HPLC METHOD FOR ESTIMATION OF DRUG RELEASE OF ENTACAPONE IN ENTACAPONE, LEVODOPA AND CARBIDOPA TABLETS
HTML Full TextHPLC METHOD FOR ESTIMATION OF DRUG RELEASE OF ENTACAPONE IN ENTACAPONE, LEVODOPA AND CARBIDOPA TABLETS
Prasoon Bhatnagar*1, 2, Deepak Vyas 1, Shailendra Kumar Sinha 2 and Archana Gajbhiye 3
Department of Chemistry 1, Department of Biotechnology 3, Sir Padampat Singhania University, Udaipur - 313601, Rajasthan, India.
Macleods Pharmaceuticals Limited 2, Mumbai - 400059, Maharashtra, India.
ABSTRACT: Today’s generic drugs manufacturing industry need robust analytical methods for analysis of generic drug products to provide best quality pharmaceutical formulation. Entacapone, Levodopa and Carbidopa tablets is such a product which is used for the treatment of Parkinson’s disease. This paper presents a robust HPLC method that has been developed for estimation of % drug release of Entacapone in this multi component drug formulation. HPLC column used for separation was Cosmosil 5PE-MS 150mm x 4.6 mm with 5 micron particle size. Combination of phosphate buffer pH 2.5, acetonitrile and methanol was used for mobile phase in isocratic mode with UV detection at 280 nm. The method was validated to ensure suitability of the method for quantitative determination of % drug release of Entacapone in presence of Levodopa and Carbidopa in this triple drug combination pharmaceutical formulation. Proposed method was found to be specific, precise, robust, accurate and linear in range 0.066 to 0.33 mg/mL of Entacapone.
Keywords: |
Pharmaceutical formulation, HPLC method, Dissolution test, Multi component drug formulation, Parkinson disease
INTRODUCTION: Parkinson’s disease is a progressive, neurodegenerative disorder of the extrapyramidal nervous system affecting the mobility and control of the skeletal muscular system. Symptoms of Parkinson’s disease are related to depletion of dopamine. But administration of dopamine is ineffective in the treatment of Parkinson's disease. This is because it does not cross the blood-brain barrier. However, levodopa (Fig. 1A), the metabolic precursor of dopamine, does cross the blood-brain barrier, and presumably is converted to dopamine in the brain.
Carbidopa (Fig. 1B) inhibits the decarboxylation of peripheral levodopa, making more levodopa available for transport to the brain. There is no pharmacopoeial or literature reference of a suitable HPLC method to estimate % drug release of Entacapone (Fig. 1C) in presence of other two drugs in the proposed triple combination formulation. Proposed method was developed to cater to this need of pharmaceutical industry Entacapone is a selective and reversible inhibitor of catechol-O-methyltransferase (COMT). When entacapone is given in conjunction with levodopa and carbidopa, plasma levels of levodopa are greater and more sustained than after administration of levodopa and carbidopa alone1. There is no pharmacopoeial or literature reference of a suitable HPLC method to estimate % drug release of Entacapone in presence of other two drugs in the proposed triple combination formulation. Proposed method was developed to cater to this need of pharmaceutical industry.
Literature survey revealed few methods for individual or combination product analysis such as; Method for in-vitro release of drugs is also found but with longer run time and for single drug product 2, 3, 4, spectroscopic methods for simultan-eous estimation of levodopa and carbidopa 5, Estimation methods by liquid chromatography for levodopa and carbidopa have been reported using electrochemical detector 6 and fluorescence detector 7. Spectrophotometric determination of entacapone was reported in single drug product 8, 9, HPLC method in combination product was also reported 10, 11, USP monographs for determination of Levodopa and carbidopa for single and dual drug combination have also been reported 12-17.
In the present study, we propose a rapid HPLC method for estimation of % drug release of Entacapone [(2E)-2-cyano-3-(3, 4-dihydroxy-5-nitrophenyl) - N, N-diethyl-2-propenamide] in presence of Levodopa and carbidopa in triple drug combination product.
