METHOD DEVELOPMENT AND VALIDATION FOR THE ESTIMATION OF RITONAVIR AND NIRMATRELVIR IN PHARMACEUTICAL DOSAGE FORM BY UV-SPECTROSCOPY
HTML Full TextMETHOD DEVELOPMENT AND VALIDATION FOR THE ESTIMATION OF RITONAVIR AND NIRMATRELVIR IN PHARMACEUTICAL DOSAGE FORM BY UV-SPECTROSCOPY
M. P. Chaithanya, S. Vijaya Bhaskar, Ayushi Chawla *, K. R. Puneeth Kumar, Shachindra L. Nargund and Shravan L. Nargund
Department of Pharmaceutical Quality Assurance, Nargund College of Pharmacy, Bengaluru, Karnataka, India.
ABSTRACT: The present study describes the development and validation of a simple, economical, accurate, and precise UV spectrophotometric method for the simultaneous estimation of Ritonavir and Nirmatrelvir in active pharmaceutical ingredients and tablet dosage forms, in accordance with ICH guidelines. Ritonavir, an HIV protease inhibitor, and Nirmatrelvir, a SARS-CoV-2 protease inhibitor, were selected for their combined therapeutic significance in COVID-19 antiviral therapy. Solubility profiling revealed that Ritonavir was freely soluble in acetonitrile and methanol, while Nirmatrelvir was soluble in methanol. Considering cost and solubility, acetonitrile and methanol were used as solvents for Ritonavir and Nirmatrelvir, respectively. The λmax values were found to be 239 nm for Ritonavir and 228 nm for Nirmatrelvir. Linearity was established in the ranges of 20–80 μg/mL for Ritonavir and 2–18 μg/mL for Nirmatrelvir, with excellent correlation coefficients (r² = 0.999 and 0.9998, respectively). The method was validated for accuracy, precision, and sensitivity. LOD and LOQ were 0.01285 μg/mL and 0.03894 μg/mL for Ritonavir, and 0.13419 μg/mL and 0.40662 μg/mL for Nirmatrelvir. Recovery studies confirmed accuracy within 98.06–102.56%, and %RSD values were within acceptable limits (<2%), indicating good precision. The assay of marketed tablets showed drug contents of 99.39% for Ritonavir and 99.41% for Nirmatrelvir. Thus, the proposed UV spectrophotometric method is reliable, cost-effective, and suitable for routine quality control analysis of both drugs in bulk and tablet forms.
Keywords: Virus, Antiviral agents, Analytical method validation, Ritonavir, Nirmatrelvir
INTRODUCTION:
Virus: A virus is a microscopic infectious agent. It consists of a segment of nucleic acid (either DNA or RNA) surrounded by a protein coat. Viruses can infect hosts such as humans, animals, and plants. They cannot reproduce without a host 1.
Antiviral drugs assist your body in fending off specific viruses that might lead to illness. Additionally, antiviral medications are preventative. In other words, they can shield you from contracting a virus or infecting others. Antiviral therapy is what medical professionals sometimes refer to as antiviral medication 2.
Analytical Method Development: The present work is to develop and validate UV Spectroscopic method for the determination of Ritonavir Fig. 1 and Nirmatrelvir Fig. 2 in API and pharmaceutical dosage formas per ICH Guidelines for solubility, selection of solvent, selection of Wavelength, Linearity, Range, Accuracy, Precision, Limit of Detection (LOD) and Limit of Quantification (LOQ) were assessed 3. Ritonaviris an HIV protease inhibitor 4 and Nirmatrelvir is an orally bioavailable inhibitor of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) 5.
