QUANTITATIVE EVALUATION OF IMPURITIES G AND S IN VALACICLOVIR HYDROCHLORIDE HYDRATE ACTIVE PHARMACEUTICAL INGREDIENT USING LIQUID CHROMATOGRAPHY-MASS SPECTROMETRY ANALYTICAL METHOD
HTML Full TextQUANTITATIVE EVALUATION OF IMPURITIES G AND S IN VALACICLOVIR HYDROCHLORIDE HYDRATE ACTIVE PHARMACEUTICAL INGREDIENT USING LIQUID CHROMATOGRAPHY-MASS SPECTROMETRY ANALYTICAL METHOD
Subodhkumar M. Baluni * and Virupaksha A. Bastikar
Amity Institute of Biotechnology, Amity University, Mumbai-Pune Expressway, Bhatan, Panvel, Maharashtra, India.
ABSTRACT: Current work makes known that Liquid Chromatography-Mass Spectrometry (LC-MS) method development and validation for the impurity G (N, N -dimethylpyridin-4-amine) and impurity S (2-[(2-amino-6-oxo-1,6-dihydro-9H-purin-9-yl) methoxy] ethyl N-[1,1-dimetylethoxy) carbonyl]-L-valinate) in Valaciclovir Hydrochloride Hydrate active pharmaceutical ingredient (API). The developed LC-MS analytical method is complimentary to the thin layer chromatography (TLC) method available for impurity G and impurity S quantitation in the Valaciclovir Hydrochloride Hydrate API monograph. Impurity G and Impurity S were determined by the LC-MS method in Q1 Multiple ion mode using Ascentis Express C18 (15cm X 4.6 mm) 2.7µm analytical HPLC column. A gradient system was employed for the elution of analytes using acetonitrile (Eluent B) and 0.01M Ammonium Formate, LC-MS compatible volatile buffer, pH 3.0 (Eluent A) in distinct compositions. The gradient system (T/%B) was applied as 0.01/5, 4.00/5, 7.50/80, 10.00/80, 12.50/5, and 20.00/5. A method developed was validated considering the International Conference on Harmonization pharmaceutical guidelines. The quantitation limit found for Impurity G and Impurity S were 207.20ppm and 216.00 ppm.
Keywords: Impurity G, Impurity S, Valaciclovir Hydrochloride Hydrate API, LC-MS, Q1 Multiple Ion, ICH guidelines
INTRODUCTION: L-valyl ester hydrochloride salt of acyclovir is Valaciclovir. Chemical name of Valaciclovir is 2-[(2-amino-6-oxo-1, 6-dihydro-9H-purin-9-yl) methoxy] ethyl L-valinate Fig. 1C. Valaciclovir Hydrochloride Hydrate is widely used to kill simplex and zoster herpes virus. Viral thymidine kinase phosphorylated valaciclovir to acyclovir triphosphate and this help in inhibition of herpes viral DNA replication.
The active ingredient Valaciclovir Hydrochloride, Hydrate has two process-related impurities, Impurity G and Impurity S. Impurity G is N, N -dimethylpyridin-4-amine with the molecular formula C7H10N2 Fig. 1A. Impurity S is 2-[(2-amino-6-oxo-1,6-dihydro-9H-purin-9-yl) methoxy] ethyl N-[1,1-dimetylethoxy) carbonyl]-L-valinate. Molecular formula of impurity S is C18H28N6O6 Fig. 1B.
Impurity G is comparatively highly harmful, is destructive to the lungs and eyes, and gets absorbed through the skin 8. Impurities S and G are official impurities of Valaciclovir Hydrochloride, Hydrate API as per European pharmacopeia. Chiral method development and method validation research article is available for Valacyclovir drug and its associated constituent’s guanine, acyclovir and unknown impurities are controlled using a high-performance liquid chromatograph method. The Valacyclovir and its associated impurities were well separated in the published method. The HPLC method is also published for assay and purity determination of Acyclovir and Valacyclovir. Literature for evaluating antiviral drugs acyclovir and valacyclovir with their impurity guanine using micellar electrokinetic chromatography (MEKC) is available. Literatures available related to Valaciclovir Hydrochloride, Hydrate are less in number, but Acyclovir related literatures are available in public domain. p-Toluenesulfonic acid is a potent impurity and method developed to quantify the impurity in acyclovir drug substancesat trace or residual level by using API-4000 LC-MS/MS is also available.