FIG. 1A: CHEMICAL STRUCTURES OF LEVODOPA
FIG. 1B: CHEMICAL STRUCTURES OF CARBIDOPA
FIG. 1C: CHEMICAL STRUCTURES OF ENTACAPONE
MATERIAL AND METHODS:
Reagents and Materials: All analytical reagent grade (AR Grade) reagents were used for method development purpose. Acetonitrile (Merck) was used for standard solution preparation. Orthophosphoric acid (Rankem) and Potassium dihydrogen orthophosphate (Merck) were used for mobile preparation. Milli-Q water (HPLC grade) was used for all solution preparations. Working standards of Entacapone, Levodopa and Carbidopa were obtained from Macleods Pharmaceuticals Limited, Mumbai, India.
Chromatographic System and Conditions: Development study was performed on Shimadzu HPLC, consisting of UV-Visible, photodiode array detector and a quaternary gradient pump. Sample loop in the system was of 100µl capacity. Cosmosil 5PE-MS 150 x 4.6 mm, 5 µ (Nacalai Tesque, USA) HPLC column was used for chromatographic separation. Mobile phase consisted of mixture of phosphate buffer, acetonitrile and methanol in isocratic mode. Buffer consisted of 10 mM potassium dihydrogen orthophosphate solution with pH adjusted to 2.5 using orthophosphoric acid. Flow rate was 1.0 mL/min and detection was carried out at 280 nm based on their wavelength maxima as per UV spectrum. Lab solutions software was used for data collection. For intermediate precision study, Shimadzu HPLC system with gradient pump, UV-visible detector and Chemstation software was used.
Solution Preparation:
Standard Preparation: About 45 mg of Entacapone was accurately weighed and transferred to a 200 mL volumetric flask. 10 mL of acetonitrile was added and sonicated to dissolve Entacapone completely. Further diluted to volume with dissolution medium and mixed.
Sample Preparation: 900 mL of dissolution medium was poured in each vessel. Sufficient time was allowed for the dissolution medium to equilibrate at 37 °C ±0.5 °C. Stirring element speed was adjusted to 125 rpm. One tablet was placed in each basket. The apparatus was lowered in the dissolution medium and started.
At the end of specified time, 10 mL aliquot was withdrawn from a zone midway between the surface of the dissolution medium and the top of the rotating basket using a sampling cannula with pre - filter attached to the end of it and filtered immediately through Whatman GF/C (25 mm) filter, discarding first 5 ml of the filtrate.
Method Validation: Once optimum separation conditions are achieved, method was validated to ensure its suitability and reliability for routine use in estimation of % drug release of active ingredient in a pharmaceutical formulation. Validation parameters adopted are as follows:
Specificity: Specificity for blank, placebo, Levodopa and Carbidopa was established by injecting blank solution, placebo solution, Levodopa standard solution, Carbidopa standard solution and Entacapone standard solution.
Solution Stability: Solution stability was evaluated by storing sample solution at 25 °C till 48 hrs.
Filter Compatibility: Sample solution was prepared by spiking Entacapone into placebo powder containing Levodopa and Carbidopa equivalent to one dosage unit. At filtration stage, solution was filtered through Whatman GF/C filter (25 mm). First 5.0 mL of filtrate was discarded. The filtrate was collected for further analysis. The unfiltered sample solution was centrifuged. Each of the above solutions thus obtained (filtered sample solution and centrifuged sample solution) were analysed as described in the methodology. The results were calculated.
Filter Saturation: Sample solution was prepared by spiking Entacapone into placebo powder containing Levodopa and Carbidopa equivalent to one dosage unit. At filtration stage, three filtrates were obtained using three separate Whatman GF/C filters (25 mm) by discarding 1 mL, 3 mL and 5 mL respectively. The filtrates were collected for further analysis.
Accuracy: Accuracy study was performed at 50%, 100% and 120 % of the target concentration of Entacapone. Recovery solutions were prepared by spiking Entacapone API to placebo powder containing levodopa and Carbidopa in dissolution medium.