FIG. 1: THE STRUCTURE OF RITONAVIR1,3-THIAZOLE-5-YLMETHYL N-[(2S, 3S, 5S)-3-HYDROXY-5- [[(2S)-3METHYL-22 [[METHYL-[(2-PROPAN-2-YL-1, 3-THIAZOL-4YL) METHYL] CARBAMOYL] AMINO] BUTANOYL] AMINO]1,6-DIPHENYLHEXAN 2-YL] CARBAMATE 4
FIG. 2: THE STRUCTURE OF NIRMATRELVIR(1R,2S,5S)-N-[(1S)-1-1CYANO-2-[(3S)-2-OXOPYRROLIDIN-3-YL] ETHYL] 3-[(2S)-3,3-DIMETHYL-2-[(2,2,2-TRIFLUOROACETYL) AMINO] BUTANOYL] 6,6-DIMETHYL-3-AZABICYCLO [3.1.0] HEXANE-2-CARBOXAMIDE 5
MATERIALS AND METHODOLOGY:
Materials: The Glassware’s and Apparatus used for the Research work were made up of borosilicate and that were calibrated before used in the research. For weighing the required chemicals in the range of (10 mg –700 mg)a calibrated digital weighing balance (sartorious-TE-214S) was utilized (max capacity – 200 gm)A Shimadzu (UV-1700Pharmaspec) double beam UV-Visible Spectrophotometer with spectral width of 2mm, Wavelength Accuracy of 0.5nm and a pair of 10mm matched quartz cell were used. A Shimadzu (FTIR-8400S) was used for determining FT-IR range for the APIs with palletization sampling technique using KBR (AR Grade). Ultra-Sonicator (RC system-MU1700) was used for getting the uniform mixture of solvent and the sample. Ritonavir API and Nirmatrelvir API were purchased from Dhamtec Pharma Consultants Navi Mumbai, Maharashtra. India. Tablets of brand Combovir (Nirmatrelvir 150mg and Ritonavir 100mg) were purchased from Bharat Pharma New Delhi, India.
Methodology:
Solubility: 10 mg of RTV was weighed and added to 3 different 10 ml volumetric flasks and add the small quantity of Water, Methanol and Acetonitrile respectively and shaken for 1-2 min and then made up the volume with respective solvents where results were found that the drug RTV was insoluble in water and freely soluble in Methanol and Acetonitrile.
10 mg of NMV was weighed and added to 3 different 10 ml volumetric flasks and add the small quantity of Water, Methanol and Acetonitrile respectively and shaken for 1-2 min and then made up the volume with respective solvents where results were found that the drug NMV was insoluble in water and freely soluble in Methanol.
Selection of Solvent: RTV was soluble in both Acetonitrile and Methanol so in the present research work Acetonitrile was selected as a solvent to make the method economical (i.e. cheaper than the methanol).
NMV was soluble in Methanol so in the present research work Methanol was selected as a solvent.
Selection of Wavelength: The standard stock solution- 1 of Ritonavir (1000μg/ml) was prepared by dissolving 10 mg of pure drug in 10 ml of Acetonitrile , from the 1st stock solution transfer 1 ml of solution to a 10 ml volumetric flask and made up the volume with water and resulting solution with the concentration (100μg/ml i.e. stock solution -2 ), then the final solution was prepared by taking 1 ml of from the 2nd stock solution to the 10 ml volumetric flask and made up the volume to 10 ml with water and the resulting solution of 10μg/ml. The solution (10μg/ml) was scanned from 200 to 400nm and selected 239nm with the absorbance value of 1.023 for RTV. The standard stock solution of Nirmatrelvir (1000μg/ml) was prepared by dissolving 10 mg of API in 10 ml of Methanol from the 1st stock transfer 1 ml of solution to a 10 ml volumetric flask and make up the volume with water making the solution concentration (100μg/ml), the final solution was prepared by taking 1 ml of solution from the 2nd stock to 10 ml volumetric flask and make up the volume to 10 ml with water to obtain the final concentration of 10μg/ml. The solution was scanned from 200 to 400nm and found 228nm with absorbance value of 0.948 for NMV.
Selection of Analytical Concentration Range:
Selection of Analytical Concentration Range for RTV: 10 volumetric flasks (each 10 ml) were taken and labelled as C1-C10. From the 2nd stock solution 10-90 μg/ml conc’s were prepared and made up the volume with water and all the solutions were scanned from the 200nm - 400nm then the absorbancewas measured at 239nm and calibration curve was plotted for Abs v/s Conc with sextuplicate results Table 1 and Fig. 3.
