HPLC METHOD FOR SIMULTANEOUS ESTIMATION OF DRUG RELEASE OF LEVODOPA AND CARBIDOPA IN ENTACAPONE, LEVODOPA AND CARBIDOPA TABLETS
HTML Full TextHPLC METHOD FOR SIMULTANEOUS ESTIMATION OF DRUG RELEASE OF LEVODOPA AND CARBIDOPA IN ENTACAPONE, LEVODOPA AND CARBIDOPA TABLETS
Prasoon Bhatnagar * 1, 2, Deepak Vyas 1, Shailendra Kumar Sinha 2 and Archana Gajbhiye 3
Department of Chemistry 1, Department of Biotechnology 3, Sir Padampat Singhania University, Udaipur, Rajasthan, -313601, India
Macleods Pharmaceuticals Limited 2, Mumbai, Maharashtra, India.
ABSTRACT: In pharmaceutical industry, researchers aim at catering to the need of robust analytical methods for analysis of generic drug products. The paper deals with method of analysis of pharmaceutical formulation - Entacapone, Levodopa and Carbidopa tablets for the treatment of Parkinson’s disease. The paper presents a simple and efficient HPLC method that has been developed for a multi component drug formulation for estimation of % drug release of levodopa and carbidopa. This HPLC method uses ‘Cosmosil 5PE-MS 150 x 4.6 mm, 5 µ’ HPLC column, combination of phosphate buffer pH 2.5 and methanol as mobile phase in gradient mode with UV detection at 280 nm. The method was validated and found to be precise, robust, accurate, linear (in range 0.020 to 0.40 mg/mL and 0.005 to 0.100 mg/mL of Levodopa and Carbidopa respectively), and specific for blank and placebo solution ensuring suitability of the method for quantitative determination of % drug release of Levodopa and Carbidopa in presence of Entacapone in multi component pharmaceutical formulation.
Keywords: |
Pharmaceutical formulation, HPLC, Dissolution test, Multi component drug formulation, Parkinson’s disease
INTRODUCTION: Parkinson’s disease is a progressive, neurodegenerative disorder of the extrapyramidal nervous system affecting the mobility and control of the skeletal muscular system. This is caused due to depletion of dopamine. Administration of dopamine is ineffective for treatment of Parkinson's disease because it does not cross the blood-brain barrier. Levodopa, the metabolic precursor of dopamine, has the ability to cross the blood-brain barrier, and converts into dopamine in the brain.
Carbidopa inhibits the decarboxylation of peripheral levodopa, making more levodopa available for transport to the brain. Entacapone is a selective and reversible inhibitor of catechol-O-methyltransferase (COMT). Plasma levels of levodopa are greater and more sustained when entacapone is given in conjunction with levodopa and carbidopa.
No literature reference of a suitable chromatographic method for estimation of drug release of levodopa and carbidopa in presence of Entacapone in triple drug combination product. Proposed method caters to this need of pharmaceutical industry for simultaneous estimation of levodopa and carbidopa.
Few spectroscopic/chromatographic methods are available for individual and or combination product for estimation of levodopa and carbidopa, spectroscopic methods for simultaneous estimation of levodopa and carbidopa 4. Simultaneous estimation of levodopa and carbidopa by UV spectroscopy 2, Estimation methods by liquid chromatography for levodopa and carbidopa have been reported using ECD 3 and FLD 4, LC method were reported for in-vitro release in single drug product 5, 6, 7, HPLC method in combination product was also reported 8-11, USP monographs for determination of Levodopa and carbidopa for single and dual drug combination have also been reported 12, 13, 14, 15, 16, 17 single and dual drug.
In the present research work, we propose a simple and robust HPLC method for simultaneous estimation of Levodopa {(2S)-2-amino-3-(3,4-dihydroxyphenyl) propanoic acid}[LD] and Carbidopa {(2S)-3-(3,4-dihydroxyphenyl)-2-hydra zino-2-methylpropanoic acid} [CD] in presence of Entacapone {(2E)-2-cyano-3-(3,4-dihydroxy-5-nitrophenyl)-N,N-diethyl-2-propenamide} [EN]. (Fig. 1)
FIG.1 CHEMICAL STRUCTURES OF LEVODOPA, CARBIDOPA AND ENTACAPONE
MATERIAL AND METHODS:
2.1. Reagents and Materials: All analytical reagent grade (AR Grade) reagents were used for method development purpose. Acetonitrile (Merck) and tetrahydrofuran (Merck) were used for standard solution preparation. Orthophosphoric acid (Rankem) and potassium dihydrogen orthophosphate (Merck) were used for mobile preparation. Milli-Q water (HPLC grade) was used for all solution preparations. Working standards of entacapone, levodopa and carbidopa were obtained from Macleods Pharmaceuticals Limited, Mumbai, India.