MATERIALS AND METHODS:
Chemicals & Regents: Procured Impurity G and Impurity S from authorised supplier of pharmacopeial impurities. Ammonium Format of LCMS grade and acetonitrile were purchased from Honeywell, India. Arranged Valaciclovir Hydrochloride Hydrate API sample from Pharma manufacturer, India.
TABLE 1: IMPURITY G, IMPURITY S AND VALACICLOVIR HYDROCHLORIDE HYDRATE STRUCTURES
Impurity G(1A) | Impurity S(1B) | Valaciclovir Hydrochloride Hydrate(1C) |
Analytical Instrumentation: Analytical instrument and analytical method parameter used for the impurity G and impurity S quantification method development in Valaciclovir Hydrochloride Hydrate API are given below in Table 2.
TABLE 2: ANALYTICAL INSTRUMENT AND METHOD PARAMETER DETAILS
Liquid Chromatograph | |
Pump Details | Shimadzu LC-20AD Pump |
Detector Details | Shimadzu SPD-20A Detector |
Auto Sampler | Shimadzu SIL-20AC/HT |
Colum Thermostat | Shimadzu column Thermostat CTO-10ASvp |
Chromatography Method Details | |
Eluent A | 0.01M ammonium Formate, LCMS compatible volatile buffer, pH 3.0 |
Eluent B | Acetonitrile (Cyanomethane)100% v/v |
Analytical HPLC Column | Ascentis Express Octadecylsilane (15cm × 4.6mm) 2.7µm |
Flow Rate | 1.0mL/min, Flow Splitter used, Pass 0.5mL/min into the MS source |
Thermostat temperature | 15 degrees Celsius |
SamplerCooler Temperature | 5degrees Celsius |
The Injection amount | 5.0μl |
System Runtime | 20.0Minutes |
Mass Spectrometer Parameter | |
Mass Spectrometer | AB Sciex API 4000 model (Made in Singapore) |
Ionization Probe | Electrospray ionization |
Ionization Mode | Positive |
Scan Type | Q1 Multiple Ions |
Impurity G Molecular Mass Details | 123.2 Dalton in Positive Mode |
Impurity S Molecular Mass Details | 425.2 Dalton in Positive Mode |
Declustering potential | 50 V |
Entrance potential | 10V |
Curtain gas flow | 35 (PSI) |
Ion Spray Voltage(V) | 5500V |
Ion Source Gas 1 | 30 Nebulization pressures (PSI) |
Ion Source Gas 2 | 50 Nebulization pressures (PSI) |
Valco Valve Details | Venting was given from 4.1 to 7.4 minutesfirst then 9.9 to 18 minutes. |
Data Acquisition & Processing Software | Analyst 1.6.3 |
Solutions Preparation for Standard and Sample: Prepared different type of standard and sample solution to carry out the study as per given below. All the prepared solutions given below in Table 3 were sonicated well before the analysis.
