DEVELOPMENT AND VALIDATION OF HPLC METHOD FOR THE DETERMINATION OF ALVIMOPAN IN BULK AND PHARMACEUTICAL FORMULATION
HTML Full TextDEVELOPMENT AND VALIDATION OF HPLC METHOD FOR THE DETERMINATION OF ALVIMOPAN IN BULK AND PHARMACEUTICAL FORMULATION
Ramesh Devi
Government Polytechnic for Women, Warangal, Telangana, India.
ABSTRACT: Objective: The present investigation demonstrates a simple, sensitive and accurate high-pressure liquid chromatographic (HPLC) method for the determination of Alvimopan (AMP) in bulk and dosage form. The chromatographic separation was achieved within 10 minutes by using acetonitrile: potassium dihydrogen phosphate buffer pH 3.0 adjusted with orthophosphoric acid (50:50) as mobile phase on Altima Grace Smart C-18 column (5μ; 250 × 4.6 mm) at flow rate of 1.0 ml/min with injection volume 20µL at wavelength of 261nm. The retention time of the drug was found to be 5.02 min. This method was validated as per United States Food & Drug Administration (US-FDA) guidelines. The results of the validation parameters were found to be within the acceptance limits. The method was linear in the concentration range from 500-20000 ng/mL (r2 = 0.9998) and the accuracy of drug was found to be 95.83-104.63% for AMP. The lower limit of quantification was found to be ng/ml and the stability of recovered samples at different conditions was found to be more than 95%. The developed method possesses good selectivity, specificity, there was no interference found in the blank sample at retention times of AMP. We found a good correlation between the peak area and concentration of the drug under prescribed conditions. Furthermore, the method can also be used to estimation of drug in the dosage form for future studies of AMP.
Keywords: Alvimopan, Method development, Recovery, Stability, Validation
INTRODUCTION: Narcotic medicines that are often used to reduce the pain caused by gastrointestinal surgery. However, these medicines can cause a variety of side effects, such as nausea, vomiting, constipation and stomach pain, which are further leads to delay recovery in patients undergoing gastrointestinal surgery. Therefore, medicines those are required to prevent the above-mentioned side effects without diminishing the pain-relieving effect of narcotic medicines.
Alvimopan (AMP) (Entereg), the only drug approved by the Food & Drug Administration (FDA) for the treatment of postoperative ileus 1, 2. This drug behaves as a peripherally acting μ-opioid antagonist. Since, the AMP has limited ability to cross the blood–brain barrier, many of the undesirable side-effects of the narcotic medicines are minimized without affecting analgesia or precipitating withdrawal 3, 4.
In order to comprehend the antagonist effect of AMP further, it is imperative to determine the pharmacokinetic parameters through estimation of the AMP in plasma using simple estimation methods. Several liquid chromatography-tandem mass spectrometric (LC–MS/MS) assay methods have been employed for the determination of various drugs in the form of single and combined dosage forms 5-9. In addition, estimation of AMP in plasma and pharmaceutical formulation has also been carried out using LC-MS/MS method. In general, this method is highly sensitive to separate and identify a multitude of compounds in low concentration in a complex mixture with little assay optimization 10, 13.
The present contribution provides simple and regular estimation method for determination of AMP in using high pressure liquid chromatography (HPLC). Fig. 1 gives the molecular structure of AMP. As the above-mentioned, AMP is the only narcotic antagonist approved by the FDA for the treatment of postoperative surgery of gastro intestine.
FIG. 1: CHEMICAL STRUCTURE OF 2-([(2S)-2-([(3R, 4R) - 4 - (3 - HYDROXYPHENYL) - 3, 4-DIMETHYL –PIPERIDIN - 1 - YL] METHYL) - 3 - PHENYL-PROPANOYL] AMINO) ACETIC ACID
MATERIALS AND METHODS:
Instrumentation: The instruments employed in this study are; HPLC- Shimadzu HPLC SPD 10A VP UV-Visible Detector with manual injector. Sonicator- Sharp Analytical, Hyderabad, India. Analytical balance- Sartorius, German. Millipore Direct-Q 3 U.V. USA. pH meter- Systronics, Ahmadabad, India.
Standards and Chemicals: Alvimopan (AMP) was gift sample obtained from Aurobindo Pharma (Hyderabad, India). Purified water was obtained from a Millipore Direct-Q 3 U.V. Acetonitrile of HPLC grade, o-phosphoric acid and potassium dihydrogen phosphate were of A.R. Grade were purchased from Merck, Pvt. Ltd (Mumbai, India).
Stock and Working Solution Preparation:
Preparation of Standard Stock Solution: Accurately weighed and transferred 10mg of AMP into a 10mL volumetric flask dissolved with small amount of methanol and made up of the volume with methanol. Daily working standard solutions of mixture were prepared by suitable dilution of the stock solution with the mobile phase.