2.2 Spectrophotometric analysis for λmax: Solutions of levodopa, carbidopa and entacapone were prepared in diluent containing orthophosphoric acid and tetrahydrofuran and the absorption spectrum was obtained in the UV range using UV-Vis Spectrophotometer (Make). λmax was
determined from the respective absorption spectrum of the drugs. Both the drugs show λmax at about 280nm.
2.3. Chromatographic System and Conditions:
Development study was performed on Shimadzu HPLC, consisting of UV-Visible, photodiode array detector and a quaternary gradient pump. Sample loop in the system was of 100µl capacity. Cosmosil 5PE-MS 150 x 4.6 mm, 5 µ (Nacalai Tesque, USA) HPLC column was used for chromatographic separation. Phosphate buffer and methanol were used as mobile phase in gradient mode. Buffer was composed of 10 mM potassium dihydrogen orthophosphate solution with pH adjusted to 2.5 using orthophosphoric acid. Flow rate was 1.5 mL/min and detection was carried out at 280 nm based on their wavelength maxima as per UV spectrum. Labsolutions software was used for data collection.
2.4. Solution Preparation:
2.4.1 Standard Preparation:
Levodopa standard solution: About 42 mg of Levodopa was accurately weighed and transferred to a 25 mL volumetric flask. About 15 mL of diluent was added and sonicated to dissolve Levodopa completely. This solution was allowed to equilibrate to room temperature, diluted to volume with diluent and mixed.
Carbidopa standard solution: About 45 mg of Carbidopa was accurately weighed and transferred to a 100 mL volumetric flask. About 70 mL of diluent was added and sonicated to dissolve Carbidopa completely. This solution was allowed to equilibrate to room temperature, diluted to volume with diluent and mixed.
Standard solution: 5 mL of each of Levodopa standard solution and Carbidopa standard solution was transferred to a 50 mL volumetric flask, diluted to volume with dissolution medium and mixed.
2.4.2 Sample Preparation: 750 mL of dissolution medium was poured in each vessel. Sufficient time was allowed for the dissolution medium to equilibrate at 37°C + 0.5°C. Stirring element speed was adjusted to 50 rpm. One tablet was placed in each basket. The apparatus was lowered in the dissolution medium and started.
At the end of specified time, 10 mL aliquot was withdrawn from a zone midway between the surface of the dissolution medium and the top of the rotating basket using a sampling cannula with pre - filter attached to the end of it and filtered immediately through Whatman GF/C (25 mm) filter, discarding first 5 ml of the filtrate.
2.5. Method Validation: Once optimum separation conditions are achieved, method was validated to ensure its suitability and reliability for routine use in estimation of % release of active ingredients by HPLC in a pharmaceutical formulation. Validation parameters adopted are as follows:
2.5.1 Specificity: Specificity for blank, placebo and Entacapone was established by injecting blank solution, placebo solution, Levodopa standard solution, Carbidopa standard solution and Entacapone standard solution. (Fig. 3)
2.5.2 Solution Stability: Solution stability was evaluated by storing sample solution at 10°C till 24 hrs.
2.5.3 Filter Compatibility: Sample solution was prepared by spiking Levodopa and Carbidopa into placebo powder containing Entacapone equivalent to one dosage unit. At filtration stage, solution was filtered through Whatman GF/C filter (25 mm). First 5.0 mL of filtrate was discarded. The filtrate was collected for further analysis. The unfiltered sample solution was centrifuged.
Each of the above solutions thus obtained (filtered sample solution and centrifuged sample solution) were analysed as described in the methodology. The results were calculated.