Linearity: A series of solutions were prepared by quantitative dilutions of the stock solution of standard to obtain solutions as mentioned in the following table at 30% to 150 % of the target concentration of Entacapone. Each solution was injected and the peak area was recorded. Slope, Y-intercept and Correlation coefficient of the regression line were calculated.
Repeatability:
Precision: Precision test was carried out by spiking Entacapone into placebo with levodopa and Carbidopa powder equivalent to one dosage unit to obtain solutions at 100 % level of target concentration. Six sample preparations were prepared and injected. The mean and relative standard deviation of the % drug release was calculated.
Intermediate Precision: For intermediate precision, analysis was carried out by different analyst, on a different day, using a different HPLC and different dissolution apparatus. The absolute difference between the mean % release obtained in precision and intermediate precision was calculated.
Robustness: The Dissolution method was carried out as described in the methodology and by making the following alterations in the dissolution conditions. Changing the volume of dissolution medium (890 mL, 910 mL), Changing the pH of dissolution medium (pH = 5.3, pH = 5.7)
RESULTS AND DISCUSSION:
Preliminary Studies: Selection of Dissolution Medium: Dissolution medium was chosen based on USFDA recommendation of Phosphate Buffer pH 5.5 for Entacapone in Carbidopa, Levodopa and Entacapone Tablets.
Selection of Wavelength: Wavelength was selected based on absorbance maxima of Entacapone as per UV spectrum. 280 nm was found to be optimum (Fig. 2).
FIG. 2: UV ABSORPTION SPECTRA OF LEVODOPA, CARBIDOPA AND ENTACAPON
Selection of Mobile Phase: Due to non-polar nature of Entacapone, mobile phase was optimized with 50% of methanol and 50% buffer. Looking at the pH range of HPLC column, buffer pH 2.5 was evaluated and found to be optimum.
Selection of HPLC Column: Entacapone is non-polar in nature which makes it elute late on a octadecyl phase. In order to elute it early, a more polar phase was evaluated and selected for method development. Cosmosil 5PE-MS, 150 x 4.6 mm, 5 µ was the column of choice. Conventional Phenyl phase is polar in nature but do not last long at low pH due to its weak bonding. Cosmosil PE column has an ethyl group attached to phenyl group which makes this column a rugged stationary phase with better column life. 150 mm column was chosen to achieve a shorter run time.
Selection of HPLC Pump Mode: Mobile phase with higher ratio of solvent was optimized in isocratic mode for estimation of active ingredients with a flow rate of 1 ml/min and run time of 10 minutes.
Selection of Diluent: Due to its low solubility in dissolution medium, Entacapone was first dissolved in acetonitrile and then finally diluted with dissolution medium for standard preparation.
Method Validation:
Specificity: As shown in Table 1 and Fig. 3, No interference from blank, placebo, Levodopa and Carbidopa was observed at retention time of Entacapone peak.
TABLE 1: METHOD VALIDATION
Sr. No | Sample Details | Retention Time (min) |
1 | Blank | No peak detected |
2 | Placebo Solution | No peak detected |
3 | Levodopa | 1.79 |
4 | Carbidopa | 1.82 |
5 | Entacapone | 4.61 |
FIG. 3: CHROMATOGRAMS OF SPECIFICITY TO CONFIRM NO INTERFERENCE AT RETENTION TIME OF PEAK OF INTEREST
Solution Stability: The absolute difference between the % release of sample solution when stored for 48 hours at 25 °C and % release of initial was within the acceptance criteria of not more than 2 (Table 2).
TABLE 2: OBSERVATION OF SOLUTION STABILITY
Time (hours) |
Area |
% Release |
Absolute difference |
Initial | 6782464 | 100.7 | - |
24 | 6833824 | 99.4 | 1.3 |
48 | 6597937 | 99.1 | 1.6 |
The sample solution was found to be stable till 48 hours, when stored at 25°C.
Filter Compatibility: The absolute difference between the results obtained for filtered solution and centrifuged solution was calculated (Table 3).