TABLE 3: STANDARD AND SAMPLE SOLUTION PREPARATION DETAILS
Solvent Mixture | Water: Ethanol mixture (20:80 v/v). |
Stock solution Preparation | |
0.52mg/mL stock solution of impurity G and 0.54mg/mL stock solution Impurity S prepared in solvent mixture | |
Diluent | Acetonitrile: Mobile phase A (10:90) |
0.05 mg/mL stock solution preparation | |
Transferred 1.0ml of stock solution of impurity G and impurity S into 10mL volumetric flask and then diluted to 10mL with diluent to prepare 0.05mg/mL solution | |
0.001 mg/mL stock solution preparation | |
Transferred 1.0ml of 0.05mg/mL solutioninto 50mL volumetric flask and then diluted to 50mL with diluent to prepare 0.001mg/mL solution | |
Standard solution preparation | |
Transferred 0.50ml of 0.001 mg/mL solutioninto 10ml volumetric flask and then diluted to 10ml with diluent. Concentration of impurity G and Impurity was 518ppm and 540ppm in standard solution with respect to Valaciclovir Hydrochloride, Hydrate API concentration. | |
Sample solution Preparation | |
Prepared Valaciclovir Hydrochloride, Hydrate sample solution at 0.1 mg/mL by dissolving suitablequantity of drug substance in the diluent | |
Recovery studies solution preparation | |
Prepared LOQ level-1, 50% level-2, 100% level-3 and 150% level-4 accuracy solution forimpurities G and S with respect to Valaciclovir Hydrochloride, Hydrate API concentration for recovery studies by diluting the impurities stock solution with the essential amount of diluent | |
Linearity Study solutions preparation | |
Prepared Linearity solutions for impurity G at 207.20, 259.00, 414.40, 518.00, 621.60and777.00 ppm and for impurity S at 216.00, 270.00, 432.00, 540.00, 648.00and810.00 ppm by diluting the stock solution with requisite amount of diluent |
Analytical Method Development Details: To obtain sensitivity, recovery and separation between impurity G and impurity S along with Valaciclovir Hydrochloride, Hydrate API several analytical columns like Waters Xbridge C18 (150mm X 4.6mm,5.0μm), Ascentis Express Octadecylsilane (10cm X 4.6mm, 2.7μm), Ascentis Express Octadecylsilane (50mm X 4.6mm, 2.7μm), Inertsil ODS (25cm X 4.6 mm,5.0μm) were evaluated during method development along with the different type of mobile phase A.
The recovery of impurity G and impurity S was found within the acceptance range on Ascentis Express Octadecylsilane (15cm × 4.6mm, 2.7μm) analytical column, with as an eluent A (10mM Ammonium Formate LCMS Compatible buffer, pH 3.0) and as an eluent B (Acetonitrile) along with gradient system. On Ascentis Express C18 (15cm X 4.6mm, 2.7μm), injected impurity G, impurity S and sample solution using MRM mode by keeping transition 123.20 -> 107.20 for impurity G and transition 425.20 -> 369.30 for impurity S. Responses for impurity G and impurity S were found satisfactory. Mobile phases of multiple compositions were verified with different mobile phase flow rate. 0.02% Trifluoracetic acid and 0.1% Formic acid in water as an eluent A and Methanol as an eluent B were used in another method development trial. The isocratic method was tried to set initially but it later confirmed the gradient method. Method development trials were taken using Multi reaction monitoring (MRM) mode, but the response of impurity and accuracy was not found in an acceptable range. Finally, looking at the analytical method development study data, selected an eluent A (0.01M Ammonium Formate LCMS compatible volatile buffer, pH 3.0) an eluent B (Acetonitrile (100 v/v)) with gradient run. Gradient programme was provided in Table 4.
RESULTS AND DISCUSSION:
Advantage of LC-MS Method over TLC Method: Thin layer chromatography (TLC) method was available for Impurities G and S determination in the European pharmacopeia monograph of Valaciclovir Hydrochloride Hydrate API, and the limit of impurity G and impurity S are 0.05% or 500ppm with respect to Valaciclovir Hydrochloride Hydrate API sample concentration.
Thin layer chromatography plate image Fig. 1 as per the monograph method, is given below.
FIG. 1: IMPURITY G AND IMPURITY S TLC PLATE
TLC quantitation method showing reproducibility issues. The length of the TLC plate is restricted and therefore achieved restricted separation quality. Time consumption is a big concern associated with the TLC method. Looking at the disadvantage connected with the present method available in monograph, the current LC-MS analytical method was developed by testing different column chemistry stationary phases to obtain significant separation of the impurity G and impurity S with Valaciclovir Hydrochloride Hydrate API and recovery within the acceptance criteria. A method developed using LC-MS was selective and sensitive compared to the available European pharmacopeia TLC method. It can be used for exact evaluation of impurity G and impurity S in valaciclovir Hydrochloride hydrate antiviral drug. LC-MS method run time was 20 minutes and the mobile phase flow rate was 1.0ml. Impurity G and Impurity S were well separated in LCMS method and proved orthogonality compared to the available TLC method.