Preparation of a Buffer: Accurately weighed 2.72g of potassium dihydrogen orthophosphate dissolved in 1000ml of HPLC grade water and pH was adjusted to 3.0 by using orthophosphoric acid.
Chromatographic Conditions: The chromato-graphic separation was done by using Grace Smart Altima C18 column (250X4.6 mm, 5µ) with mobile phase acetonitrile: phosphate buffer pH 3.0 (50:50 % v/v) at a flow rate of 1mL/min and detection wavelength was 261nm with 20µL of injection volume.
Method Development:
Method Optimization: The method development, top priority was given for the sharp and separation of drug from solvent peaks. The chromatographic method was optimized by changing various parameters, such as pH of the mobile phase, organic solvent and buffer used in the mobile phase and composition of the mobile phase on trial error basis by varying one parameter and keeping all other conditions constant.
Method Validation: The validation parameters like linearity, sensitivity, accuracy, precision, recovery and stability of drugs were studied according to the ICH guidelines 14.
Selectivity: Selectivity was studied by comparing the chromatograms obtained from the blank sample with the chromatogram obtained from a standard drug mixture.
Linearity: The linearity of this method was evaluated by linear regression analysis, using least square method and found linear in the concentration range of 0.5-20µg/mL for AMP. Calibration standards are prepared by spiking required volume of working standard (100µg/mL) solution into different 10 mL volumetric flasks and volume made up with mobile phase to yield concentrations of 0.5, 1, 2, 5, 10, 15, 20 µg/mL of AMP. The resultant peak area of drug was measured. Calibration curve is plotted between peak areas of drug against concentration of the drug.
Sensitivity: The LOD and LOQ of this method were verified based on the standard deviation of response, slope.
Intra-day and Inter-day Precision and Accuracy, Recovery: Intra- and inter-day accuracy and precision of this method was determined at three different concentration levels in 3 different days. On each day, three replicates were analyzed with independently prepared calibration curves. The accuracy and precision were expressed as percentage accuracy and relative standard deviation (R.S.D., %) respectively.
Robustness: Robustness of the method was done by changing slight variation in the parameters like mobile phase composition, flow rate and wavelength. Present method didn’t show any significant change when the critical parameters were modified (i.e. Mobile phase composition, flow rate and pH of buffer).
Solution Stability: The stability of the drug solution was determined for the short-term stability, auto-sampler stability. Short-term stability was carried out by keeping at room temperature (25°C) for 24 h to determine any degradation during the analysis. Each sample injected three times into HPLC, and concentrations obtained were compared with the nominal values of the quality control (QC) samples.
Analysis of Marketed Formulation: 20 Capsules (Entereg) were weighed, finely powdered and an accurately weighed sample of powdered tablets equivalent to 12mg of AMP [equivalent to one capsule] was extracted with different extraction solvents like acetonitrile, methanol, water and mobile phase.
The powder equivalent one tablet was transferred and extracted with 50:50 methanol and acetonitrile in a 100ml volumetric flask and sonicated for 15 min. This solution was filtered through Whatmann No.1 filter paper. The solution obtained was diluted with the mobile phase so as to obtain a concentration in the range of linearity previously determined, filtered through 0.22µ syringe filter and injected into HPLC. The amount of drug recovered was calculated from the respective linear graph.
RESULTS AND DISCUSSION: During the method development, top priority was given for the complete separation of drugs. The chromatographic method was optimized by changing various parameters, such as pH of the mobile phase, organic solvent and buffer used in the mobile phase and composition of the mobile phase on trial error basis. Phosphate buffer in various strengths are tried along with methanol and acetonitrile as organic solvent.
A mixture of acetonitrile and phosphate buffer with different pH was tried. At pH 3.0 the separation was good enough, and then the proportions of acetonitrile and phosphate buffer pH 3.0 were tested as a mobile phase with Grace Smart C-18 column. The mobile phase composition of 50:50 v/v phosphate buffer: acetonitrile was shown good resolution, retention time with minimal tailing factor in acceptable range.
The method was optimized with the mobile phase composition of acetonitrile and phosphate buffer 50:50 (v/v). Buffer molarity of 10, 20 and 50 mM was tested. There were no significant changes in the chromatographic response and peak shape with change in buffer molarity. A buffer molarity of 20 mM was selected for further analysis.
After several trials, the method was optimized as a mixture of 20mM potassium dihydrogen phosphate buffer (pH 3.0) and acetonitrile (50:50 v/v), at a flow rate of 1mL/min, at 261nm by using Grace Smart, Altima C-18 column.
These chromatographic conditions achieved satisfactory resolution, retention time and tailing for AMP. The Fig. 2 shows that chromatogram of standard drug.
