DEVELOPMENT AND VALIDATION OF A STABILITY INDICATING RP-HPLC METHOD OF APIXABAN IN COMMERCIAL DOSAGE FORM
HTML Full TextDEVELOPMENT AND VALIDATION OF A STABILITY INDICATING RP-HPLC METHOD OF APIXABAN IN COMMERCIAL DOSAGE FORM
Israa Al-Ani 1, Mohammed Hamad 2, Ramadan Al-Shdefat 3, Kenza Mansoor 4, Felicia Gligor 5 and Wael Abu Dayyih * 4
Faculty of Pharmacy 1, Al-Ahliyya Amman University, Jordan.
Department of Basic Sciences 2, College of Science and Health Professions, King Saud Bin Abdulaziz University for Health Sciences, Jeddah, Saudi Arabia.
Faculty of Pharmacy 3, Jadara University, Jordan.
Faculty of Pharmacy and Medical Sciences 4, University of Petra, Jordan.
Faculty of Medicine 5, Lucian Blaga University of Sibiu, Romania.
ABSTRACT: The aim of this study is to develop a sensitive, specific, rapid, and precise reverse-phase high-performance liquid chromatography (RP-HPLC) method and validate it for the quantification of process-related and degradation impurities of apixaban; an anticoagulant drug. The chromatographic separation was achieved on a Sigma-Aldrich’s Ascentis Express® C18 (4.6 mm × 100 mm, 2.7 µ) HPLC column with a runtime of 40 min. Mobile phase-A and mobile phase-B were phosphate buffer and acetonitrile, respectively. The column oven temperature was set at 35 °C, and the photodiode array detector was set at 225 nm. The newly developed method was utilized to detect nine process-related impurities (Imp-1 to Imp-9) in a test sample of Apixaban. Forced degradation study was carried out under acidic, alkaline, oxidative, photolytic, and thermal conditions to demonstrate the stability-indicating nature of the developed RP-HPLC method. The developed method was validated as per ICH guidelines and found to be specific, precise, sensitive, and robust. In conclusion, the RP-HPLC method was successfully developed and validated then effectively applied to analyze both; Apixaban drug substance and product.
Keywords: |
Apixaban, Validation, Stability, HPLC, Eliquis
INTRODUCTION: Apixaban is an anticoagulant drug chemically known as l-(4-methoxyphenyl)-7-oxo-6-[4-(2-oxopiperidin-l-yl) phenyl]- 4, 5, 6, 7-tetrahydro- lH-pyrazolo [3, 4-c] pyridine- 3- carbo-xamide Fig. 1. 1
FIG. 1: APIXABAN CHEMICAL STRUCTURE 1
“Eliquis”, the trade name of apixaban, is prescribed to treat atrial fibrillation (a heart rhythm disorder) and to lower the risk of stroke caused by blood clots. It is a selective, reversible, direct inhibitor of factor Xa indicated to reduce the risk of stroke and systemic embolism in patients with non-valvular atrial fibrillation 2, 3. Furthermore, it is indicated to prevent deep vein thrombosis (DVT), which may lead to pulmonary embolism, especially after hip or knee replacement surgery 4-6. It was designed by Aderis Pharmaceuticals and jointly developed by Pfizer and Bristol-Myers Squibb. “Eliquis” was officially permitted in the USA and Europe on December 2012 and January 2010, respectively 7.
During the development of any drug substance or product in the pharmaceutical industry, the development of an accurate and efficient analytical method to determine the quality of the product is a crucial step 8. The current research focused on the development of an efficient analytical method using a core-shell chromatography column for the quantitative analysis. Silica particles are mostly used in ultra-pressure liquid chromatography (UPLC) columns. They are porous with a particle size of sub-2 µ, thus yields good resolution, speed, sensitivity, and at the same time, they offer high backpressure 9-11. The cost of UPLC investment and its columns for routine analysis is higher than that of conventional HPLC and its columns. Hence, core-shell columns have been preferred to overcome such limitations 12, 13. Several analytical methods reported in the literature described the investigation of the pharmacokinetics of apixaban. Most of them were in the content of Apixaban and/or its metabolites investigated in human plasma by liquid chromatography-mass spectrometry method 14-16. Nevertheless, researchers did not succeed in providing the details of process and degradation related impurities due to the employed stress conditions. One of the articles reported on stability, indicating the HPLC method for apixaban bulk drug sample, found the method to be non-selective at our end 17. Hence, there is a need for the development of a selective, fast, and stability-indicating HPLC method on the core shell column.
To the best of our knowledge, no method on the core-shell column has been reported for the determination of apixaban and its potential process-related impurities in drug substance and drug product for regular analysis and stability studies in quality control laboratories.