TABLE 3: FILTER COMPATIBILITY RESULTS OF THE PROPOSED METHOD
Filter Type | Area | % Release | Absolute Difference |
Centrifuge | 6713359 | 99.6 | - |
Whatman GF/C filter |
6782464 |
100.7 |
1.1 |
Since the absolute difference between the results obtained for filtered sample solution and centrifuged sample solution was within acceptance criteria of NMT 2, Whatman GF/C (25 mm) filter is considered as suitable for sample filtration.
Filter Saturation: Each of the above solutions thus obtained were analysed as described in the methodology. The absolute difference between the results obtained for consecutive filtered solutions was calculated (Table 4).
TABLE 4: FILTER SATURATION RESULTS OF THE PROPOSED METHOD
Volume Discarded |
Area |
% Release |
Absolute Difference |
1 mL | 6727730 | 99.8 | - |
3 mL | 6777998 | 100.6 | 0.8 |
5 mL | 6790666 | 100.8 | 0.2 |
Since the absolute difference between the results obtained for two consecutive filtered solutions was within acceptance criteria of NMT 2, volume of 5 mL was considered as sufficient to saturate the filter.
Accuracy: The % recovery was within 95-105% (Table 5). Hence method is considered to be accurate.
TABLE 5: ACCURACY RESULTS OF THE PROPOSED METHOD
Level | API Spiked (mg) | Area | % Recovery |
50 % | 100.15 | 3308363 | 98.9 |
100.15 | 3307003 | 98.9 | |
100.15 | 3306039 | 98.8 | |
100% | 200.31 | 6724202 | 100.5 |
200.31 | 6701853 | 100.2 | |
200.31 | 6720652 | 100.5 | |
120% | 240.37 | 8020915 | 99.9 |
240.37 | 8060047 | 100.4 | |
240.37 | 8062883 | 100.4 | |
Mean % Recovery | 99.8 |
Linearity: Linearity plot (Fig. 4) and results (Table 6) shows that the correlation coefficient is within acceptance criteria of not less than 0.99. Hence the method is linear.
Repeatability: The absolute difference between the mean % release results obtained in precision (Table 7). and intermediate precision (Table 8) was within the acceptance criteria of not more than 5.0 Also difference between precision and intermediate precision was within 5% (Table 9).
TABLE 6: LINEARITY RESULTS OF THE PROPOSED METHOD
% Level | Concentration (ppm) |
Area |
30 | 67.4 | 2095012 |
50 | 112.3 | 3499256 |
80 | 179.73 | 5566098 |
100 | 224.7 | 6893248 |
120 | 269.6 | 8283081 |
150 | 336.9 | 10366642 |
Slope | 30592.4 | |
Y-Intercept | 46167.6 | |
Correlation coefficient | 1.000 |
FIG. 4: LINEARITY PLOT
Precision:
TABLE 7: PRECISION RESULTS OF THE PROPOSED METHOD
Area | % Release | |
Sample-1 | 6688850 | 100.4 |
Sample-2 | 6660375 | 100.0 |
Sample-3 | 6707099 | 100.7 |
Sample-4 | 6684719 | 100.4 |
Sample-5 | 6698032 | 100.6 |
Sample-6 | 6666779 | 100.1 |
Mean | 100.4 | |
% RSD | 0.27 |
Intermediate Precision:
TABLE 8: INTERMEDIATE PRECISION RESULTS OF THE PROPOSED METHOD
Area | % Release | |
Sample-1 | 6892912 | 98.8 |
Sample-2 | 6908731 | 99.0 |
Sample-3 | 7003650 | 100.4 |
Sample-4 | 6998750 | 100.3 |
Sample-5 | 7016739 | 100.6 |
Sample-6 | 7015409 | 100.6 |
Mean | 100.0 | |
% RSD | 0.82 |
TABLE 9: COMPARISON OF PRECISION AND INTERMEDIATE PRECISION RESULTS OF THE PROPOSED METHOD
Content | Mean % Release
in Precision |
Mean % Release in Intermediate Precision | Absolute difference |
Entacapone | 100.4 | 100.0 | 0.4 |
Hence, the method for estimation of % release is precise.