Selection of Analytical Concentration Range for NMV: 10 volumetric flasks (each 10 ml) were taken and labelled as C1-C10. From the 2nd stock solution 2-14 μg/ml conc’s were prepared and made up the volume with water and all the solutions were scanned from the 200nm - 400nm then the absorbance was measured at 228nm and calibration curve was plotted for Abs v/s Conc with sextuplicate results Table 2 and Fig. 4.
Analysis of Tablet Formulation:
Analysis of Ritonavir Tablets (100 mg): Twenty Tablets of RTV (Combovir) were weighed, average weight was calculated crushed using mortar and pestle to obtain fine powder of RTV. Accurately weighed tablet powder equivalent to 10mg of powder was transferred to 25 ml volumetric flask with 10 ml of ACN and sonicated for 20 min at 26 0C for the extraction of drug and volume was made up to 25 ml with ACN. Then resulting solution was filtered through Whatman filter paper no. 41, where initial filtrate drops were discarded and rest of the solution was used as Sample stock A (400μg/ml). From the stock A 1 ml of the aliquot was pipetted out and transferred to 10 ml volumetric flask and made up the volume with water to obtain final concentration of 40μg/ml. The resulting solution was scanned from 200-400nm and absorbance was measured at 239nm 9.
Analysis of Nirmatrelvir Tablets (150mg): Twenty Tablets of NMV (Combovir) were weighed, average weight was calculated crushed using mortar and pestle to obtain fine powder of RTV. Accurately weighed tablet powder equivalent to 10mg of powder was transferred to 25 ml volumetric flask with 10 ml of Methanol and sonicated for 20 min at 26 0C for the extraction of drug and volume was made up to 25 ml with Methanol. Then resulting solution was filtered through Whatman filter paper no. 41, where initial filtrate drops were discarded and rest of the solution was used as Sample stock A (400μg/ml). From the Stock Solution A, 1 ml of the aliquot was pipetted out and transferred to 10 ml volumetric flask and made up the volume with water to obtain final concentration of 40μg/ml. The resulting solution was scanned over 200-400nm and absorbance was measured at 228nm 10.
Validation of Spectrophotometric Method:
Accuracy Studies of Ritonavir and Nirmatrelvir:
Ritonavir: To study the accuracy assay of synthetic mixture was carried out. Recovery studies were caried out by Standard addition method by adding the known amount of RTV (Std) separately to the pre-analyzed sample at three different concentration levels i.e., 80%, 100%, and 120% of assay concentration and percentage recovery was calculated Table 3.
Nirmatrelvir: To study the accuracy assay of synthetic mixture was carried out. Recovery studies were caried out by Standard addition method by adding the known amount of NMV (Std) separately to the pre-analyzed sample at three different concentration levels i.e., 80%, 100%, and 120% of assay concentration and percentage recovery was calculated Table 4.
Precision Studies of Ritonavir and Nirmatrelvir: The precision of an analytical method was studied by performing intermediate precision and Repeatability.
Precision Study of Ritonavir:
Intermediate Precision:
Intra-day Precision: Variation of results were analyzed by measuring the Abs of Std solution of RTV (40,50,60 μg/ml Conc’s) at 239nm on three different time intervals (0hr, 2hr, 4hr) of same day. The % RSD was calculated by using the following formula: Table 5.
%RSD = (SD/mean) × 100
Inter-day Precision: Variation of results were analyzed by measuring the Abs of Std solution of RTV (40,50,60 μg/ml Conc’s) at 239nm on same time of three consecutive days. The % RSD was calculated using the following formula: Table 7.
%RSD = (SD/mean) × 100
Repeatability: Standard solution of 40μg/ml of RTV was prepared in triplicate and absorbance was measured at 239nm. %RSD was calculated Table 9.
Limit of Detection (LOD) and Limit of Quantification (LOQ): Detection limit and quantification limit were calculated from the standard intercept of y- intercepts of six calibration curves and average slope of six calibration curves. The formula is given as follows:
LOD = 3.3 x (standard Deviation of y-intercepts of six calibration curves) / (Average Slope of six calibration curves)
LOQ = 10 x (Standard Deviation of y-intercepts of six calibration curves) /Average Slope of six calibration curves
Precision Study of Nirmatrelvir:
Intermediate Precision:
Intra-day Precision: Variation of results were analyzed by measuring the Abs of Std solution of NMV (40,50,60 μg/ml Conc’s) at 228nm on three different time intervals (0hr, 2hr, 4hr) of same day. The % RSD was calculated using the following formula: Table 6.
