15.
An extensive review of the literature did not revealed any HPLC method for simultaneous determination of both drugs. Therefore, attempts were made to develop and validate simple, precise, and sensitive, isocratic reverse phase stability indicating high performance liquid chromatographic method for simultaneous determination of both drugs in pharmaceutical formulations.
The proposed methods were validated in compliance with the ICH guidelines and were successfully applied for determination of Moxifloxacin Hydrochloride and Ketorolac Tromethamine in their pharmaceutical formulations.
MATERIALS AND METHODS:
Chemicals and reagents:
MOX, KET were procured from pharmaceuticals industry. Acetonitrile HPLC Grade, Phosphoric acid analytical grade from Merck (Mumbai, India), Potassium dihydrogen phosphate analytical grade from Merck (Mumbai, India), Triethylamine analytical grade from Merck (Mumbai, India), HPLC-grade water (resistivity 18.2 MΩ) cm was generated from a Milli-Q water purification system, was used throughout the analysis.
Samples are procured from pharmaceutical industry and they are considered as Sample I and Sample II respectively and both the samples are ocular formulations.
Instrumentation and chromatographic conditions:
HPLC apparatus consisted of a Agilent Technology (USA) Model, G1311A Quaternary pump, G1365D variable wave length UV detector, Auto-sampler (G1329A), Column oven (G13368) and EZ CHROM ELITE Version 331SOP software. Chromatographic separation was performed isocratically at room temperature using a Agilent Zorbax eclipse Plus, C18 (250 mm x 4.6 mm, 5 µm particle size) Mobile phase consisted of a mixture of Potassium dihydrogen phosphate buffer (2.7363 gm to 1000ml+ 1 ml Triethylamine, pH 3.14 by dil H3PO4) and Acetonitrile (40:60, v/v) at a flow rate of 0.5 ml min-1 and sample injection of 20 µL was injected at 25oC. Eluent was monitored with a UV detector set at 302 nm.
Preparation of stock and working solutions:
24.5 gm of MOX and 25.5 gm of KET taken in a 25 ml volumetric flask and dissolving in mobile phase to obtain concentration of 1000 µg/mL. The stock solution stored in amber colored labeled volumetric flask at 8 0C.
Preparation of calibration standards and quality control (QC) samples:
Five calibration standards (CC) of both MOX and KET at concentration of: 20, 40, 60, 80, and 100 µg mL-1 were prepared by spiking 0.2, 0.4, 0.6, 0.8, 1.0 ml respectively to 10 ml by Mobile phase. Three QC sample 40, 60, 80 µg mL-1 were used. All standards stored in amber colored labeled volumetric flask at 8 0C.
Sample preparation:
2ml of sample diluted with mobile phase mixed properly. Samples were further diluted by mobile phase which have final concentration 20µg mL-1 of both MOX and KET and then injected into the HPLC system.
Method validation:
The proposed methods were validated in compliance with the ICH guidelines and were successfully applied for determination of Moxifloxacin Hydrochloride and Ketorolac Tromethamine in their pharmaceutical formulations.
This method was validated to meet the acceptance criteria with the ICH guidelines of method validation. 16
Selectivity:
Selectivity of the method was determined by analyzing blank (mobile phase), to demonstrate the lack of chromatographic interference at the retention time of the analytes.
Limit of detection (LOD), Limit of quantitation (LOQ) and Linearity:
Limit of detection (LOD), Limit of quantitation (LOQ) was determined by the following equation 3.3xσ/S and 10xσ/S, where as σ = standard deviation of the response and S = slope of the calibration curve. Calibration curves were acquired by plotting the peak-area of the analytes against the nominal concentration of calibration standards. Analyte concentration of different CC and QC samples were prepared as mentioned above.
Accuracy and precision:
Accuracy of an analytical procedure is the closeness of agreement between accepted conventional true values (reference values) and the values found. The accuracy of the proposed methods was tested by the determination of MOX and KET at different concentration levels within the linear range of each compound.