2.5.4 Filter Saturation: Sample solution was prepared by spiking Levodopa and Carbidopa into placebo powder containing Entacapone equivalent to one dosage unit. At filtration stage, three filtrates were obtained using three separate Whatman GF/C filters (25 mm) by discarding 1 mL, 3 mL and 5 mL respectively. The filtrates were collected for further analysis.
2.5.5 Accuracy: Accuracy study was performed from 12.5% to 120% of the target concentration of individual active ingredient. Recovery solutions were prepared by spiking levodopa and Carbidopa to placebo powder containing Entacapone in dissolution medium. (Table 6).
2.5.6 Linearity: A series of solutions were prepared by quantitative dilutions of the stock solution of standard solutions to obtain solutions as mentioned in the following table from 8% to 150% of the target concentration of individual active ingredient. Each solution was injected and the peak area was recorded. Slope, Y-intercept and Correlation coefficient of the regression line were calculated. (Table 7)
2.5.7 Repeatability:
2.5.7.1 Precision: Precision test was carried out by spiking levodopa and Carbidopa to placebo powder containing Entacapone equivalent to one dosage unit to obtain solutions at 100 % level of target concentration.
Six sample preparations were prepared and injected. The mean and relative standard deviation of the results was calculated. The results obtained for % release are tabulated in Table 8.
2.5.7.2 Intermediate precision: For intermediate precision, analysis was carried out by different analyst, on a different day, using a different HPLC and different dissolution apparatus. The absolute difference between the mean % release results obtained in precision and intermediate precision was calculated. (Table10)
2.5.8 Robustness: The Dissolution method was carried out as described in the methodology and by making the following alterations in the dissolution conditions.
Changing the volume of dissolution medium (742.5 mL, 757.5 mL
Changing the strength of dissolution medium (0.08N, 0.12N)
RESULTS AND DISCUSSION:
3.1. Preliminary studies:
3.1.1 Selection of dissolution medium: Dissolution medium was chosen based on USFDA recommendation of 0.1 N Hydrochloric acid for Carbidopa and Levodopa in Carbidopa, Levodopa and Entacapone tablets.
3.1.2 Selection of wavelength: Wavelength was selected based on absorbance maxima of both the drugs as per UV spectrum. 280 nm was optimum for both the active ingredients. (Fig. 2)
FIG. 2: UV ABSORPTION SPECTRA OF LEVODOPA, CARBIDOPA AND ENTACAPONE.
3.1.3 Selection of mobile phase: Due to difference in acidity of levodopa/carbidopa and entacapone, low pH was selected to achieve optimum separation of all the peaks. With reference to the specified pH range of HPLC column, pH 2.5 was evaluated and found to be optimum.
3.1.4 Selection of HPLC column: Levodopa and Carbidopa elute early on an octadecyl phase. In order to retain them, a more polar phase was evaluated and selected for method development. Cosmosil 5PE-MS 150 x 4.6 mm, 5 µ was the column of choice.
Conventional Phenyl phase is polar in nature but do not last long at low pH due to its weak bonding. Cosmosil PE column has an ethyl group attached to phenyl group which makes this column a rugged stationary phase with better column life. 150 mm column was chosen to achieve a shorter run time.
3.1.5 Selection of HPLC pump mode: Entacapone do not elute early with a low solvent mobile phase. Hence, gradient mode was chosen and optimized to elute Entacapone in the same run and for separation of active ingredients with a flow rate of 1.5ml/min and run time of 9 minutes. (Table 1)
TABLE 1: GRADIENT TIME PROGRAM
Time
(min) |
Buffer
(% v/v) |
Methanol (% v/v) |
0 → 3 | 100 | 0 |
3 → 3.1 | 100 → 10 | 0 → 90 |
3.1 → 6 | 10 | 90 |
6 → 6.1 | 10 → 100 | 90 → 0 |
6.1→ 9 | 100 | 0 |
3.1.6 Selection of diluent: For better solubility and stability of Levodopa and Carbidopa, combination of Orthophosphoric acid and Tetrahydrofuran in ratio 70:30 was chosen as diluent.
3.2. Method Validation:
3.2.1 Specificity:
TABLE 2: VALUES OF RETENTION TIME OBTAINED
Sr. No. | Sample Details | Retention Time (min) |
1 | Blank | No interference observed |
2 | Placebo Solution | No interference observed |
3 | Levodopa | 2.25 |
4 | Carbidopa | 4.24 |
5 | Entacapone | 5.83 |
As shown in Table 2, No interference from blank, placebo and Entacapone was observed at retention times of Levodopa peak and Carbidopa peak.