Robustness: The absolute difference in the results obtained under normal condition and robustness study of change in dissolution medium volume was within the acceptance criteria of NMT 10 (Table 10 and 11). System suitability parameters were also within acceptance limits (Table 12). The absolute difference in the results obtained under normal condition and robustness study of pH 5.7 of dissolution medium was within the acceptance criteria of NMT 10. The absolute difference in the results obtained under normal condition and robustness study of pH 5.3 of dissolution medium was not within the acceptance criteria of NMT 10.
TABLE 10: ROBUSTNESS RESULTS OF THE PROPOSED METHOD
% Release of Entacapone | |||||
Unit |
Unaltered |
Volume
890 mL |
Volume
910 mL |
pH of
medium 5.3 |
pH of
medium 5.7 |
1 | 95.2 | 97.3 | 91.7 | 82.3 | 101.7 |
2 | 96.7 | 97.4 | 92.0 | 82.9 | 99.7 |
3 | 94.9 | 95.4 | 90.5 | 83.3 | 99.9 |
4 | 97.7 | 96.2 | 92.1 | 83.3 | 101.8 |
5 | 95.9 | 96.1 | 91.4 | 84.6 | 98.9 |
6 | 95.4 | 96.9 | 92.4 | 83.2 | 99.3 |
Mean | 96.0 | 96.6 | 91.7 | 83.3 | 100.2 |
% RSD | 1.10 | 0.81 | 0.73 | 0.91 | 1.23 |
TABLE 11: COMPARATIVE % RELEASE RESULTS FOR ROBUSTNESS
Sr. No. | Changed Parameter | Mean % Release | Absolute Difference |
1 | Unaltered | 96.0 | - |
2 | Volume 890 mL | 96.6 | 0.6 |
3 | Volume 910 mL | 91.7 | 4.3 |
4 | pH 5.3 | 83.3 | 12.7 |
5 | pH 5.7 | 100.2 | 4.2 |
The system suitability parameters were not significantly changed with altered conditions.
TABLE 12: SYSTEM SUITABILITY RESULTS FOR ROBUSTNESS
Sr.
No. |
Changed Parameter |
Tailing
Factor (NMT2.0) |
Theoretical
Plates (NLT 2000) |
% RSD for Entacapone peak area (NMT 2.0) |
1 | Unaltered | 0.99 | 8934 | 0.08 |
2 | Volume 890 mL | 1.02 | 9072 | 0.03 |
3 | Volume 910 mL | 1.00 | 9050 | 0.06 |
4 | pH 5.3 | 1.00 | 9124 | 0.08 |
5 | pH 5.7 | 1.00 | 9111 | 0.02 |
This concludes that pH of dissolution medium is a critical parameter.
CONCLUSION: A simple and efficient HPLC method for estimation of % drug release of Entacapone in triple drug combination product was developed and validated for specificity, accuracy, linearity, precision and robustness ensuring suitability of the method for quantitative analysis. The results indicated that this method is suitable for estimation of % Release of Entacapone in a pharmaceutical formulation.
ACKNOWLEDGEMENT: The authors express sincere gratitude to the Research and Development Center of Macleods Pharmaceuticals Limited, Mumbai, India, for granting permission to publish the data generated at Analytical Research and Development section for this research work. The first author is also thankful to his colleagues for their consistent support to carry out this research.
CONFLICT OF INTEREST: The authors declare that they have no conflict of interests.
REFERENCES:
- Stalevo (carbidopa, levodopa and entacapone) tablets, prescribing information, Novartis pharmaceuticals limited.
- Ashish S, Doshi, kalpeshJ, Upadhyay, Tushar NM, and Nanda N: Development and application of a high- performance liquid chromatographic method for the determination of in vitro drug release of levodopa, Carbidopa, and Entacapone from a tablet formulation. Journal of AOAC International 2009; 92(2): 394-403.