Precision was studied by determination of intra-day and inter-day precision. Intra-day precision was determined by injecting five standard solutions of three different concentrations on the same day and inter-day precision was determined by injecting the same solutions for three consecutive days. Relative standard deviation (RSD %) of the peak area was then calculated to represent precision.
Extraction recovery:
Recoveries of MOX and KET were determined in the addition standard (40, 60, 80 µg mL-1) by comparing the experimental and true values (40, 60, 80 µg mL-1).
RESULTS AND DISCUSSION:
Optimization of chromatography:
In the work chromatographic conditions were optimized to obtain high sensitivity, reproducibility, and sample throughput for the determination of MOX and KET. Both analytes contain basic nitrogen atoms and therefore have the potential to cause the peak tailing due to interaction of this basic nitrogen atoms with the silanol group of the stationary phase during chromatographic separation. In order to obtain symmetrical peaks with better resolution, the chromatographic condition i.e. pH of the buffer, concentration of the organic modifier and silanol blocker were optimized.
Various chromatographic condition such as mobile phase composition, analytical column with different packing materials (C8, C18,Phenyl, Cyano) and configuration (10, 15, 25 cm ) were used to obtain sharp peak with reduce tailing, and better resolution with no peak impurity. Finally Agilent Zorbax Eclipse Plus C18 column was selected which provided reduced peak tailing and acceptable peak purity index.
Eclipse plus C18 packing is made by first chemically bonding a dense monolayer of dimethyl-n-octadecylsilane stationary phase to a specially prepared, an improved ultra-high purity (>99.995% SiO2), ZORBAX Rx-SIL porous silica support. The bonded-phase packing is doubly endcapped using proprietary reagents and procedures to obtain maximum deactivation of the silica surface. Mobile phase composition was selected base upon the peak parameter (symmetry, tailing, resolution and peak purity index etc.), run time, case of preparation and cost.
The most suitable mobile phase composition was found to be acetonitrile and phosphate buffer (pH 3.14) in the ratio of 60:40 (v/v), respectively. Under the chromatographic conditions outline, highly symmetrical, sharp peaks of MOX and KET were obtain at retention time about 4.2 and 7.0 min. respectively.
Selectivity:
The method was found to selective as no significant interfering peak are observed at the retention times of MOX and KET which were 4.2, and 7.0 min respectively. Total chromatographic run time was 10.0 min. (Figure 5 & 6) shows the representative chromatograms of blank spiked with analytes .
Limit of detection (LOD), Limit of quantitation (LOQ) and Linearity:
Limit of detection (LOD) was established 4.4 and 8.7 µg mL-1 for MOX and KET respectively. Limit of quantification (LOQ) was established 13.3 and 26.3 µg mL-1 for MOX and KET respectively. Calibration curves were linear over the concentration range 40–100 µg mL-1 of both analytes. Regression coefficient 0.9891 and 0.994 for MOX and KET respectively. (Figure 3 & 4)
Standard curve had a reliable reproducibility over the standard concentrations across the calibration range. All back-calculated concentrations did not differ from the theoretical value as no single calibration standard point was dropped during the validation.
Accuracy and precision:
The accuracy and precision of the proposed methods were tested by the determination of MOX and KET at different concentration levels within the linear range of each compound. The low SD (< 1) of six determinations indicated the high accuracy and precision of the proposed method. Collective results are shown in Tables 1 & 2.
The inter- and intra-day determination of MOX and KET over 3 consecutive days by the same analyst using the same instrument is shown in Tables 1 & 2. The low RSD (< 2%) reflects the ruggedness of the methods.
TABLE 1: ASSESSMENT OF ACCURACY AND PRECISION OF MOXIFLOXACIN HYDROCHLORIDE
QC Sample
(µg mL-1) |
Mean
(µg mL-1) |
S.D. | R.S.D.