The main objective of the current research was to develop a precise, fast, sensitive, and stability-indicating RP-HPLC method for the determination of process and degradation related impurities of apixaban. And to validate the method according to “USP Validation of Compendia Procedures” and ICH guidelines 18-23.
MATERIALS AND METHODS:
Chemicals and Reagents: Apixaban reference standard was kindly given as a gift sample from United Pharmaceuticals. Methanol and acetonitrile (HPLC grade) Milli-Q water. Hydrochloric acid HCL (AR) 37%, sodium hydroxide NaOH (AR-grade), ortho-phosphoric acid (HPLC - grade) 85%, ammonium HOAC –NH3 acetate (GR-Q), potassium dihydrogen phosphate -KHPO4(AR-grade), hydrogen peroxide H2O2 (ACS-Q) 50%, formic acid HCOOH (AR-Q), triethylamine TEA (HPLC- grade), sodium lauryl sulphate Na-L-SO4, polysorbate Na-L-PO4- 80 (GR-Q) were procured from Merck-R, Fischer Scientific-R and Sigma -J.
Instruments: The HPLC system used was an Agilent 1260 infinity equipped with a quaternary solvent pump, a thermostated autosampler, a degasser, UV/VWD detector, and a column oven. A double beam UV-Visible spectrophotometer (Shimadzu 1800), USP Type II-Paddle apparatus (Labindia DS2000 with Ismatec high precision multichannel pump as autosampler), photostability chamber (Atlas Sun test XLS+), HPLC water purifier (Millipore–Merck Millipore), orbital shaker (Thermo scientific) and a centrifuge (REMI research centrifuge) were used for this analysis.
Preparation of Solutions:
UV Studies:
Preparation of Standard Stock Solution (10 ppm): Apixaban reference standard (1 mg) was accurately weighed and transferred into a 100 ml volumetric flask. 30 ml solvent (acetonitrile: water 90:10 v/v) was added, and the mixture was sonicated. The solution was diluted up to the mark with the diluents to give the standard stock solution.
The λmax was determined using a UV-VIS spectrophotometer. A working standard range from 2 to 7 ppm was prepared from the stock solution of 10 ppm and used for linearity studies.
Preparation of Standard Stock Solution (90 ppm): Apixaban reference standard (25 mg) was weighed and transferred into a 25 ml volumetric flask to obtain a solution of 1000 ppm in duplicate. 18 ml of methanol was added and sonicated. This solution was diluted to volume with methanol and mixed well. 9.0 ml of standard stock solution was taken and diluted to 100 ml with diluent to give a standard stock solution of 90 ppm.
Preparation of Sample Solution: Five apixaban tablets were weighed and transferred to 250 ml volumetric flask. Then, 100 ml of water was added and sonicated for 1 h to disperse the contents. Further, 50 ml of water was added, and the flask was kept in an orbital shaker at 50 rpm for 2 h at room temperature (RT). The flasks were made up to the mark with water. The solution was then filtered using a 0.45 μ Whatman Teflon filter. The filter was saturated by discarding the first 10 ml of filtrate. 5.0 ml of the above filtrate was further diluted to 100 ml with diluents and mixed well to get the desired concentration.
Compatibility Studies of Filters: Filter compatibility tests were conducted to get a particle-free solution, which also should reproduce a 100% response as per the respective content. The filtrate from each filter was compared with the centrifuged sample. The filter complying with all the parameters and showing the same concentration as that of the centrifuged sample was selected for further studies. Five intact tablets were weighed and transferred to 250 ml volumetric flask. 25 ml of miliQ water was added and sonicated for 10 min. 150 ml of acetonitrile was added and sonicated for 20 min with intermittent swirling. This was further diluted with acetonitrile mixed and filtered using a 0.45μ Whatman Teflon filter. 10 ml of filtrate was discarded, and a further 5 ml was diluted to 100 ml with diluent. A 20 ml aliquot was subjected to centrifuge at 5000 rpm for 20 min, and another 20 ml per filter was passed through the filters. The four solutions were chromatographed and were compared for peak shape and peak area. The filter that matched the centrifuged sample data was selected for further studies.
Selection of Diluent (HPLC Compatible Solvent) and Solution Stability Studies: The stability of the solution under study was established by keeping the solution at room temperature for 24 h. A standard solution was prepared using acetonitrile: water (90: 10 v/v, 50: 50 v/v, and 10:90 v/v), methanol: water (90: 10 v/v, 50: 50 v/v and 10: 90 v/v). The samples were injected at different time intervals of 5 min, 1 d, and 1 w.
Selection of Flow Rate and load Volume: Flow rates (0.8 ml/min. and 1.3 ml/min.) and loading volumes (1, 4, 5, 6, 10, 20, 50 μl) were analyzed to get a precise and reproducible peak of apixaban with minimum tailing or fronting.
