IMPURITY PROFILING METHOD DEVELOPMENT AND VALIDATION FOR METFORMIN HCL AND VILDAGLIPTIN FROM COMBINATION TABLET DOSAGE FORM BY RP-HPLC COUPLED WITH UV/PDA DETECTOR
HTML Full TextIMPURITY PROFILING METHOD DEVELOPMENT AND VALIDATION FOR METFORMIN HCL AND VILDAGLIPTIN FROM COMBINATION TABLET DOSAGE FORM BY RP-HPLC COUPLED WITH UV/PDA DETECTOR
B. D. Musmade, S. G. Lokhande, M. S. Sable, R. R. Korhale, S. G. Bhope, M. Nagar and K. S. Lohar *
Department of Chemistry, Shrikrishna Mahavidyalay Gunjoti, Maharashtra, India.
ABSTRACT: A precise, accurate, robust method was developed for quantitative estimation of related impurities of Metformin HCL and Vildagliptin from combination tablet dosage form. The mixture of ammonium dihydrogen orthophosphate and octane sulfonic acid sodium salt buffer at pH 4 was used as mobile phase A and mixture of methanol and buffer in the ratio of95:5 used as a mobile phase B. The mobile phase was pumped at 0.8 ml/minutes flow rate through BDS Hypersil C8, 250 x 4.6 mm, 5µ, HPLC column operated at 35°C. All the solutions were injected at 10µl injection volume, and the chromatograms were monitored at 210 nm based on optimum response of impurities and analytes. The % recovery of Metformin HCL, was found 90.4, 92.3, 99.4 and 99.2%, while for Vildagliptin was 90.5, 94.5, 99.6 and 98.3 % from LOQ, 50,100 and 150 % levels respectively. The method was found linear from LOQ to 150% with the correlation coefficients (r2) were 0.998 for Metformin HCL and Vildagliptin. No interference observed from diluent and placebo at the retention time of all known impurities and principle analytes in specificity study and no significantly affected chromatographic pattern for standard, system suitability criteria and sample chromatograms during robustness study, hence the method is specific and robust. The method can be used in quality control laboratories for analysis of related substances from these combination drug products in pharmaceutical industries.
Keywords: Impurity profiling, Metformin HCl, Vildagliptin, Method validation, Forced degradation
INTRODUCTION: The Metformin HCL and Vildagliptin combination tablet is commonly prescribed medication worldwide to control the sugar level in the blood.
The purity, safety and efficacy of the medication is depending on the impurity level in the drug and drug products hence for batter efficacy the medication should have impurities well within the limits as per ICH guideline. It is most important to have an accurate, precise and robust method that can detect and able to accurately quantitate all the known and unknown degradants from drug and drug products.
Metformin Hydrochloride (MTF): MTF Fig. 1 is a biguanide derivative used to lower the blood glucose concentration in patients with non-insulin dependent diabetes mellitus 1. Chemically it is 1, 1-dimethylbiguanide hydrochloride easily soluble in water, marginally soluble in ethanol and almost insoluble in acetone and methylene chloride 2.
FIG. 1: STRUCTURE METFORMIN HCL (MTF)
Vildagliptin: The IUPAC name of Vildagliptin is (S)-1-[2-(3-Hydroxyadamantan-1-ylamino) acetyl] pyrrolidine-2- carbonitrileas shown in Fig. 2. It is anti-hyperglycemic agent having white to off-white solid powder that is soluble in water and Dimethyl Sulfoxide 3.