Method Validation: The method specificity was verified by infusing the diluent, individual impurities, and Valaciclovir Hydrochloride Hydrate API sample, and the chromatograms of connected solutions are available in Fig. 2-6.
FIG. 2: XTRACTED ION CHROMATOGRAM (XIC), BLANK-IMPURITY S
FIG. 3: XTRACTED ION CHROMATOGRAM (XIC), BLANK- IMPURITY G
The diluent chromatogram in Fig. 2 & Fig. 3 showed that no interfering peak was observed at the retention times of Valaciclovir Hydrochloride Hydrate APIas well as impurities. Impurities G and S, xtracted chromatograms of in Fig. 4 and Fig. 5 displayed that Impurities G and S eluted at the retention times of 2.88 min and 8.38 min, respectively. Established method chromatograms display the absent of any interfering peak at Valaciclovir Hydrochloride, Hydrate API, and impurities G and S retention time. A developed analytical method could distinguish G and S impurities with each other and with the main Valaciclovir Hydrochloride Hydratedrug.
FIG. 4: XTRACTED ION CHROMATOGRAM (XIC), IMPURITY G STANDARD
FIG. 5: XTRACTED ION CHROMATOGRAM (XIC), IMPURITY S STANDARD
FIG. 6: VALACICLOVIR HYDROCHLORIDE HYDRATE UV CHROMATOGRAM
The limit of detection (LOD) and lower limit of quantification (LLOQ) determined for impurities G and S from signal to noiseratio. Prepared the lower concentrations of standard solutions to obtain lower limit of quantification in this procedure.
The lower limit of quantification of impurities G & S are 207. 20 ppm and 216.00 ppm and LLOQ solutions of impurities G and S give signal-to-noise ratios 245.5 and 219.0, respectively.
FIG. 7: IMPURITY G S/N RATIO
FIG. 8: IMPURITY S S/N RATIO
TABLE 4: ELUENT GRADIENT PROGRAM
Time(minutes) | Eluent A(%) | Eluent B(%) |
0.01 | 95.0 | 5.0 |
4.00 | 95.0 | 5.0 |
7.50 | 20.0 | 80.0 |
10.00 | 20.0 | 80.0 |
12.50 | 95.0 | 5.0 |
20.00 | 95.0 | 5.0 |
The developed analytical method linearity in Q1 Multiple ions scan type was established by injecting impurity G and impurity S at many levels of the concentrations between LLOQ and 150 % of the target concentration. The calibration curve was designed by drawing the chart between the peak response and impurity G concentration at 207.20, 259.00, 414.00, 518.00, 621.60 and 777.00 ppm and Impurity S at 216.00, 270.00, 432.00, 540.00, 648.00 and 810.00 ppm. Carried out linear least square regression study to get the slope, correlation coefficient values, and intercept. Linearity data were accessible in Table 5.
A spiking study was carried to establish the accuracy of the newly analytically developed method by spiking the impurities G and S at LLOQ level, 100% and 150 % of the specification concentrations, concerning the Valaciclovir Hydrochloride, Hydrate API sample concentration.
LLOQ level, Limit level, and 150% Level determination were carried out, and corresponding data is presented in Table 6. The recovery of impurity G and Impurity S at three points (LLOQ level, 100 % and 150 %) should be in the limit of 85.0 % to 115.0 %, and the relative standard deviation should be not more than 10.0 %. Recovery values of 96.0 % to 98.3 % for impurity G and 98.6 % to 103.4 % for impurity S were obtained with % RSD 1.53 & 2.96, respectively.