Selection and Optimization of Mobile Phase: Various mobile-phase compositions like acetonitrile: water (50: 50v/v, 90: 10 v/v, 60: 40 v/v, 40: 60 v/v and 10: 90 v/v), acetonitrile: methanol: water (80: 10: 10 v/v, 50: 25: 25 v/v, 30: 40: 30 v/v, 40:30: 30 v/v and 10: 45: 45 v/v), acetonitrile: TEA buffer in water (50:50 v/v ratio with 0.25 ml and 0.5 ml concentration of TEA per 500 ml), acetonitrile: TEA and FA buffer in water (50: 50 v/v ratio with0.25 ml: 0.25 ml and 0.5 ml: 0.5 ml concentration of TEA and FA per 500 ml.), acetonitrile: buffer [pH range 3 to 7.5] (50: 50 v/v, 80: 20 v/v and 20: 80 v/v) were evaluated in an effort to get symmetrical, sharp and reproducible peaks of apixaban.
Selection and Optimization of Stationary Phase: Using the temporary developed chromatographic method a variety of columns with different attributes like polarity, particle size (3.5 μ and5 μ), carbon loading (C8 and C18) and dimensions were screened in an effort to arrive at the best column for the given analysis. Some of the columns studied were; Symmetry C18 50 × 4.6 mm 3.5 μ, Akzonobel - Kromasilc C18 150 × 4.6 mm 5 μ, Waters Xterrac C18 150 × 4.6 mm 3.5 μ, BDS Hypersilthermoc C8 50 × 4.6 mm 5 μ, FortiscC8 50 × 4.6 mm 5μ, and Agilent zorbaxc RX C18 150 × 4.6 mm, etc.
Selection of Dissolution Media: Apixaban standard solution was prepared, and the final dilution was done using water, 0.1 N HCl, pH 4.5 buffer, pH 6.8 buffer, SLS media and tween media, respectively. Five tablets were weighed and transferred to a 250 ml volumetric flask. 100 ml of water was added and sonicated for 1h to disperse the contents. Further, 50 ml of water were added, and the flask was kept in an orbital shaker at 50 rpm for 2 h at room temperature. The solutions were made up to the mark with water. The solutions were mixed well and filtered using a 0.45μ Whatman Teflon filter. The filter was saturated by discarding 10 ml of filtrate, and 5.0 ml of the above solution was diluted to 100 ml with diluent and mixed well. Six tablets were weighed and transferred to each jar of Dissolution Rate Apparatus USP II was chosen, and the temp. was maintained at 37 °C, stirring rate 75 rpm, and the volume of dissolution media was 900 ml. At specified intervals of time (5, 10, 20, 30, 45, and 60 min), 10 ml of sample aliquots were withdrawn and replenished with the same volume of fresh media. The samples were filtered with a 0.45μ Whatman Teflon filter. Around 8 ml of the sample was discarded to saturate the filter. Similarly, the procedure was repeated using 0.1 N HCl, pH 4.5 buffer, pH 6.8 buffer, 0.1% SLS media, 0.2% SLS media and 0.1% Tween 80 media, 0.2% Tween 80 media. The release of the apixaban into the respective media was calculated using the formula;
Release = At/As × Ws/25 × 25/5 × P/100 × 100/LC
Where, AT = Peak area response of apixaban in the chromatogram obtained from test solution, AS = Average peak area response of apixaban in the chromatograms obtained from replicate injections of standard solution, WS = Weight of apixaban standard taken in mg in the standard stock solution, P = % Purity of apixaban standard, LC =Label claim of apixaban in mg per tablet.
Assay of Apixaban Tablets: The samples were analyzed using the developed chromatographic method, and the % content of the drug in the tablets was determined.
TABLE 1: GRADIENT SEQUENCE FOR RELATED SUBSTANCES STUDIES
Time
|
% acetonitrile | % formic (FA) : Triethylamon
(TEA) 1:1 v/v in warwe |
0
10 20 30 40 |
4
20 50 80 45 |
55
80 50 20 55 |
0.1 mg of known impurity (Acetal impurity, RS-13) was accurately weighed and transferred to a 10 ml volumetric flask. 5 ml of diluent was added, sonicated, and diluted to volume with diluent to get 10 ppm of impurity solution. The chromatograms of apixaban reference standard, drug sample and known impurity solutions were compared using the gradient run below in Table 1.
Related Substances Studies: The isocratic mode of the RP-HPLC method was modified to a gradient mode as given below with a run time of 40 min.