FIG. 2: STRUCTURE VILDAGLIPTIN (VIL)
After extensive literature review we come to know that, several assay methods for estimation of metformin HCL and Vildagliptin were reported by various technique like HPLC 4, 9, HPTLC 10, 11, HPLC-MS, and by UFLC 12. Individual assay determination of Vildagliptin as well as combination with other drug also reported by RP-HPLC 13-17, Spectrophotometry and spectroflurometric 18, GC 19, electrophoresis 20. Ramzia L. El Bagry and et. al reported the Vildagliptin determination in presence of its synthetic mixture in presence of Metformin HCL and Pioglitazone 21. In this reported method, does not included the Vildagliptin known impurities as well as impurities of MTF like Melamine, Cyanoguanidine, MTF related compound B, MTF related compound C. Caroline Paola Uber and et. al developed the method by HPLC–MS for determination of Vildagliptin and Metformin HCL with its impurities, but no Vildagliptin impurities were reported and eluted by this method 22. E. Al-Qudah, and et. al, did the Forced degradation studies of Vildagliptin API in the presence of excipients, by using HPLC-UV, but known impurities not reported and eluted 23. Best of our knowledge to the date no methods are reported for the simultaneous quantitation of known and unknown impurity profiling for Metformin HCL and Vildagliptin form their combination tablet dosage forms. Present study focused on simultaneous quantitation of known and unknown degradants of MTF and VIL from their combination tablet dosage form. The developed method was validated as per ICH guideline 24, 25 and can be used in quality control laboratories for routine analysis of commercial batches.
MATERIALS AND METHOD: MTF (99.6% purity) and VIL (99.7% purity) was in-house qualified working standards used for the complete development and validation study. The AR grade octane sulfonic acid sodium salt (Merck), ammonium dihydrogen orthophosphate (RANKEM), orthophosphoric acid (RANKEM), and HPLC grade acetonitrile, methanol (RANKEM) were used for mobile phase preparation. The complete development and validation was carried out on Shimadzu LC 2010 CHT, HPLC with Lab solution software and UV detector. Wavelength selection, selectivity study and method precision study was performed on Water’s HPLC with Empower 3 software connected with 2695 PDA detector. The in- house manufactured sample batches and placebo were used for the development and validation studies. The developed method was applied for the analysis of some marketed samples like TORGLIP M 50/1000, and TORGLIP 50/500.
Method Optimization: Method development trials were initiated by using monobasic phosphate buffer with ion pair reagents like hexane sulfonic acid , octane sulphonic acid sodium salt at different pH used as mobile phase A and acetonitrileasa mobile phase B, but the noisy base line and unwanted peaks was observed. Inertsil ODS 250 x 4.6 mm, 5 µ; BDS Hypersil, C18, 250 x 4.6 mm, 5µ; Baker bond C18, 250 x 4.6 mm, 5µ HPLC columns were tried to achieve the desire resolution between impurities. Finally 40 mm of ammonium dihydrogen orthophosphate buffer with 1.0 g/loctane sulfonic acid at pH 4 was selected as a buffer for mobile phase A. The mixture of buffer and methanol in the ratio of 88:12 was selected as mobile phase A and mixture of methanol and buffer in the ratio of 95:5 was used as a mobile phase B, which significantly improved retention of MTF and resolution between the VIL impurities on Hypersil BDS C8, 250 x 4.6 mm, 5µ HPLC column operated at 35°C. The gradient was optimized for better separation and chromatograms were monitored at 210 nm based on the optimum response of impurities and principle the analytes.
Chromatographic Condition: Accurately weighed 4.6 g ammonium dihydrogen orthophosphate, and 1g octane sulfonic acid sodium salt was transferred in 1 L HPLC grade water. The mixture was mixed well to dissolved, then sonicated and filtered through 0.45µ filter. The pH was adjusted to 4.0±0.1 with dilute orthophosphoric acid and used as buffer. The well mixed and degassed mixture of buffer and methanol in the ratio of 88:12 %v/v was used as mobile phase A. Degassed mixture of methanol and buffer in the ratio of 95:5 was used as mobile phase B. The mobile phase was pumped at 0.8ml/ minutes through.
BDS Hypersil C8, 250x4.6 mm, 5 µ columns which was operated at 35°C. The 10µl injection volume was load in to the column and the generated chromatograms were interpreted at 210 nm. The gradient as shown in Table 1 was optimized for better separation.
Diluent: The mixture of water: methanol in the ratio of 80:20% v/v used as diluent.