%RSD = (SD / mean) × 100
Inter-day Precision: Variation of results were analyzed by measuring the Abs of Std solution of NMV (40,50,60 μg/ml Conc’s) at 228nm on same time of three consecutive days. The % RSD was calculated using the following formula: Table 8.
%RSD = (SD/mean) × 100
Repeatability: Standard solution of 40μg/ml of NMV was prepared in triplicate and absorbance was measured at 228nm. %RSD was calculated using the formula: Table 10.
Limit of Detection (LOD) and Limit of Quantification (LOQ): Detection limit and quantification limit were calculated from the standard intercept of y- intercepts of six calibration curves and average slope of six calibration curves. The formula is as follows 6:
LOD =3.3 x (Standard Deviation of y-intercepts of six calibration curves) / (Average Slope of six calibration curves)
LOQ = 10 x (Standard Deviation of y-intercepts of six calibration curves) / Average Slope of six calibration curves
RESULTS:
Linearity studies of Ritonavir and Nirmatrelvir:
Ritonavir:
FIG. 3: THE LINEARITY OF THIS METHOD WAS DETERMINED AT RANGES FROM 20-80 ΜG/ML FOR RITONAVIR
TABLE 1: LINEARITY RANGE OF RITONAVIR
| Conc μg/ml | Average Absorbance | SD | % RSD |
| 20 | 0.00207 | 0.00207 | 0.92367 |
| 30 | 0.00557 | 0.00557 | 1.60473 |
| 40 | 0.00343 | 0.00343 | 0.72039 |
| 50 | 0.00829 | 0.00829 | 1.37498 |
| 60 | 0.00163 | 0.00163 | 0.22147 |
| 70 | 0.00117 | 0.00117 | 0.13512 |
| 80 | 0.00476 | 0.00476 | 0.46691 |
FIG. 4: CALIBRATION CURVE OF RITONAVIR. [The Linearity for Ritonavir was found to be linear in the range of 20-80 μg/ml with R2= 0.999 and the straight-line equation as Y=0.0132x + 0.0472 and R2=0.999.]
Nirmatrelvir:
FIG. 5: THE LINEARITY OF THIS METHOD WAS DETERMINED AT RANGES FROM 2-18ΜG/ML FOR NIRMATRELVIR
TABLE 2: LINEARITY RANGE OF NIRMATRELVIR
| Conc. μg/ml | Average Absorbance | SD | % RSD |
| 2 | 0.13567 | 0.00121 | 0.89267 |
| 4 | 0.24467 | 0.00294 | 1.20324 |
| 6 | 0.37467 | 0.00151 | 0.40184 |
| 8 | 0.50483 | 0.00279 | 0.55204 |
| 10 | 0.6345 | 0.00266 | 0.41995 |
| 12 | 0.759 | 0.00303 | 0.39962 |
| 14 | 0.87633 | 0.00151 | 0.1718 |
| 16 | 1.00833 | 0.00378 | 0.37459 |
| 18 | 1.13417 | 0.00319 | 0.28113 |
FIG. 6: CALIBRATION CURVE OF NIRMATRELVIR. [The Linearity for Nirmatrelvir was found to be linear in the range of 2-18 μg/ml with R2= 0.9998 and the straight-line equation as Y=0.0629x + 0.0017 and R 2=0.9998.]
Accuracy Studies of Ritonavir and Nirmatrelvir: The accuracy of analytical method for Ritonavir and Nirmatrelvir was determined at 80%, 100% and 120% levels of standard solution. Absorbance was measured at 239nm for Ritonavir and 228nm for Nirmatrelvir. And results were expressed in terms of % studies.