CHROMATOGRAM OF BLANK
CHROMATOGRAM OF PLACEBO WITH ENTACAPONE
CHROMATOGRAM OF STANDARD
CHROMATOGRAM OF ENTACAPONE STANDARD
CHROMATOGRAM OF CARBIDOPA STANDARD
CHROMATOGRAM OF LEVODOPA STANDARD
FIG. 3: CHROMATOGRAMS FOR SPECIFICITY TO CONFIRM NO INTERFERENCE AT RETENTION TIME OF PEAKS OF INTEREST
3.3.2 Solution Stability: The absolute difference between the % release of sample solution when stored for 24 hours at 10°C and % release of initial
was within the acceptance criteria of not more than 2. (Table 3)
TABLE 3: OBSERVATION OF SOLUTION STABILITY
Time (hours) | Levodopa | Carbidopa | ||||
Area | % Release | Absolute difference | Area | % Release | Absolute difference | |
0 | 383262 | 101.9 | - | 80233 | 96.1 | - |
12 | 382619 | 101.8 | 0.1 | 79647 | 95.4 | 0.7 |
24 | 383091 | 100.2 | 1.7 | 78956 | 94.6 | 1.5 |
The sample solution was found to be stable till 24 hours, when stored at 10°C.
3.3.3 Filter Compatibility: The absolute difference between the results obtained for filtered
solution and centrifuged solution was calculated. (Table 4)
TABLE 4: FILTER COMPATIBILITY RESULTS OF THE PROPOSED METHOD
Filter Type | Levodopa | Carbidopa | ||||
Area | % Release | Absolute Difference | Area | % Release | Absolute Difference | |
Centrifuge | 364625 | 99.7 | - | 80808 | 101.3 | - |
Whatman GF/C filter | 366618 | 100.3 | 0.6 | 81157 | 101.7 | 0.4 |
Since the absolute difference between the results obtained for filtered sample solution and centrifuged sample solution was within acceptance criteria of NMT 2, Whatman GF/C (25 mm) filter is considered as suitable for sample filtration.
3.3.4 Filter Saturation:
Each of the filtered solutions thus obtained were analysed as described in the methodology. The absolute difference between the results obtained for consecutive filtered solutions was calculated. (Table 5).
TABLE 5: FILTER SATURATION RESULTS OF THE PROPOSED METHOD
Volume Discarded | Levodopa | Carbidopa | ||||
Area | % Release | Absolute Difference | Area | % Release | Absolute Difference | |
1 mL | 365740 | 100.0 | - | 81084 | 101.7 | - |
3 mL | 366187 | 100.2 | 0.2 | 81108 | 101.7 | 0.0 |
5 mL | 366624 | 100.3 | 0.1 | 80921 | 101.5 | 0.2 |
Since the absolute difference between the results obtained for two consecutive filtered solutions was within acceptance criteria of NMT 2, volume of 5 mL was considered as sufficient to saturate the filter.
3.3.5 Accuracy:
TABLE 6: ACCURACY RESULTS OF THE PROPOSED METHOD
Levodopa | Carbidopa | |||||
Level | API Spiked (mg) | Area | % Recovery | API Spiked (mg) | Area | % Recovery |
12.5 % | 25.2116 | 193570 | 100.1 | 6.251 | 44175 | 102.4 |
25.4509 | 193592 | 99.2 | 6.251 | 44055 | 103.1 | |
25.6702 | 192973 | 98.0 | 6.251 | 44000 | 103.0 | |
100% | 201.8224 | 1526569 | 98.6 | 50.008 | 347567 | 101.7 |
201.5233 | 1526541 | 98.8 | 50.008 | 347031 | 101.5 | |
201.4934 | 1525994 | 98.8 | 50.008 | 346880 | 101.5 | |
120% | 242.2168 | 1802524 | 97.1 | 60.010 | 413266 | 100.8 |
241.8978 | 1802568 | 97.2 | 60.010 | 413652 | 100.9 | |
242.0074 | 1800264 | 97.0 | 60.010 | 412909 | 100.7 | |
Mean % Recovery | 98.3 | Mean % Recovery | 101.8 |
The % recovery was within 95-105% (Table 6). Hence method is considered to be accurate.