TABLE 1: GRADIENT PROGRAM
Time in Minutes | % Mobile phase A | % Mobile phase B |
0.1 | 100 | 0 |
15.0 | 100 | 0 |
50.0 | 80 | 20 |
65.0 | 80 | 20 |
65.1 | 100 | 0 |
80.0 | 100 | 0 |
Standard Preparation:
Preparation of MTF Standard Stock Solution: Accurately weighed 5 mg of MIF working standard was transferred into a 100 ml volumetric flask; about 70 ml of diluent was added and sonicated with intermittent shaking under controlled room temperature. The volume was made up to the mark with diluent.
Preparation of VIL Standard Stock Solution: Accurately weighed 10 mg of VIL working standard was transferred into a 100 ml volumetric flask; about 70 ml of diluent was added and sonicated with intermittent shaking under controlled room temperature. The volume was made up to the mark with diluent.
Preparation of Mix Standard Solution: (5ppm of MTF and 10 PPM of VIL): Mix standard was prepared by adding 5 ml of MTF standard stock solution and 5 ml of VIL standard stock solution in a 50 ml volumetric flask and adjusting the volume with diluent.
Preparation of MTF Impurity Stock Solution: Accurately weighed about 2.5 mg each, Melamine impurity, Cyanoguanidine, MTF related compound B, MTF Related compound C was transferred into a separate 10 ml volumetric flasks, dissolved by sonication and diluted up to the mark with diluent.
Preparation of VIL Impurity Stock Solution: Accurately weighed about 2.5mg each of VIL impurity A, B, C, and D, was transferred into a separate 10 ml volumetric flasks, dissolved by sonication and diluted up to the mark with diluent.
Preparation of System Suitability Solution: In a 50 ml volumetric flask, pipette 1 ml each of Cyanoguanidine, Melamine, MTF related compound B, and MTF related compound C, 5 ml MTF standard stock solution and 2 ml each of VIL Impurity A, B, C, D and 5 ml of VIL standard stock solutions and diluted up to the mark with diluent. The volume was made up the mark with diluent, mixed well and used as a system suitability solution.
Sample Solution Preparation: Tablet powder equivalent to 1 tablet was transferred into a 100 mL volumetric flask. About 70 mL of diluent was added, mixed well and sonicated for 20 minutes under controlled room temperature with intermittent shaking. The volume was made up to the mark with diluentand centrifuged the solution for 10 minutes at 5000 rpm. Then filtered through 0.45µ filter by discarding 5 ml of filtrate and injected.
System Suitability Criteria: The relative retention times for all the known impurities were calculated and listed in Table 2. The similarity factor between standard -1 and standard -2 should be within 95%-105%. The theoretical plates should not be less than 2000 and tailing factor not more than 2 for both the analytes in standard 1 and 2.
Impurity Calculations: The known impurities were calculated against respective drug standard area from standard solution 1. The unknown impurities were confirmed by injecting individual MTF and VIL API with placebo during FD study. Hence the rest of all unknown impurities were calculated against the MTF standard area from standard solution 1 as most of the generated unknown impurities were from MTF.
Analytical Method Validation: Analytical method validation for the developed method was carried out as per ICH guideline in terms of Accuracy, Precision, selectivity, Linearity, LOD/ LOQ and range.
Accuracy: Recovery of impurities in presence of active, placebo plays a vital role to assess the level and extent of recovery and extraction efficiency of the analytical method. Accuracy in terms of recovery for all the known impurities of MTF and VIL as well as principle analytes was done by spiking in the placebo from LOQ, 50%, 100% and 150% of specification level. All the preparations were injected in triplicate injections in to the HPLC system and recovery was calculated at each level.
Precision: In the precision study, the standard -1and Standard-2 were injected and calculated the similarity factor for both the analytes. The limit of similarity factor was set 95 to 105%.