Ritonavir:
TABLE 3: ACCURACY STUDIES OF RITONAVIR
| Level of Recovery | Absorbance | % Recovery |
| 80% | 0.989 | 99.09 % |
| 100% | 1.09 | 98.75 % |
| 120% | 1.23 | 101.001 % |
Nirmatrelvir:
TABLE 4: ACCURACY STUDIES OF NIRMATRELVIR
| Sl. no. | Level of Recovery | Absorbance | % Recovery |
| 1 | 80% | 1.112 | 98.06 % |
| 2 | 100% | 1.255 | 99.62 % |
| 3 | 120% | 1.550 | 102.56 % |
Precision Studies of Ritonavir and Nirmatrelvir: In Precision the measurement of intraday, interday, repeatability and results were showed good reproducibility with the present relative standard deviation (% RSD) was below 2.0%. This indicated that this method was highly precise 7.
TABLE 5: INTRADAY PRECISION OF RITONAVIR
| Conc. μg/ml | Absorbance | |||||
| 0 hour | 2 hours | 4 hours | Average | SD | % RSD | |
| 40 | 0.476 | 0.474 | 0.472 | 0.474 | 0.002 | 0.421941 |
| 50 | 0.61 | 0.599 | 0.596 | 0.601667 | 0.007371 | 1.225116 |
| 60 | 0.733 | 0.729 | 0.726 | 0.729333 | 0.003512 | 0.48152 |
TABLE 6: INTRADAY RESULTS OF NIRMATRELVIR
| Conc. μg/ml | Absorbance | |||||
| 0 hour | 2 hours | 4 hours | Average | SD | % RSD | |
| 6 | 0.387 | 0.385 | 0.384 | 0.38533 | 0.00153 | 0.39642 |
| 10 | 0.637 | 0.635 | 0.634 | 0.63533 | 0.00153 | 0.24043 |
| 14 | 0.886 | 0.885 | 0.884 | 0.885 | 0.001 | 0.11299 |
TABLE 7: INTERDAY RESULTS OF RITONAVIR
| Conc. μg/ml | Absorbance | |||||
| Day 1 | Day 2 | Day 3 | Average | SD | % RSD | |
| 40 | 0.476 | 0.47 | 0.468 | 0.471333 | 0.004163 | 0.883309 |
| 50 | 0.593 | 0.589 | 0.584 | 0.588667 | 0.004509 | 0.766011 |
| 60 | 0.709 | 0.702 | 0.699 | 0.703333 | 0.703333 | 0.729612 |
TABLE 8: INTERDAY RESULTS OF NIRMATRELVIR
| Conc μg/ml | Absorbance | |||||
| Day 1 | Day 2 | Day 3 | Average | SD | % RSD | |
| 6 | 0.382 | 0.383 | 0.384 | 0.383 | 0.001 | 0.2611 |
| 10 | 0.63 | 0.628 | 0.629 | 0.629 | 0.001 | 0.15898 |
| 14 | 0.88 | 0.878 | 0.882 | 0.88 | 0.002 | 0.22727 |
Repeatability:
TABLE 9: REPEATABILITY OF RITONAVIR
| Conc μg/ml | Average Absorbance | SD | %RSD |
| 40 | 0.571667 | 0.001528 | 0.267206 |
TABLE 10: REPEATABILITY RESULTS OF NIRMATRELVIR
| Conc μg/ml | Average Absorbance | SD | %RSD |
| 10 | 0.627333 | 0.000577 | 0.092032 |
Acceptance Criteria:
TABLE 11: ACCEPTANCE CRITERIA FOR ALL THE PARAMETERS
| Sl. no. | Parameters | Limit |
| 1 | Linearity Range (%RSD) | <2 |
| 2 | Correlation Coefficient (r2) | ≤1 |
| 3 | Analysis of Tablets (Mean % Assay) | 98-105% |
| 4 | Accuracy(%Recovery) | 98-105% |
| 5 | Intra Day Precision (%RSD) | <2 |
| 6 | Inter Day Precision (%RSD) | <2 |
| 7 | Repeatability (%RSD) | <2 |
CONCLUSION: The present study successfully developed and validated a simple, precise, accurate, and cost-effective UV spectrophotometric method for the quantitative determination of Ritonavir (RTV) and Nirmatrelvir (NMV) in bulk APIs and tablet formulations, following ICH guidelines. The method showed excellent linearity, precision, accuracy, and sensitivity. The results confirmed that both drugs could be accurately and reliably estimated in bulk and pharmaceutical dosage forms. The low %RSD values indicated good repeatability and intermediate precision, while the high recovery rates confirmed the accuracy of the method. Thus, this UV method can be effectively employed for routine quality control analysis of Ritonavir and Nirmatrelvir in pharmaceutical industries.