3.3.6 Linearity:
TABLE 7: LINEARITY RESULTS OF THE PROPOSED METHOD
% Level | Levodopa | Carbidopa | ||
Concentration (ppm) | Area | Concentration (ppm) | Area | |
8 | 20.8 | 115019 | 5.06 | 25219 |
50 | 133.7 | 735280 | 33.41 | 162512 |
80 | 213.92 | 1168438 | 53.66 | 257718 |
100 | 267.4 | 1465216 | 66.82 | 321435 |
120 | 320.88 | 1736532 | 80.99 | 393997 |
150 | 401.11 | 2179072 | 100.23 | 478962 |
Slope | 5417.001 | 4790.148 | ||
Y-Intercept | 7383.111 | 1729.732 | ||
Correlation coefficient | 1 | 1 |
FIG. 4: LINEARITY PLOT
Linearity plot of Levodopa and Carbidopa (Fig. 4) and results (Table 7) shows that the correlation coefficient is within acceptance criteria of not less than 0.99. Hence the method is linear.
3.3.7 Repeatability:
3.3.7.1 Precision:
TABLE 8: PRECISION RESULTS OF THE PROPOSED METHOD
Levodopa | Carbidopa | |||
Area | % Release | Area | % Release | |
Sample-1 | 1443722 | 98.1 | 298726 | 95.1 |
Sample-2 | 1469334 | 99.9 | 297066 | 94.6 |
Sample-3 | 1461117 | 99.3 | 298870 | 95.1 |
Sample-4 | 1444689 | 98.2 | 298671 | 95.1 |
Sample-5 | 1471629 | 100.0 | 296375 | 94.3 |
Sample-6 | 1461307 | 99.3 | 298047 | 94.9 |
Mean | 99.1 | Mean | 94.9 | |
% RSD | 0.82 | % RSD | 0.35 |
3.3.7.2 Intermediate precision:
TABLE 9: INTERMEDIATE PRECISION RESULTS OF THE PROPOSED METHOD
Levodopa | Carbidopa | |||
Area | % Release | Area | % Release | |
Sample-1 | 1510380 | 100.4 | 325657 | 97.5 |
Sample-2 | 1533018 | 101.9 | 325950 | 97.6 |
Sample-3 | 1510459 | 100.4 | 325677 | 97.5 |
Sample-4 | 1532532 | 101.9 | 325973 | 97.6 |
Sample-5 | 1512934 | 100.6 | 326026 | 97.6 |
Sample-6 | 1532413 | 101.9 | 325942 | 97.6 |
Mean | 101.2 | Mean | 97.6 | |
% RSD | 0.78 | % RSD | 0.05 |
TABLE 10: COMPARISON OF PRECISION AND INTERMEDIATE PRECISION RESULTS OF THE PROPOSED METHOD
Content | Mean % Release
in Precision |
Mean % Release
in Intermediate Precision |
Absolute difference |
Levodopa | 99.1 | 101.2 | 2.1 |
Carbidopa | 94.9 | 97.6 | 2.7 |
The absolute difference between the mean % release results obtained in precision (Table 8) and intermediate precision (Table 9) was within the acceptance criteria of not more than 5.0 Also difference between precision and intermediate precision was within 5% (Table 10). Hence, the method for estimation of % release is precise.