% |
Accuracy
% |
|
Intra Day
(n=6) |
40.00 | 39.71 | 0.70 | 1.76 | 99.28 |
60.00 | 59.52 | 1.28 | 2.15 | 99.19 | |
80.00 | 79.38 | 1.25 | 1.57 | 99.22 | |
Inter Day
(n=18)
|
40.00 | 39.54 | 0.70 | 1.78 | 98.86 |
60.00 | 59.43 | 1.16 | 1.95 | 99.05 | |
80.00 | 79.22 | 1.35 | 1.70 | 99.03 |
S.D. = Standard deviation; R.S.D. (%) (Relative standard deviation) = [(S.D./Mean) X 100]; Accuracy (%) = [(Mean / Conc. Added) X 100]; n = number of replicates.
TABLE 2: ASSESSMENT OF ACCURACY AND PRECISION OF KETOROLAC TROMETHAMINE
QC Sample
(µg mL-1) |
Mean
(µg mL-1) |
S.D. | R.S.D.
% |
Accuracy
% |
|
Intra Day
(n=6) |
40.00 | 39.90 | 0.56 | 1.40 | 99.75 |
60.00 | 59.26 | 0.78 | 1.31 | 98.76 | |
80.00 | 79.08 | 0.75 | 0.95 | 98.84 | |
Inter Day
(n=18)
|
40.00 | 39.43 | 0.70 | 1.78 | 98.58 |
60.00 | 58.88 | 0.74 | 1.25 | 98.13 | |
80.00 | 78.78 | 1.00 | 1.28 | 98.47 |
S.D. = Standard deviation; R.S.D. (%) (Relative standard deviation) = [(S.D./Mean) X 100]; Accuracy (%) = [(Mean / Conc. Added) X 100]; n = number of replicates.
Extraction recovery:
Recovery results were found to be satisfactory as these were consistent, precise and reproducible are summarized in Table 3.
TABLE 3: EXTRACTION RECOVERY OF ANALYTES (n = 6)
Analyte | QC Sample
(µg mL-1) |
Extraction recovery (%) | R.S.D.
(%) |
40.00 | 98.68 | 0.64 | |
MOX | 60.00 | 98.55 | 1.06 |
80.00 | 97.82 | 1.07 | |
40.00 | 98.07 | 0.88 | |
KET | 60.00 | 98.35 | 0.80 |
80.00 | 98.14 | 0.68 |
R.S.D. (%) (Relative standard deviation) = [(Standard deviation /Mean) X 100]; n = number of replicates.
Implementation to Pharmaceutical formulation:
This newly developed method was applied to determine the MOX and KET in pharmaceutical formulation (eye drops). Result ware summarized in Table 4.
TABLE 4: ESTIMATION MOXIFLOXACIN AND KETOROLAC IN DIFFERENT FORMULATION
Sample | Analyte | Concentration found
%(w/v) |
% |
Sample I
Sample II |
MOX
KET
MOX KET |
0.511
0.519 0.493 0.494 |
102.18
103.98 98.69 98.85 |
CONCLUSION: Here, we have developed and validated a HPLC-UV method that has significant advantages over the previously published method as it provides simple mobile phase composition for chromatographic separation, shorter run time for analysis, simple sample preparation as well as improved sensitivity. Therefore, this new method leads to a simple, feasible, cost effective, rapid method with high degree of accuracy and specificity to quantify simultaneously MOX and KET in pharmaceutical formulations with HPLC-UV. It will be extremely helpful for successfully analyzing the MOX and KET in ocular formulations.
ACKNOWLEDGEMENT: The authors are thankful to The Director, Central Drugs Laboratory (CDL), Govt. of India, Kolkata and Dr. Ayan Das.
REFERENCES:
- B. Raju, M. Ramesh, R. M. Borkar, R. Padiya, S. K. Banerjee and R. Srinivas, “Development and validation of liquid chromatography–mass spectrometric method for simultaneous determination of moxifloxacin and ketorolac in rat plasma”, Journal of Biomedical Chromatograghy, (2012) 1341–1347.
- S. k. Sahu, Md. Afzal Azam, Dipansu Sahu, and M. Banarjee Spectrophotometric estimation of Moxifloxacin in bulk and its pharmaceutical formulations, pharmacologyonline 2:491-502 (2011).