Forced Degradation Studies: With slight modifications, the developed chromatographic method was used to get accurate results. The optimized chromatographic method was kept same and run time was 12.0 min. Forced degradation studies of apixaban were conducted using solution stressors like distilled water, 0.1 N HCl, 0.1 N NaOH, hydrogen peroxide 50%. Photostability studies were conducted at a flux of 4.6 × 1014 photons for 24 h, and thermostability studies were conducted at a temperature of 80 °C for 24 h. The results were compared with the untreated standards. Five tablets were weighed accurately and transferred to 250 ml volumetric flask. 25 ml miliQ water was added and sonicated for 10 min. 150 ml of acetonitrile was added and sonicated for 20 min with intermittent swirling. 25 ml of the respective stressor was added and heated at 80 °C for 24 h on a water bath previously maintained at 80 °C. The sample was cooled to RT and neutralized. It was filtered using a 0.45 μ Whatman Teflon filter. 10 ml of filtrate was used for filter saturation, and the remaining 7 ml was used. 5 ml of filtrate was diluted to100 ml with diluent. The results were compared with the untreated sample. For the solid degradation analysis, the tablets were treated with the respective stressor, and the results were compared with the untreated sample.
Validation Studies:
Accuracy: The accuracy of the method was determined by calculating % recovery. A known amount of apixaban was added to a placebo and the amounts were estimated by measuring the peak area. These studies were carried out in triplicate over the specified concentration range and the amount of apixaban was estimated by measuring the peak area ratios. The percentage recovery and standard deviation of percentage recovery was calculated.
Precision: The precision of the method was determined in terms of Intra-day and inter-day precision. For intra-day precision studies, a standard solution of 90 ppm was injected at various time intervals and percent related standard deviation (%RSD) was estimated. The inter-day precision was studied by injecting the same concentration of the standard solution on consecutive days, and the % RSD of the signal was calculated. The repeatability, intermediate precision, and reproducibility of the developed method was determined
Specificity: Specificity is the ability to assess the analyte unequivocally in the presence of components etc. The blank (diluent), placebo, standard (90 ppm), sample (90 ppm) were prepared and injected to prove that the method developed was specific to apixaban
Linearity and Range: The linearity of the method was determined at six concentration levels ranging from 1-7 ppm of apixaban. A regression line was plotted of peak area v/s concentration. The correlation coefficient and equation of the regression line were calculated. The interval of lowest assessed concentration to the highest is the linearity range of the procedure.
Limit of Detection (LOD) and Limit of Quantification (LOQ): The detection limit of an individual analytical procedure is the lowest amount of analyte in a sample which can be detected but not necessarily quantitated as an exact value. The quantitation limit of an individual analytical procedure is the lowest amount of analyte in a sample, which can be quantitatively determined with suitable precision and accuracy. Where, = the standard deviation of the response, Slope = slope of the calibration curve.
Robustness: Robustness of the developed method was studied by changing the flow rate and column temperature. The effect of flow rate was studied by keeping all chromatographic conditions the same except the flow rate, i.e., 1.2 ml/min and in the next run 1.4 ml/min, respectively. Similarly, the effect of temperature was studied by keeping all chromatographic conditions the same except the temperature, i.e., 35 °C and in the next run with 25 °C respectively.
System Suitability: The system suitability parameters like retention time, the number of USP theoretical plates, USP tailing, peak area, and peak height were evaluated.
RESULTS AND DISCUSSION:
UV Studies: The UV spectrum of apixaban in acetonitrile: water (90: 10 v/v) indicated the λmax to be 256 nm Fig. 2. The response was linear, and the regression equation was found to be y = 0.1331x + 0.0858, as seen in UV-scan, Fig. 2.
FIG. 2: UV-SPECTRUM SCAN OF APIXABAN IN ACETONITRILE TO WATER (90:10 v/v) IN CALIBRATION CURVE
Filter Compatibility Studies and Stability: A temporary chromatographic method was used for the selection of the best filter. Whatman Nylon w/GMF membrane filter pore size of0.45μ was selected after comparing the peak area and concentration Table 2 of centrifuged drug solution to other drug samples.
TABLE 2: FILTER COMPATIBILITY STUDIES
Sample no. | Sample analyzed form | Concentration (ppm) | Peak Area |
1
2 |
Centrifuge
Whatman |
89.631134583
89.621134437 |
0.45µ
Nylon w/GMF |
Based on these observations, acetonitrile: water; 90: 10 v/v was selected as the diluent for further studies. The stability of the drug sample was determined by keeping the sample in contact with the diluent for 5 min, 24 h and 1 w, which indicated no change in the concentration Fig. 3-5 and Table 3.