Method Precision and Intermediate Precision: Six samples were separately prepared, injected and %impurities were calculated. The relative standard deviation of % individual known impurities, single maximum impurity and total impurities were calculated from the six sample preparations. The intermediate precision was conducted as like method precision but by different the analyst, different date, HPLC system, and column lots. Same study was repeated by spiking the known impurities at 100% level in the sample preparations.
Selectivity: Selectivity of the developed method was checked by specificity and by forced degradation study.
Specificity: Specificity of the method was checked by selectivity and force degradation study. In the selectivity all the individual solutions of each impurities and principle analytes were injected and confirmed the any interference from placebo, diluent at the retention time of each analytes.
Force Degradation Study: The stress study was applied for the developed method to check the stability indicating nature of the method. The sample, placebo and diluent were treated in acid, base, peroxide, condition, and thermal conditions. For acidic degradation 0.05NHCl at room temperature 6 Hrs and for base degradation study
0.05 N NaOH at room temperature for 15 minutes, were optimized. The solutions were neutralized; volume was adjusted and injected to check the degradation. The sample, placebo was kept at 60°C for 7 days in thermal degradation, after 7 days the samples were removed, prepared as per method of analysis and injected. Oxidation stress study was optimized by adding10% of total volume 30 % H2O2 and the solutions were kept at room temperature for 12Hrs. The mass balance was checked at every condition by using assay method and it was controlled between 95 to 105% of initial value.
LOD and LOQ: LOQ value was considered where the single to noise ratio (S/N) is more than 10 and for LOD it was consider 3. Limit of detection and limit of quantitation was determine by injecting minimum four levels below 50% level of working concentration. Each level was injected triplicate and LOD/LOQ was calculated by 10σ/S and 3σ/S method as per ICH guideline.
Linearity and Range: Linearity study prove that the method is capable to accurate measure the response of impurities and analytes which is directly proportional to it concentration. The study was conducted at various levels from LOQ to 150%. All the impurities solution and standard solution were spiked from LOQ to 150% and volume was makeup with diluent. Each level was injected in triplicate in the sequence and the correlation coefficient of each individual impurities as well as principle analytes was calculated and controlled by the limit r2=0.998.
The range of the method was established from lower to upper limits in which the response is directly proportional to analytes concentration.
Robustness: Ability of the developed method to remains unaffected by deliberate changes in the critical method parameters. Robust of the method was proved by varying the method parameters like change in column temperature ±5°C, wave length ±3nm, flow rate ± 0.1 ml, organic composition in mobile phase ±5% etc. The system suitability parameters were checked at each condition.
Solution Stability: Solution stability of the prepared samples, impurity standard solution, working standard solutions was monitored at room temperature. The solutions were injected at different time intervals like initial, after 12, 24, 48 Hrs. and check the system suitability criteria and any significant variation in peak response under the curve.
RESULTS AND DISCUSSION:
Selectivity: All the known impurities MTF and VIL along with principle analytes were well separate from each other’s and no interference was observed from placebo and diluent at the retention time of each impurities and principle analytes.
The representative chromatograms of diluent as blank, placebo, system suitability, and sample were shown in Fig. 3 to Fig. 6 respectively. The retention time and relative retention time were calculated for the each known impurity against its respective API as shown in Table 2.
The force degradation study shows that the no significant degradation hence the product is stable at thermal condition. In the basic condition 11.9% degradation was found while 19.5 % degradation observed in peroxide condition and 12.8% in degradation observed in acidic condition.