3.3.8 Robustness:
TABLE 11: ROBUSTNESS RESULTS OF THE PROPOSED METHOD
Unit | % Release of Levodopa | ||||
Unaltered | Dissolution Volume
742.5 mL |
Dissolution Volume
757.5 mL |
Strength of Dissolution medium 0.08N | Strength of Dissolution medium 0.12N | |
1 | 103.8 | 97.3 | 100.6 | 96.8 | 89.9 |
2 | 98.1 | 102.1 | 99.1 | 102.1 | 97.7 |
3 | 98.0 | 98.0 | 97.0 | 92.1 | 91.3 |
4 | 91.0 | 104.1 | 100.3 | 91.6 | 101.6 |
5 | 103.1 | 100.6 | 97.4 | 92.0 | 96.6 |
6 | 100.8 | 99.4 | 101.1 | 101.0 | 98.4 |
Mean | 99.1 | 100.3 | 99.3 | 95.9 | 95.9 |
% RSD | 4.7 | 2.5 | 1.7 | 5.0 | 4.6 |
Unit | % Release of Carbidopa | ||||
Unaltered | Dissolution Volume
742.5 mL |
Dissolution Volume
757.5 mL |
Strength of Dissolution medium 0.08N | Strength of Dissolution medium 0.12N | |
1 | 104.5 | 97.2 | 99.3 | 94.4 | 88.7 |
2 | 97.8 | 102.0 | 99.0 | 99.3 | 95.5 |
3 | 97.6 | 99.4 | 96.2 | 91.7 | 88.5 |
4 | 91.0 | 103.7 | 100.6 | 91.6 | 98.3 |
5 | 102.4 | 100.6 | 96.7 | 89.2 | 97.3 |
6 | 101.4 | 99.7 | 101.6 | 98.0 | 95.8 |
Mean | 99.1 | 100.4 | 98.9 | 94.0 | 94.0 |
% RSD | 4.8 | 2.2 | 2.2 | 4.2 | 4.6 |
TABLE 12: COMPARATIVE % RELEASE RESULTS FOR ROBUSTNESS
Sr. No. | Changed Parameter | Levodopa | Carbidopa | ||
Mean % Release | Absolute Difference | Mean % Release | Absolute Difference | ||
1 | Unaltered | 99.1 | 99.1 | - | |
2 | Volume 742.5 mL | 100.3 | 1.2 | 100.4 | 1.3 |
3 | Volume 757.5mL | 99.3 | 0.2 | 98.9 | 0.2 |
4 | 0.08N | 95.9 | 3.2 | 94 | 5.1 |
5 | 0.12N | 95.9 | 3.2 | 94 | 5.1 |
TABLE 13: SYSTEM SUITABILITY RESULTS FOR ROBUSTNESS
Sr.
No. |
Changed Parameter | Levodopa | Carbidopa | ||||
Tailing
Factor (NMT2.0) |
Theoretical
Plates (NLT 2000) |
% RSD of peak area
(NMT 2.0) |
Tailing
Factor (NMT2.0) |
Theoretical
Plates (NLT 2000) |
% RSD of peak area
(NMT 2.0) |
||
1 | Unaltered | 1.10 | 6371 | 0.04 | 1.02 | 6678 | 0.11 |
2 | Volume 742.5 mL | 1.07 | 5576 | 0.21 | 1.01 | 6628 | 0.26 |
3 | Volume 757.5mL | 1.07 | 5678 | 0.18 | 1.02 | 6659 | 0.19 |
4 | 0.08N | 1.07 | 5422 | 0.10 | 1.04 | 6296 | 0.43 |
5 | 0.12N | 1.13 | 5390 | 0.12 | 1.03 | 6286 | 0.29 |
The system suitability parameters (Table 13) were not significantly changed with altered conditions.
The absolute difference in the results obtained under normal condition and robustness study of change in dissolution medium volume and dissolution medium strength (Table 11, 12) were within the acceptance criteria of NMT 10.
CONCLUSION: A simple and efficient method for estimation of % drug release of Levodopa and Carbidopa in triple drug combination product was developed and validated for specificity, accuracy, linearity, precision and robustness ensuring suitability of the method for quantitative analysis. The results indicated that this method is suitable for estimation of % drug release of Levodopa and % drug release of Carbidopa in a pharmaceutical formulation.
ACKNOWLEDGEMENTS: The authors express sincere gratitude to the Research and Development Center of Macleods Pharmaceuticals Limited, Mumbai, India, for granting permission to publish the data generated at Analytical Research and Development section for this research work. The first author is also thankful to his colleagues for their consistent support to carry out this research.
CONFLICT OF INTEREST: The authors declare that they have no conflict of interests.
REFERENCES:
- Stalevo (carbidopa, levodopa and entacapone) tablets, prescribing information, Novartis pharmaceuticals limited. http://www.accessdata.fda.gov/drugsatfda_docs/label/2010/020796s015lbl.pdf
- Madrakian, A. Afkhami, M. Borazjani, and M. Bahram: Simultaneous derivative spectrophotometric determination of levodopa and Carbidopa in pharmaceutical preparations. Bull. Korean Chem. Soc. 2009; 25(12): 1764- 1768.