TABLE 3: EFFECT OF DILUENT ON CHROMATOGRAM AFTER A SPECIFIC TIME INTERVALS
Sample no. | Analysis time (after) | Retention time (min) | Peak Area |
1
2 2 |
5 min
24 min one min |
9.010
9.010 8.010 |
1134437
1134437 1134432 |
Selection of Mobile Phase and Stationary Phase: According to the observations of peak shape and retention time as Fig. 6, a flow rate of 1.3 ml and loading volume 10 μl was selected. Based on the data in Table 5, acetonitrile: formic acid: tri-ethylamine in water 50: 50 v/v (0.5 ml: 0.5 ml) was selected as the mobile phase for the analysis as shown in Fig. 7.
FIG. 6: FLOW RATE EFFECT AND VOLUME LOADING ON SHAPE OF PEAK FOR APIXABAN
FIG. 7: APIXABAN CHROMATOGRAM WITH ACETONIT-RILE ACN: FORMIC ACID: TRI-ETHYLAMINE (BUFFER) IN H2O2 50: 50 v/v (0.5 ml: 0.5 ml) -MOBILE PHASE
A variety of columns with different attributes like polarity, particle size, carbon loading, and dimensions were screened in an effort to arrive at the best column for developing the analytical method of apixaban in API and tablet formulation. Table 4 Akzonobel Kromasilc C18 150 × 4.6 mm 5μ was selected as the column for the analysis of Apixaban as Fig. 8.
Table 5 shows the results of the effect of the stationary phase on the chromatogram.
TABLE 4: EFFECT OF MOBILE PHASE COMPOSITION ON CHROMATOGRAM
Sample no. | Mobile phase composition Observation | ||
Retention time | Peak shape | ||
1
2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 |
Acetonitrile: water in 50:50 v/v
Acetonitrile: water 90 : 10 v/v Acetonitrile: water 60:40 v/v Acetonitrile: water in 40:60 v/v Acetonitrile: water in 10:90 v/v Acetonitrile: methanol : water 80:10 :10 v/v Acetonitrile: methanol : water 50:25:25 v/v Acetonitrile: methanol : water 30:40:30 v/v Acetonitrile: methanol : water 40:30:30 v/v Acetonitrile: methanol : water 10:45:45 v/v Acetonitrile: triethylamine in water 9.991 (TEA buffer) 50:50 v/v Acetonitrile: triethylamine in water 9.100 (TEA buffer) 50:50 v/v Acetonitrile: triethylamine in water 9.931 (TEA buffer) 50:50 v/v Acetonitrile: triethylamine in water 9.434 (TEA buffer) 50:50 v/v Acetonitrile: buffer pH 3.0 in 50:50:v/v Acetonitrile: buffer pH 4.5 IN 50:50 v/v Acetonitrile: buffer pH 6.8 in 50:50 v/v Acetonitrile: buffer pH 7.5 in 50:50 v/v |
8.913
6.341 6.341 12.252 11.472 9.001 7.882 9.624 12.113 8.624 0.25 ml 0.5 0.25ml : 0.25 ml 0.5ml : 0.5 ml 7.021 9.982 3.718 - |
Tailing
Sharp Tailing Sharp Shoulder Sharp Sharp Sharp Sharp Sharp Tailing Tailing Sharp Sharp Merged Merged Tailing - |
FIG. 8: APIXABAN CHROMATOGRAM WHEN AKZONOBEL - KROMASIL© C18 150 × 4.6 mm, 5μ APPLIED
TABLE 5: EFFECT OF STATIONARY PHASE ON CHROMATOGRAM
Sample no. | Stationary phase | Observation | |
Retention time (min) | Peak shape | ||
1 | Symmetry © C18 50 × 4.6 mm 3.5 µ | 9.850 | Sharp |
2 | Waters Xterra© C18 150 × 4.6 mm 3.5µ | 8.989 | Broad peak |
3 | Fortis© C8 50 × 4.6 mm 5 µ | 8.989 | Symmetric |
4 | inertsil ODS© C18 250 × 4.6mm 5 µ | 9.434 | Sharp |
5 | AkzonobelKromasil© C18 150 × 4.6 mm 5 µ | 9.001 | Sharp |
6 | BDS hypersil thermo© C8 50 × 4.6 m 5 µ | 13.001 | Fronting |
7 | YMC© c 18 150 × 4.6 mm 5 µ | 14.101 | Broad peak |
8 | supelcosilSupelco©C18 300 × 3mm 3 µ | 9.130 | Sharp peak |
9 | Hypersil thermo© C18 300 × 3mm 5 µ | 8.022 | Sharp peak |
10 | Agilent ZORBAX©SB C18 150 × 4.6 mm 5 µ | 9.031 | Sharp |
11 | Agilent zorbax©RX C18 150 × 4. Mm 5 µ | 7.812 | Sharp |
12 | Synergy hydro phenomenex© C18 50 × mm 5 µ | 7.992 | Broad peak |
Selection of Dissolution Media: Different dissolution media were tried. After a series of trials using different solvents, Table 6 0.2% tween 80 was selected as the solvent which shown in Fig. 9.