The standard mixture solution was used as a system suitability solution, and it was injected into HPLC five times. The retention time, tailing factor, resolution and theoretical plates of drug were observed. Then, calculated the percentage relative standard deviation (% RSD) of five consecutive injections for each parameter. The system’s suitability parameters of the present method were found to be within acceptable limits.
FIG. 2: STANDARD CHROMATOGRAM OF AMP
TABLE 1: SYSTEM SUITABILITY PARAMETERS OF AMP
Parameters (n=5) | Results | Required limits |
Retention time in minutes (Rt) | 5.02±0.01, 0.30 | RSD≤2 |
Theoretical plates (N) | 6433±99, 1.54 | N>2000 |
Tailing Factor (T) | 1.17±0.01, 1.48 | T≤2 |
Peak Area | 1125898±13464, 1.19 | - |
The data of system suitability parameters were represented in Table 1. The acceptable limits of the resolution between two adjacent peaks should be ≥2 and tailing factor should be ≤2 15 and the %RSD of these values should be ≤2. System suitability tests confirmed that the chromatographic system was adequate for the analysis planned to be done.
The method was found to be selective for AMP as there is no interference that was found at the retention time of drug (AMP) when comparing the chromatograms of blank (mobile phase) and drug solution (AMP). The linearity was performed, and calibration curve is plotted between peak areas of drug against concentration of the drug. The curve was linear over the range of 0.5-20µg/mL for AMP and the graph shown in Fig. 3.
FIG. 3: LINEARITY GRAPH OF AMP
The regression equations of drug was y=59.898x – 3437.3(r2=0.9998) for AMP. The results of intra and inter day precision was shown in Table 2. The % RSD was found to be less than 2 for all the drugs which indicating that the method is precise.
TABLE 2: INTRA AND INTER-DAY ACCURACY AND PRECISION OF AMP
Concentration (ng/mL) | Intra-day (n=6) | Inter-day (n=9) | ||
mean±SD | RSD (%) | mean±SD | RSD (%) | |
1500 | 104.63±0.55 | 0.53 | 104.58±1.20 | 1.15 |
8000 | 98.15±1.34 | 1.36 | 95.83±1.77 | 1.85 |
15000 | 100.19±0.40 | 0.40 | 99.35±0.18 | 0.19 |
Values expressed Mean±SD
Present method didn’t show any significant change when the critical parameters were modified. The tailing factor for the drug was always less than 2.0 and the components were well separated under all the changes carried out (i.e. Mobile phase composition, flow rate and pH of buffer). Considering the modifications in the system suitability parameters and the specificity of the method, as well as carrying the experiment at room temperature may indicate the method found to be robust. The robustness data of the method was presented in Table 3.
The stability of the drug was studied for short-term stability using the QC samples. The samples were analyzed and compared with freshly analyzed QC samples, no differences were found in accuracy and precision. The stability studies indicate there were no major changes observed in this study.
The method was applied for the determination of AMP in formulation (Entereg) and the amount of drug (AMP) was found to be 12.07mg, the accuracy was found to be 100.05%. The chromatogram of formulation was shown in Fig. 4 and the data was presented in Table 4.
TABLE 3: ROBUSTNESS DATA OF AMP
Parameter | Variation | System suitability | ||
Retention time | Theoretical plates | Peak Asymmetry | ||
Flow Rate (mL/min) |
0.9 | 5.04 | 6560 | 1.19 |
1.0 | 5.02 | 6433 | 1.17 | |
1.1 | 5.00 | 6469 | 1.20 | |
Mean±SD | 5.02±0.02, 0.39 | 6487±67, 1.00 | 1.20±0.04, 1.28 | |
% ACN
|
52 % v/v | 4.99 | 6498 | 1.14 |
50 % v/v | 5.02 | 6433 | 1.17 | |
48 % v/v | 5.03 | 6539 | 1.15 | |
Mean±SD | 5.01±0.02, 0.41 | 6490±53, 0.82 | 1.15±0.01, 1.32 | |
pH | 2.92 | 5.02 | 6589 | 1.17 |
3.00 | 5.02 | 6433 | 1.17 | |
3.11 | 5.04 | 6487 | 1.18 | |
Mean±SD | 5.02±0.01, 0.22 | 6503±79, 1.21 | 1.17±0.00, 0.49 |
FIG. 4: CHROMATOGRAM OF AMP FROM FORMULATION
TABLE 4: RECOVERY STUDY FROM FORMULATION (N=5) OF AMP
Brand Name | Labeled amount (mg) | Peak Area | Amount of drug (mg) | Percentage of drug recovery |
EnteregTM |
12 |
69808 | 12.22 | 101.89 |
68975 | 12.08 | 100.73 | ||
68124 | 11.95 | 99.55 | ||
68974 | 12.08 | 100.73 | ||
68457 | 12.00 | 100.01 | ||
Mean±SD | 12.07±0.10 | 100.58±0.88 |
Values are expressed in Mean ±SD, n=5
CONCLUSION: The developed method possesses good selectivity, specificity, there is no interference found in the blanks at retention time of AMP and good correlation between the peak area and concentration of the drug under prescribed conditions. The recovery studies are found to be 100.58% for the drug. The observation of % RSD less than 2 for both intra- and inter-day measurements indicates a high degree of precision. In the present method, a Grace Smart, Altima C-18 column has been used and at a flow rate of 1mL/min and pH of the buffer were 3.0 which are within the limits specified by the manufacturers. The method was optimized with low injection volume. The stability of AMP was found to be within the limits indicating that there is no degradation of drug during the daily analysis. This method was applied for the determination of AMP in tablet dosage form. From the results we are concluding that the developed method was accurate, precise, and robust. Hence the method can be applied for further investigation of AMP.