- Ravindra kakade, Ghodekar sn., Wagh p., jagdale s. HPTLC method for the simultaneous estimation of Moxifloxacin Hydrochloride and Ketorolac Tromamethamine in opthalmic dosage form.
- Dinesh M. Dhumal, Atul A. Shirkhedkar and Sanjay J. Surana, Quantitative determination of Moxifloxacin hydrochloride in bulk and ophthalmic solution by UV-spectrophotometry and first order derivative using area under curve, Scholars Research Library Der Pharmacia Lettre, 2011, 3(3):453-456.
- Najma sultana, Mahwish Aktar, Sana shamim, Somia gul, Muhammad saeed arayne, Moona mehaboob khan HPLC assay for moxifloxacin in bulk and pharmaceutical formulations and serum application to invitro metal interactions Journal of Chinese Chemical Society, Vol- 57, No. 4A, 2010.
- Xu YH, Li D, Liu XY, Li YZ, Li J HPLC assay with ultraviolet detection for Moxifloxacin; validation and application to a pharmacokinetic study in Chinese volunteers.
- Vandana dhillon, Alok kumar chaudary, A validated HPTLC method for the estimation of moxifloxacin hydrochloride in tablets, pharmaceutical methods October-December 2010 vol-1, issue-1, pg.no:-54-56.
- Ghandi.L.R, Dewani A.P, Bakal R.L and Dr Chandewar A.V, Absorption ratio method for the estimation of moxifloxacin hcl & ketorolac tromethamine in their combined dosage form by uv-visible spectroscopy, IJPRD 2011; Vol 3(7): September 2011(21- 26).
- G. Sunil, M. Jambulingam, S. A. Thangadurai, D. Kamalakannan, R. Sundaraganapathy and C. Jothimanivannan, “Development and Validation of Ketorolac Tromethamine in Eye Drop Formulation by RP- HPLC Method”, Arabian Journal of Chemistry, available online January 2013.
- R. S. Chaudhary, S. S. Gangwal, K. C. Jindal and S. Khanna, “Reversed-phase high performance liquid chromatography of ketorolac and its applicationto bioequivalent studies in human serum”, Journal of Chromatography B: Biomedical Sciences and Applications, 1993;180–184.
- W. At and I. J. Massey, “Simultaneous determination of Ketorolac and its hydroxylated metabolite in plasma by high performance liquid chromatography”, Journal of Chromatography B, 1990; 241–246
- S. N. Razzaq, I. U. Khan, M. Ashfaq and I. Mariam, “Stability indicating HPLC method forsimultaneous determination of Moxifloxacin hydrochloride and Ketorolac Tromethamine in pharmaceutical formulations”, Química Nova, 2012; 1216-1221.
- D. Patel, M. Patel and K. Patel “Simultaneous RP-HPLC Estimation of Moxifloxacin Hydrochloride and Ketorolac Tromethamine in Ophthalmic Dosage Forms”, Asian Journal of Research In Chemistry, 2012; 697
- R.S. Kakade, S. N. Ghodekar, P. Wagh and S. Jagdale, “HPTLC Method for Simultaneous Estimation of Moxifloxacin HCl and Ketorolac Tromethamine in Ophthalmic Dosage Form”, Journal of R.J.S.P.M.S College of Pharmacy, available online February 2012.
- Validation of Analytical Procedures, Proceedings of the International Conference on Harmonisation (ICH), Geneva. Commission of the European Communities 1996.
- Martidale The complete Drug Reference, 36th edition.
- High Performance Liquid Chromatography, Quantitative Analysis of Pharmaceutical Formulations, P.D.Sethi.
Article Information
36
2902-2907
467
2463
English
IJPSR
Tapas Majumder, Sarbojit Kundu*, Subrata Kumar Ray and Prasanta Kumar Barat
Central Drugs Laboratory, (Govt. of India), 3, Kyd Street, Kolkata-700 016, India.
sarbojit18@yahoo.co.in
24 March, 2014
19 June, 2014
29 June, 2014
http://dx.doi.org/10.13040/IJPSR.0975-8232.5(7).2902-07
01 July, 2014