FIG. 9: APIXABAN CHROMATOGRAM IN 0.2% TWEEN 80 IN WATER AS SOLVENT RATIO
TABLE 6: EFFECT OF VARIOUS MEDIA ON THE STABILITY OF APIXABAN
Sample no. | Solvent | Peak area | % content | |
Standard | Sample | |||
1 | Distilled water | - | - | No peak |
2 | 0.1 N HCl | 1921 | 1850 | 101.23 |
3 | pH 4.5 buffer | 10421 | 10393 | 99.73 |
4 | pH 6.8 buffer | 120105 | 121021 | 100.283 |
5 | pH 7.5 buffer | - | - | No peak |
6 | 0.1 % SLS | 1042301 | 1042498 | 99.63 |
7 | 0.2% SLS | 1042598 | 1042502 | 10.00 |
8 | 0.1% Tween 80 | 1075400 | 1075412 | 99.61 |
9 | 0.2% Tween 80 | 1134682 | 1132817 | 99.48 |
Dissolution Studies and Assay of Apixaban Tablets: The results of dissolution studies indicated that the developed RP-HPLC method was suitable for this purpose in Table 7.
TABLE 7: RELEASE STUDIES OF APIXABAN IN 0.2% TWEEN 80
Sample no.
Time (min) |
Peak area | % release | |
Standard | Sample | ||
15
210 315 430 545 660 |
11346804
11346805 11346805 11346807 11346801 11346801 |
81342
01212 41822 55817 084912 132817 |
42.26
44.02 47.58 66.38 95.27 99.48 |
The dissolution results indicated a release of 99.4% in 60 min at 75 rpm and the chromatogram in Fig. 10 indicates this. The assay of apixaban tablets was conducted using the developed and validated RP-HPLC method, and the drug content was found to be 99.9%. Fig. 9 and Table 7 show the results.
Related Substances (RS) and Forced Degradation Studies (FD): The modified analytical method with a run time of 12.0 min was used for the determination of impurities generated in the apixaban API sample after conducting forced degradation using various stressors. The results obtained in Table 9 indicated that the developed method was able to resolve all the degradation peaks from the Apixaban, which indicted in Fig. 12-17.
TABLE 8: FORCED DEGRADATION STUDIES OF APIXABAN
Sample no. | Stressor condition | Retention time of API (min) | Impurities
generate |
1 | Thermo stability | 6.071 | 3 |
2 | Photostability | 6.076 | 3 |
3 | Hydrolytic degradation | 6.102 | 10 |
4 | Oxidative degradation | 6.100 | 11 |
5 | Acid hydrolysis | 6.089 | 13 |
And 1 major impurity | |||
6 | Alkaline hydrolysis | 6.085 | 19 |
Validation Studies: The developed RP-HPLC method was validated according to the International Conference on Harmonization (ICH) Q2 guidelines for various parameters like accuracy, precision, specificity, linearity and range, LOD, and LOQ, robustness, and system suitability.
Accuracy: The accuracy of the developed method was evaluated by conducting recovery studies. The accuracy was calculated by assay method at concentrations of 50%, 100%, and 150%. The calculated amount of apixaban stock solutions were spiked to get the concentrations. The % mean recovery of spiked samples was found to be 99.82% Table 10.
TABLE 9: ACCURACY STUDIES OF APIXABAN
Level | Mean of concentration added (µ/mL) | Mean of concentration obtained (µ/mL) | % mean Recovery |
50% | 45.052 | 44.810 | 99.793 |
100% | 89.990 | 89.953 | 99.930 |
150% | 135.009 | 134.678 | 99.750 |
Precision: The intra-day precision studies were performed using apixaban reference standard solution of 90 ppm. The solution was injected at various time intervals, and the percent related standard deviation (% RSD) was determined. The inter-day precision was studied by injecting the same concentration of the standard solution on consecutive days, and the % RSD was calculated. The relative standard deviation obtained from 12 assay results was found to be 0.069% Table 11.
TABLE 10: PRECISION STUDIES OF APIXABAN
Precision parameter | Peak area | Mean | SD | %RSD | ||
Run 1 | Run 1 | Run 1 | ||||
Reproducibility | 1389984 | 1389953 | 1388300 | 1389412 | 963.433 | 0.069 |
Intermediate precision | 1389986 | 1389958 | 1388303 | 1389416 | 963.699 | 0.069 |
Repeatability | 1389983 | 1389956 | 1388309 | 1389416 | 958.785 | 0.069 |
Specificity: No interferences were observed in the chromatogram of apixaban standard due to the presence of excipients and blank Table 12.