- Clésio S. Paim, Magda T. Martins, Marcelo D. Malesuik, Martin Steppe LC: Determination of entacapone in entacapone tablets; in-vitro dissolution studies. Journal of Chromatographic Science 2010; 48: 755-759.
- Prafullachandra T, Mhatre Vinayak S, Pai NR, Maurya C, Smrut T: Estimation of Entacapone tablets by reverse phase high performance liquid chromatographic method. Bioscience Discovery 2011; 2(3): 294-298.
- Madrakian T, Afkhami A, Borazjani M and Bahram M: Simultaneous derivative spectrophotometric determination of levodopa and Carbidopa in pharmaceutical preparations. Bull. Korean Chem. Soc. 2009; 25(12): 1764-1768.
- Bugamelli F, Marcheselli C, Barba E and Raggi MA: Determination of l-dopa, carbidopa, 3-O-methyldopa and entacapone in human plasma by HPLC-ED. Journal of Pharmaceutical and Biomedical Analysis 2011; 54(3): 562-567.
- Raut PP and Charde SY: Simultaneous estimation of levodopa and carbidopa by RP-HPLC using a fluorescence detector: its application to a pharmaceutical dosage form. Luminescence 2014; 29(7):762-771.
- Gadallah MN and Mohamed MS: Determination of drug Entacapone. Journal of the Chilean Chemical Society 2010; 55(1): 85-89.
- Mohamed Rizk, Ali K. Attia, Mona S. Elshahed and Amir SF: Validated voltammetric method for the determination of antiparkinsonism drug entacapone in bulk, pharmaceutical formulation and human plasma. Journal of Electroanalytical Chemistry 2015; 743(112-119).
- Bhatnagar P, Vyas D, Sinha SK and Chakrabarti T: Stability Indicating HPLC Method for Simultaneous Estimation of Entacapone, Levodopa and Carbidopa in Pharmaceutical Formulation”, Journal of Chromatography and Separation Techniques 2015; 9(6): 304.
- Rama Krishna VR, Murali Krishna KB, Hari Babu B: Development and validation of liquid chromatographic method for the simultaneous estimation of Levodopa, Carbidopa and Entacapone in the combined dosage form. Journal of Pharmacy Research 2014; 8(3): 281-288.
- Monograph for levodopa and carbidopa orally disintegrating tablets, USP 37 NF 2014; 32: 2134-2136.
- Monograph for levodopa and carbidopa extended release tablets, USP 37 NF 2014; 32: 2129-2134.
- Monograph for levodopa and Carbidopa tablets, British pharmacopoeia 2014; 3: 357-358.
- Monograph for levodopa and Carbidopa tablets, Indian pharmacopoeia 2014; 2:2081-2084.
- Monograph for Entacapone tablets, USP 37 NF 2014; 32: 2801-2802.
- Monograph for levodopa tablets, USP 37 NF 2014; 32: 3535-3536.
How to cite this article:
Bhatnagar P, Vyas D, Sinha SK and Gajbhiye A: HPLC method for estimation of drug release of entacapone in entacapone, levodopa and carbidopa tablets. Int J Pharm Sci Res 2017; 8(4): 1734-43.doi: 10.13040/IJPSR.0975-8232.8(4).1734-43.
All © 2013 are reserved by International Journal of Pharmaceutical Sciences and Research. This Journal licensed under a Creative Commons Attribution-NonCommercial-ShareAlike 3.0 Unported License.
Article Information
29
1734-1743
618
1224
English
IJPSR
Prasoon Bhatnagar*, Deepak Vyas, Shailendra Kumar Sinha and Archana Gajbhiye
Research Scholar, Department of Chemistry, Sir Padampat Singhania University, Bhatewar, Udaipur, Rajasthan, India.
prasoon.bhatnagar@spsu.ac.in
15 August, 2016
07 December, 2016
24 February, 2017
10.13040/IJPSR.0975-8232.8(4). 1734-43
01 April, 2017