FIG. 3: REPRESENTATIVE CHROMATOGRAM OF DILUENT AS A BLANK
FIG. 4: REPRESENTATIVE CHROMATOGRAM OF PLACEBO
FIG. 5: REPRESENTATIVE CHROMATOGRAM OF SYSTEM SUITABILITY
FIG. 6: REPRESENTATIVE CHROMATOGRAM OF SAMPLE
TABLE 2: RETENTION TIME AND RELATIVE RETENTION TIME OF IMPURITIES
Sr. no. | Name of Impurities | Retention time in minutes | RRT against respective API |
1 | Cyanoguanidine | 4.1 | About 0.22, w.r.t. MTF. |
2 | Melamine | 13.2 | About 0.78, w.r.t. MTF. |
3 | MTF Related Comp B | 15.6 | About 0.85, w.r.t. MTF. |
4 | Metformin HCL | 18.2 | 1.0 |
5 | MTF Related Comp. C | 44.3 | About 2.4, w.r.t. MTF. |
6 | Vil. Imp. A | 5.2 | About 0.105 w.r.t. VIL. |
7 | Vil. Imp. B | 9.4 | About 0.189 w.r.t. VIL |
8 | Vil. Imp. D | 45.7 | About 0.92 w.r.t. VIL |
9 | Vildagliptin | 49.6 | 1.0. |
10 | Vil. Imp. C | 59.6 | About 1.20 w.r.t. VIL. |
Precision: In precision study the % of each known impurity, highest unknown impurity and total impurities were calculated form the six sample preparations in method precision and six sample preparation in intermediate precision.
The % RSD of total impurities in method precision is 6.5% while in intermediate precision was found 4.8% which was well within the limit (NMT15.0%). All the known impurities were observed below detection limits while the % RSD of known impurities in spike study were found below 15% hence the developed method is precise.
Linearity: The method found linear to measure the accurate response of analytes over the range of LOQ to 150% of impurity limit. The correlation coefficient (r2) for MTF and VIL is more than 0.998 as shown in Fig. 7 and 8 the linearity plots for both the analytes.
FIG. 7: LINEARITY OF METFORMIN HCL
FIG. 8: LINEARITY OF VILDAGLIPTIN
Accuracy: Recovery of MTF and VIL in presence of placebo from LOQ to 150% was found between 85 to 115%. Hence, the developed method has ability to recover all the impurities as well as analytes in selected diluent and volume and accurately quantitate from LOQ to 150% as shown in Table 3.
TABLE 3: RECOVERY STUDY
Impurity name | LOQ Level | 50 % Level | 100% Level | 150% Level |
MTF | 90.4 | 92.3 | 99.4 | 99.2 |
Melamine | 104.1 | 106.3 | 105 | 106 |
Cyanoguanidine | 104.8 | 106.5 | 100.8 | 103.2 |
MTF Related Comp B | 89.2 | 92.1 | 92.1 | 91.1 |
MTF Related Comp C | 94.5 | 88.3 | 91.4 | 85.5 |
VIL. Impurity A | 89.6 | 92.3 | 91.5 | 90.6 |
VIL. impurity B | 100.7 | 102 | 102.1 | 100.3 |
VIL. Impurity C | 88.8 | 94 | 101.8 | 1006 |
VIL. Impurity D | 95.7 | 104.7 | 103.6 | 103.2 |
VIL | 90.5 | 94.5 | 99.6 | 98.3 |
LOD and LOD Determination: The LOD and LOQ values for MTF and VIL were calculated as per ICH guideline. The observed LOD and LOQ values of MTF is 0.33µg/ml and 1.03µg/ml where as for VIL is 0.19µg/ml and 0.57µg/ml.
Robustness and Solution Stability: During the robustness study, it was observed that no significant variation observed in system suitability criteria, like resolution, retention time and RRT. Hence the developed method is robust. In solution stability at room temperature the solutions were found stable up to 3 days at room temperature.
ACKNOWLEDGMENT: The authors wish to thank Mr. Vinod Ramchandra Jadhav, chairman and Mr. Dinesh Kapoor, CEO Sava Healthcare Limited for their constant support and encouragement. Thanks are also due to Mr. Sachin Margaj for formatting the figures and graphs in this article.
Funding: The research was conducted without any funding source.
CONFLICT OF INTEREST: All authors declare no conflict of interest.