Linearity and Range: A graph of peak area versus concentration (in ppm) was plotted for apixaban at a concentration range between 20-90 ppm. The linear regression equation and correlation coefficient (R2) were y = 8268.2x + 41452 and 0.995 respectively Fig. 17.
TABLE 11: SPECIFICITY STUDIES OF APIXABAN
Sample no. | Standard
(Area AU) |
Sample
(Area AU) |
1 | 138668 | 138650 |
2 | 138764 | 138758 |
3 | 138599 | 138596 |
Mean | 138677 | 138668 |
SD | 82.86736 | 82.48636 |
%RSD | 0.059756 | 0.059485 |
FIG. 17: APIXABAN LINEARITY CURVE, WHERE R2 = 0.9956
LOD and LOQ: The results of signal to noise ratio was compared with the response of Apixaban standard. The LOD and LOQ were found to be 0.2887 ppm and 0.8749 ppm, respectively.
System Suitability: System suitability tests were performed using Apixaban standard and test solutions to check for compliance with specified parameters in Table 13.
TABLE 12: SYSTEM SUITABLE STUDIES OF APIXABAN
Sample no. | Parameter | Apixaban |
1 | Retention time | 3.305 |
2 | Number of the theoretical plates | 5538 |
3 | Tailing factor | 0.85 |
4 | Peak area | 1134680 |
5 | Peak height | 190457 |
Robustness Studies: The changes were applied, and system suitability parameters were checked, found to be within the acceptable limits. It was noted that trivial changes in temperature and flow rate does not affect the method and produces results, which passes system suitability. Hence, the method was robust, as indicated in Table 14.
TABLE 13: ROBUSTNESS STUDIES OF APIXABAN
Factor | Level | TR | Tailing factor |
Parameter : flow rate | |||
1.2 ml/min | -1 | 3.53 | 0.86 |
1.3 ml/min | 0 | 3.31 | 0.92 |
1.4 ml/min | +1 | 3.20 | 0.74 |
Mean | 3.34 | 0.84 | |
Parameter :column temperature | |||
25 ºC | -5 | 3.48 | 0.75 |
30 ºC | 0 | 3.32 | 0.88 |
35 ºC | +5 | 314 | 0.81 |
Mean | 3.31 | 0.79 |
CONCLUSION: A new RP-HPLC method for the determination of apixaban was developed and validated. The developed method was accurate, precise, specific, and sensitive for the quantitative analysis of apixaban both in bulk drugs and tablets. Validation studies indicated the method to be robust on minor variations in the chromatographic parameters. No attempt was made to quantify the degradation of products in this research project.
This method can be used for routine quality monitoring of apixaban and its tablets. This method can be extended for application in LC-MS for quantitative estimation of known and unknown impurities generated during forced degradation and related substance studies both for API and finished products of apixaban.
ACKNOWLEDGEMENT: The authors wish to thank the University of Petra-Amman, Faculty of Pharmacy and Medical Sciences, and Lucian Blaga University-Sibiu, Faculty of Medicine, for supporting this work. The authors are also thankful to colleagues in the management of their cooperation in carrying out this work.
CONFLICTS OF INTEREST: The authors mention no conflict of interest.
REFERENCES:
- British national formulary: BNF 76 (76 ed.). Pharmaceutical Press 2018; 124-25.
- Skelley JW, Thomason AR and Hammond LN: Extended-duration use of direct oral anticoagulants to prevent VTE in Acutely III medical patients. Sr Care Pharm 2019; 34(2): 99-108.
- Powell J, Taylor J and Garland SG: Andexanetalfa: A Novel Factor Xa Inhibitor Reversal Agent. Ann Pharmacother 2019; 53(9): 940-46.
- Knotts TL and Mousa SA: Anticoagulation in Venous Thromboembolism Prophylaxis in Medically Ill Patients: Potential Impact of NOACs. Am J Cardiovasc Drugs 2019; 19(4): 365-76.
- Weitz JI and Chan NC: Long-Term Management of Venous Thromboembolism: Lessons from EINSTEIN CHOICE and Other Extension Trials. Thromb Haemost 2019; 119(5): 689-94.
- Young M: Atrial Fibrillation. Crit Care Nurs Clin North Am 2019; 31(1): 77-90.
- FDA Approves Eliquis (apixaban) for the Treatment of Deep Vein Thrombosis and Pulmonary Embolism. Article on https://www.drugs.com/newdrugs/fda-approves-eliquis-apixaban-deep-vein-thrombosis-pulmonary-embolism-4073.html.
- Ponnekanti K and Sundararajan R: Development and validation of new RP-UPLC method for the determination of cefdinir in bulk and dosage form. Int J Pharm Pharm Sci 2018; 10(1): 178.