- Bugamelli F. Marcheselli C, Barba E, Raggi MA: Determination of l-dopa, carbidopa, 3-O-methyldopa and entacapone in human plasma by HPLC-ED. Journal of Pharmaceutical and Biomedical Analysis 2011; 54(3): 562-567.
- Raut PP, Charde SY Simultaneous estimation of levodopa and carbidopa by RP-HPLC using a fluorescence detector: its application to a pharmaceutical dosage form. Luminescence 2014; 29(7):762-771.
- Ashish S. Doshi, kalpesh j. Upadhyay, tushar n. Mehta, and nageshnanda: Development and application of a high- performance liquid chromatographic method for the determination of in vitro drug release of levodopa, Carbidopa, and Entacapone from a tablet formulation. Journal of AOAC International 2009; 92(2): 394-403.
- Clésio S. Paim, Magda T. Martins, Marcelo D. Malesuik, Martin Steppe: LC determination of entacapone in entacapone tablets; in-vitro dissolution studies. Journal of Chromatographic Science 2010; 48:755-759.
- Tekale Prafullachandra, Mhatre Vinayak S, Pai N R, Maurya Chandrabhanu, Tekale Smrut: Estimation of Entacapone tablets by reverse phase high performance liquid chromatographic method. Bioscience Discovery 2011; 2(3): 294-298.
- Rama Krishna, K. Bala Murali Krishna, B.Hari Babu: Development and validation of liquid chromatographic method for the simultaneous estimation of Levodopa, Carbidopa and Entacapone in the combined dosage form. Journal of Pharmacy Research 2014; 8(3):281-288.
- Bujji babu, P. Srinivasa rao and R. Ramesh raju: Development of new robust RP-HPLC method for analysis of levodopa in formulations. International Journal of Science Invention Today 2012; 1(2): 130-143.
- Dhawan Raj Kumar, R Ravi, T Subburaju, H Revathi, C Arul and K. Gopalkrishnan: Development and validation of stability indicating assay method for levodopa and carbidopa in levodopa, carbidopa and entacapone ER tablets. International Journal of Pharmaceutical Research and Development 2013; 4(11): 098 – 105.
- Sravanthi D, Anusha M., Madhavi S, Firdose Shaik and Nalluri Buchi N: Simultaneous estimation of levodopa and Carbidopa in bulk, pharmaceutical dosage forms and dissolution sample analysis by RP-HPLC-PDA method. Journal of Chemical and Pharmaceutical Research 2013; 5(11): 422-428.
- Monograph for Entacapone tablets, USP 37 NF 32, 2014: 2801-2802.
- Monograph for levodopa tablets, USP 37 NF 32, 2014: 3535-3536.
- Monograph for levodopa and carbidopa orally disintegrating tablets, USP 37 NF 32, 2014: 2134-2136.
- Monograph for levodopa and carbidopa extended release tablets, USP 37 NF 32, 2014: 2129-2134.
- Monograph for levodopa and Carbidopa tablets, British pharmacopoeia vol. 3, 2014: 357-358.
- Monograph for levodopa and Carbidopa tablets, Indian pharmacopoeia vol. II, 2014: 2081-2084.
How to cite this article:
Bhatnagar P, Vyas D, Sinha SK and Gajbhiye A: HPLC method for simultaneous estimation of drug release of levodopa and carbidopa in entacapone, levodopa and carbidopa tablets. Int J Pharm Sci Res 2017; 8(3): 1091-01.doi: 10.13040/IJPSR.0975-8232.8(3).1091-01.
All © 2013 are reserved by International Journal of Pharmaceutical Sciences and Research. This Journal licensed under a Creative Commons Attribution-NonCommercial-ShareAlike 3.0 Unported License.
Article Information
14
1091-1101
978
1661
English
IJPSR
P. Bhatnagar *, D. Vyas, S. Kumar Sinha and A. Gajbhiye
Macleods Pharmaceuticals Limited, Mumbai, Maharashtra, India.
prasoon.bhatnagar@spsu.ac.in
13 August, 2016
12 October, 2016
19 October, 2016
10.13040/IJPSR.0975-8232.8(3).1091-01
01 March, 2017