REFERENCES:
- Kommineni V, Chowdary KPR and Prasad SVUM: Development of new stability indicating RP-HPLC method for simultaneous stimation of Metforminhydrochloride and Teneligliptin Hydrobromide and its validation as per ICH guideline. Indo Am J Sci 2017; 4(05): 1109-1119. doi: http://doi.org/10.5281/zenodo.582315
- El-Bargy RI, Elkady EF and Ayoub BM: Liquid chromatographic methods for the determination of vildagliptinin the presence of its synthetic intermediate and the simultaneous determination of Pioglitazonehydrochloride and Metformin Hydrochloride. Int J Biomed Sci 2011; 7(3): 201-208. DOI: 10.59566/IJBS.2011.7201
- Sultana R, Bachar SC and Rahman F: Development and validation of stability indicating assay method of vildagliptin in bulk and tablet dosage form by RP-HPLC. Int J of Pharm & Life Sci 2013; 4(4): 2530-2534.
- Shakoor A, Ahmed M, Ikram R, Hussain S, Tahir A, Jan BM and Adnan A: Stability-Indicating RP-HPLC method for simultaneous determination of Metformin hydrochloride and Vildagliptin in Tablet and Biological samples. Acta Chromatographica 2020; 32(1): 39–43. DOI:10.1556/1326.2019.00555
- Baokar SB, Mulgund SV and Ranpise NS: Development and validation of RP-HPLC method for simultaneous estimation of Vildagliptin and Metformin. Research J Pharma Dosage Forms and Tech 2013; 5(2): 95-98.
- Varma D, Atmakuri L and Dinda SC: Simultaneous determination of Metformin and Vildagliptinin solid dosage form by stability indication RP-HPLC method. International Journal of Pharmacy and Pharmaceutical Science 2013; 4(1): 122-12.
- Thangabalan B, Manohar K, Babu S and Nagar M: Method development and validation for the simultaneous estimation of Vildagliptin and Metformin in Tablet dosage form by RP-HPLC. International Journal of Research in Pharmaceutical and Nano Sciences 2014; 3(2): 80-87.
- Anisetti R and Sundri C: Validated method for the simultaneous estimation of Metformin Hydrochloride and Vildagliptin by RP-HPLC in bulk and pharmaceutical dosage form. International Research Journal of Pharmaceutical and Applied Sciences 2012; 2(3): 22-28.
- Nagalakshmi V, Shrinivas Rao G, Gayatridevi N and Mohan S: RP-HPLC method for simultaneous estimation of Vildagliptin and Metformin inbulkand pharmaceutical formulations. International Journal of Current Research and Review 2021; 13(7): 112-117. Doi: http://dx.doi.org/10.31782/IJCRR.2021.13710.
- Shirode AR, Maduskar PD, Deodhar MS and Kadam VJ: RP-HPLC and HPTLC methods for simultaneous estimation of Metformin Hydrochloride and Vildagliptin from bulk and marketed formulation: development and validation. British Journal of Pharmaceutical Research 2013; 4(20): 2370-2386. DOI: 10.9734/BJPR/2014/12820.
- El-Kimary EI, Hamdy DA, Mourad SS and Barary MA: HPTLC Determination of Three gliptins in binary mixtures with Metformin. Journal of Chromatographic Science. 2016; 54(1): 79–87. doi: 10.1093/chromsci/bmv106.
- Mendhule RB, Warokar AS, Mahajan UN, Mahajan NM and Barde LN: New stability indicating UFLC method for simultaneous estimation of Metformin HCL and Vildagliptin in bulk and solid dosage. IJPSR 2021; 12(4): 2289-2295.
- Pednekar S, Lokhande R, Sutar R, Kolhal S, Surve S and Gudekar S: Simultaneous determination of Metformin, Sitagliptin, Saxagliptin, Linagliptin and Vildagliptin in multicomponent pharmaceutical preparations by RP-HPLC. International Journal of Pharmaceutical Sciences Review and Research 2014; 28(1): 128–133.