ACKNOWLEDGMENT: I would like to thank the Dept. of Pharmaceutical Quality Assurance, Nargund college of Pharmacy, Bengaluru for providing the research facilities for conducting my research work. I would like to thank Dhamtec Pharma Consultancy, Mumbai for providing API on time. I would like to thank my Principal, Guide and Co-Guide for continuous encouragement and support.
CONFLICT OF INTEREST: The authors declare that no conflicts of interest exist related to this research. The study was carried out independently as part of an academic research project at the Department of Pharmaceutical Quality Assurance, Nargund College of Pharmacy, Bengaluru. No external funding, financial assistance, or commercial sponsorship was received for conducting or publishing this work. All materials and facilities were provided by the institution for academic purposes only. The authors alone are responsible for the content and writing of this manuscript.
REFERENCES:
- Irshad Anwar: Virus: Types, Characteristics, Facts, Topics, Structure [Internet]. Careers360;2025[cited 2025 jun 21].Avilable from:https://www2.careers360.com/biology/virus-topic-pge
- Full Classification of Antiviral Medications with Examples [Internet]. Pharmchoices [cited 2025 jun 21].Avilable fromhttps://pharmchoices.com/antiviral-medications-classifications-examples/
- International council for harmonisation of technical requirements for pharmaceuticals for human use. Validation of Analytical Procedures [Internet]. ICH 2023[cited 2025 March15]. Avilable from:https://database.ich.org/sites/default/files/ICH_Q2%28R2%29_Guideline_2023_1130.pdf
- Cayman Chemicals [Internet].2023[cited 2024 JULY 30]. Available from:https://www.caymanchem.com/product/13872/image
- Cayman Chemicals [Internet].2023[cited 2024 JULY 30]. Available from: (https://www.caymanchem.com/product/35257/pf-07321332)
- https://www.ich.org/page/ich-guidelines
- Swartz ME and Krull IS: Analytical Method Development and Validation. 2023 ed. London (UK): T&F India 2023; 1-120.
- Peerzade MY, Memon S, Bhise K and Aamer AI: Development and validation of UV-Visible spectrophotometric method for estimation of ritonavir in bulk and formulation. The Phar Inno J 2019; 8(4): 30-4.
- Skoog DA, Holler FJ and Crouch SR: Principles of Instrumental Analysis. 7th ed. India release: Cengage India 2020; 988.
- Tripathi KD: Essentials of Medical Pharmacology. 8th ed. New Delhi: Jaypee Brothers Medical Pub 2019; 234-36.
- Nash G: Uncoding Everything about Obesity: 1sr ed. India: amazon Kindle 2023; 1-28.
How to cite this article:
Chaithanya MP, Bhaskar SV, Chawla A, Kumar KRP, Nargund SL and Nargund SL: Method development and validation for the estimation of ritonavir and nirmatrelvir in pharmaceutical dosage form by UV-spectroscopy. Int J Pharm Sci & Res 2026; 17(3): 946-53. doi: 10.13040/IJPSR.0975-8232.17(3).946-53.
All © 2026 are reserved by International Journal of Pharmaceutical Sciences and Research. This Journal licensed under a Creative Commons Attribution-NonCommercial-ShareAlike 3.0 Unported License.
Article Information
18
946-953
722 KB
5
English
IJPSR
M. P. Chaithanya, S. Vijaya Bhaskar, Ayushi Chawla *, K. R. Puneeth Kumar, Shachindra L. Nargund and Shravan L. Nargund
Department of Pharmaceutical Quality Assurance, Nargund College of Pharmacy, Bengaluru, Karnataka, India.
ayushi.pharma2905@gmail.com
11 September 2025
10 November 2025
31 December 2025
10.13040/IJPSR.0975-8232.17(3).946-53
01 March 2026