- Haque A, Soundharya R, Venu J, Lakshmi M and Bakshi MV: Method development and validation of apixaban using RP-HPLC method and its stress stability study. Int J Chem Pharm Anal 2017; 5(1): 1-11.
- Katari VB and Jonnalagadda SB: Stability‐indicating RP‐HPLC method development and validation for determination of nine impurities in apixaban tablet dosage forms. Robustness study by quality by design approach. Biomed Chromatogr 2020; 34(1): e4719.
- Shalini B, Vandana A, Vijay B and Gupta MK: Ultra performance liquid chromatography: a revolutionized LC technique. Int J Drug Regul Aff 2014; 2(3): 83-87.
- Preti R: Core-Shell Columns in High-Performance Liquid Chromatography: Food Analysis Applications. Int J Anal Chem 2016; 3189724.
- Samanidou VF: Core-Shell Particle Technology in Pharmaceutical Analysis. Samanidou, Pharmaceut Anal Acta 2013; 4: 4.
- Zhang WL, Lou D, Zhang DT, Zhang Y and Huang HJ: Determination of rivaroxaban, apixaban and edoxaban in rat plasma by UPLC-MS/MS method. J Thromb Thrombolysis 2016; 42(2): 205-11.
- Abu-Dayyih W, Zakarya Z, Abu-Dayyih A and Al-Ani I: Reversed Phase HPLC for a Validation and Determination of Udenafil (Zydena®) and Dapoxetine (Priligy®) Simultaneously in Tablet Dosage Form in Jordanian Market. Der Pharma Chemica 2018; 10(4): 68-74.
- Shin‐Yi L, Ching‐Hua K, Shin‐Joe Y, Li‐Kai T, Yen‐Bin L, Chih‐Fen H, Sung‐Chun T and Jiann‐Shing J: Real‐world rivaroxaban and apixaban levels in asian patients with atrial fibrillation. Clin Pharmacol Ther 2020; 107(1): 278-86.
- Pursley J, Shen JX, Schuster A, Dang OT, Lehman J, Buonarati MH, Song Y, Aubry AF and Arnold ME: LC-MS/MS Determination of Apixaban (BMS-562247) and Its Major Metabolite in Human Plasma: An Application of Polarity Switching and Monolithic HPLC Column Bioanalysis 2014; 6(15): 2071-82.
- Gous T, Couchman L, Patel JP, Paradzai C, Arya R and Flanagan RJ: Measurement of the direct oral anticoagulants apixaban, dabigatran, edoxaban, and rivaroxaban in human plasma using turbulent flow liquid chromatography with high-resolution mass spectrometry. Ther Drug Monit 2014; 36(5): 597-605.
- Prabhune SS, Jaguste RS, Kondalkar PL and Pradhan NS: Stability-Indicating High Performance Liquid Chromatographic Determination of Apixaban in the Presence of Degradation Products. Sci Pharm 2014; 82(4): 777-85.
- The United States Pharmacopeia. 2015. Validation of Compendial Methods. USP38 NF-33, Chapter 1225.
- ICH Q2 (R1). 2005 Validation of Analytical Procedure: Text and Methodology. International Conference on Harmonization (ICH).
- Ruan J, Tattersall P, Lozano R and Shah P: The Role of Forced Degradation Studies in Stability Indicating HPLC Method Development. Am Pharm Rev 2006; 9: 46-53
- Bakshi M and Singh S: Development of Validated Stability-Indicating Assay Methods-Critical Review. J Pharm Biomed Anal 2002; 28(6): 1011-40.
- Carstensen JT and Rhodes CT: Drug Stability Principles and Practices. 3rd Edition, 2000, Marcel Dekker, New York.
How to cite this article:
Al-Ani I, Hamad M, Al-Shdefat R, Mansoor K, Gligor F and WA Dayyih: Development and validation of a stability indicating RP-HPLC method of apixaban in commercial dosage form. Int J Pharm Sci & Res 2021; 12(1): 241-51. doi: 10.13040/IJPSR.0975-8232.12(1).241-51.
All © 2013 are reserved by the International Journal of Pharmaceutical Sciences and Research. This Journal licensed under a Creative Commons Attribution-NonCommercial-ShareAlike 3.0 Unported License.
Article Information
24
241-251
1084
1566
English
IJPSR
I. Al-Ani, M. Hamad, R. Al-Shdefat, K. Mansoor, F. Gligor and W. A. Dayyih *
Department of Pharmaceutical Medicinal Chemistry and Pharmacognosy, Faculty of Pharmacy & Medical Sciences, University of Petra, Amman - Jordan.
wabudayyih@uop.edu.jo
29 December 2019
05 April 2020
10 October 2020
10.13040/IJPSR.0975-8232.12(1).241-51
01 January 2021