- Attimarad M, Nagaraja SH, Al-Dhubaib BE and Al-Najjar A: Development of a rapid reversed phase-high performance liquid chromatography method for simultaneous determination of Metformin and Vildagliptin in formulation and human plasma. Journal of Young Pharmacists 2014; 6(4): 40-46. DOI: 10.5530/jyp.2014.4.7
- Peraman R, Peruru KK, P. Reddy RY and Goera C: Stability indicating RP-HPLC method for simultaneous determination of Metformin Hydrochloride and Pioglitazone Hydrochloride indosage form. Malaysian Journal of Pharmaceutical Sciences 2014; 12(1): 33–46.
- Malakar A, Bokshi B and Nasrin D: Development and validation for estimation of Vildagliptin form tablet dosage form. International Journal of Pharmaceutical and Life Sciences 2012; 1(2): 1-8. DOI:10.3329/ijpls.v1i1.12947
- Boovizhikannan T and Palanirajan VK: RP-HPLC determination of Vildagliptin in pure and in tablet formulation. Journal of Pharmacy Research 2013; 7(1): 113–116. DOI:10.1016/j.jopr.2013.01.001
- Moneeb MS: Spectrophotometric and spectroflurometric methods for the determination of Saxagliptin and Vildagliptin in bulk and pharmaceutical preparations. Bulletin of Faculty of Pharmacy Cairo University 2013; 51(2): 139-150. doi: http://dx.doi.org/10.1016/j.bfopcu.2013.03.003
- Uçaktürk E: Development of sensitive and specific analysis of Vildagliptin in pharmaceutical formulation by Gas chromatography-mass spectrometry. Journal of Analytical Methods in Chemistry 2015; 1-7. doi:https://doi.org/10.1155/2015/707414
- Barden AT, Piccoli BL, Volpato NM, Schapoval EES and Steppe M: Capillary zone electrophoresis for determination of Vildagliptin (a DPP-4 inhibitor) in pharmaceutical formulation and comparative study with HPLC. Pharmazie 2014; 69(2): 86–91.
- Al-Rimawi F: Development and validation of an analytical method for Metformin hydrochloride and its related compound (1-cynogaunidine) in tablet formulation by HPLC-UV. Talanta 2009; 79(5): 1368-1371.doi:10.1016/j.talanta.2009.06.004.
- Uber CP, Degaut Pontes FL, Gasparetto JC, Guimares TM, Francisco DE and Piantavini MS: HPLC-MS/MS method for simultaneous quantitation of Vildagliptin and Metformin related compound in Tablets. Int J Pharm Sci 2014; 6(11): 203-207.
- Al-Qudah E, Arar S and Sweidan K: Forced degradation studies of Vildagliptin raw material alone and in the presence of excipients using HPLC-UV analysis. J. Excipients and Food Chem 2020; 11(2): 29-41.
- ICH-Q2A: Text on validation of analytical procedures. U.S. Department of Health and Human Services. USFDA, (R1); 1995; 1-13.
- ICH-Q2B: Validation of analytical procedures methodology. U. S. Department of Health and Human Services. USFDA 1996; (1): 1-10.
How to cite this article:
Musmade BD, Lokhande SG, Sable MS, Korhale RR, Bhope SG, Nagar M and Lohar KS: Impurity profiling method development and validation for metformin HCl and vildagliptin from combination tablet dosage form by RP-HPLC coupled with UV/PDA detector. Int J Pharm Sci & Res 2024; 15(2): 460-67. doi: 10.13040/IJPSR.0975-8232.15(2).460-67.
All © 2024 are reserved by International Journal of Pharmaceutical Sciences and Research. This Journal licensed under a Creative Commons Attribution-NonCommercial-ShareAlike 3.0 Unported License.
Article Information
19
460-467
699KB
377
English
IJPSR
B. D. Musmade, S. G. Lokhande, M. S. Sable, R. R. Korhale, S. G. Bhope, M. Nagar and K. S. Lohar *
Department of Chemistry, Shrikrishna Mahavidyalay Gunjoti, Maharashtra, India.
Dr.kslohar@rediffmail.com
06 June 2023
12 January 2024
28 January 2023
10.13040/IJPSR.0975-8232.15(2).460-67
01 February 2024