ACKNOWLEDGEMENT: Author is thankful to Aurabindo Pharma Ltd., Hyderabad, Telangana, India for providing drugs and to Adept Pharma and Bioscience Excellence Private Limited for providing all types of facilities for the research.
CONFLICT OF INTERESTS: The authors report no conflict of interest.
REFERENCES:
- Sharma A and Jamal MM: Opioid induced bowel disease: a twenty-first century physicians’ dilemma. Considering pathophysiology and treatment strategies”. Current Gastroenterology Reports 2013; 15(7): 334.
- FDA press release-FDA Approves Entereg to Help Restore Bowel Function Following Surgery. http://www.fda.gov/NewsEvents/Newsroom/PressAnnouncements/2008/ucm116899.htm.
- Alvimopan Product Label as approved by the FDA on May 20, 2008.
- Neary P and Delaney P: “Alvimopan”. Expert opinion on Investigational Drugs 2005; 14(4): 47988.
- Schmidt WK: “Alvimopan * (ADL 8-2698) is a novel peripheral opioid antagonist”. American Journal Surgery 2001; 182 (5): 27-38.
- Vrushali ST, Deodhar MN and Vijayalakshmi Prakya: LC and LC-MS study for simultaneous determination of tramadol hydrochloride and ketorolac tromethamine in bulk and formulation with major degradation products. Bulletin of Faculty of Pharmacy, Cairo University 2016; 54: 87-97.
- Essam Ez, Nisreen F Abo-Talib, Marwa HT and Abdelaaty AS: Development and validation of LC/MS/MS method for the determination of montelukast, gliclazide, and nifidipine and its application to a pharmacokinetic study. Chemistry of Central Journal 2014; 8: 17.
- Jaivik VS, Priyanka AS, Mallika S and Pranav SS: Fast and sensitive LC-MS/MS method for simultaneous determination of lisinopril and hydrochlorthiazide in human plasma. Journal of pharmaceutical Analysis 2017; 7: 163-169.
- Tejas D, Komal Ch and Priti M: Development of LC-MS/MS method for determination of iloperidone in rabbit plasma: Application to pharmacokinetic study. International Journal of Pharmacy and Pharmaceutical Sciences 2015; 7: 294-297.
- Ravi SO and H. Rush A: Strong EW Quantitative determination of Alvimopan in human plasma by LC-MS-MS. Worldwide Clinical Trials Drug Development Solutions-Bioanalytical Sciences, Austin, Texas, T2296.
- Padmini A, Sudharani U, Prasad G, Bennayyanaidu GA, Raju K and Rambabu K: A New L.C. method development and validation for analysis of Atvimopin in Formulations. Research Desk 2012; 1(2): 60-65.
- Majumdar Manish NR and Sunitha HB: "Development of newer validated HPLC method for the determination of Alvimopan in Rat plasma," SSRG International Journal of Pharmacy and Biomedical Engineering 2020; 3(7): 4-14.
- Ramesh D and Habibuddin M: Development and validation of RP-HPLC method for the determination of Alvimopan in rat plasma. International Journal of Pharmacy and Pharmaceutical Sciences 2018; 10(10): 124-129.
- International Conference on Harmonization (ICH). Validation of Analytical Methods: Methodology. ICH Q2 B., 1996.
- USP (The United States pharmacopoeia convention) 30-NF 25, 2007. Rockville MD. 1005, 1776.
How to cite this article:
Ramesh D: Development and validation of HPLC method for the determination of alvimopan in bulk and pharmaceutical formulation. Int J Pharm Sci & Res 2025; 16(6): 1645-50. doi: 10.13040/IJPSR.0975-8232.16(6).1645-50.