TABLE 5: RESULTS OF LINEARITY FOR IMPURITY G AND IMPURITY S
Linearity Level | Impurity G Concentration in ppm | Impurity G Area | Impurity S
Concentration in ppm |
Impurity S Area |
Level 1 | 207.20 | 2263698 | 216.00 | 231327 |
Level 2 | 259.00 | 2750265 | 270.00 | 326601 |
Level 3 | 414.00 | 4525288 | 432.00 | 558096 |
Level 4 | 518.00 | 5577955 | 540.00 | 705811 |
Level 5 | 621.60 | 6567234 | 648.00 | 829608 |
Level 6 | 777.00 | 8691013 | 810.00 | 1137051 |
Correlation Coefficient(r2) | 0.9984 | Correlation Coefficient(r2) | 0.9976 | |
Slope | 11096.19 | Slope | 1472.18 | |
Intercept | -110971.06 | Intercept | -84064.92 |
FIG. 9: XTRACTED ION (XIC) OF IMPURITY G-LLOQ SPIKED SAMPLE
FIG. 10: XTRACTED ION (XIC) OF IMPURITY S-LLOQ SPIKED SAMPLE
TABLE 6: RESULTS OF ACCURACY FOR IMPURITY G AN DIMPURITY S
Impurity Name | Impurity G | Impurity S | |||||
Level | Theoretical
Conc in ppm wrtSm |
Measured
Conc in ppm wrtSm |
Recovery %
|
Theoretical
Conc in ppm wrtSm |
Measured
Conc in ppm wrtSm |
Recovery % | |
LLOQ (40%) | 200.07 | 196.69 | 98.3 | 198.89 | 196.12 | 98.6 | |
100% | 248.05 | 245.69 | 98.8 | 246.59 | 241.64 | 98.0 | |
150% | 744.12 | 714.29 | 96.0 | 739.71 | 764.69 | 103.4 | |
% R.S.D. | 1.53 | % R.S.D. | 2.96 | ||||
Repeatability and ruggedness study was carried out to check method precision of developed analytical method. Repeatability was checked by spiking specification level impurity G and impurity S, standard concentration in six freshly prepared sample solutions on the same day and RSD of content of impurity G and impurity S were checked. Relative standard deviation should be not more than 10.0 % and the corresponding data is presented in Table 7. The relative Standard deviation of impurity G and impurity S obtained 1.9% and 2.1% respectively. LLOQ level precision was also checked and % relative standard deviation of six replicate injections was 1.67 for impurity G and 3.92 for impurity S, respectively. Connected data of LLOQ precision are presented in Table 8.
TABLE 7: RSULTS OF SPIKE PRECISION FOR IMPURITY G AND IMPURITY S
Injection | Impurity G Concentration obtained in sample | Impurity S Concentration obtained in sample |
1 | 467.18 | 461.52 |
2 | 465.93 | 462.86 |
3 | 489.98 | 461.82 |
4 | 478.62 | 475.45 |
5 | 474.84 | 486.09 |
6 | 478.76 | 472.33 |
Mean | 475.885 | 470.012 |
S.D. | 8.8361 | 9.84 |
R.S.D.% | 1.9 | 2.1 |
TABLE 8: RESULTS OF PRECISION AT LLOQ LEVEL
Injection | Area of Impurity G in LLOQ Solutions | Area of Impurity Sin LLOQ Solutions |
1 | 2253187 | 217423 |
2 | 2267775 | 236221 |
3 | 2197894 | 243270 |
4 | 2274616 | 235322 |
5 | 2312591 | 225579 |
6 | 2276122 | 230144 |
Mean | 2263697.50 | 231326.50 |
S.D. | 37739.51 | 9058.77 |
R.S.D.% | 1.67 | 3.92 |
Ruggedness was checked by spiking the specification level standard concentration of impurity G and impurity S in six freshly prepared sample solutions on a different day, and the commutative relative standard deviation of the content of each impurity between spike precision and intermediate precision should be not more than 10.0 %. Connected data are presented in Table 9.
TABLE 9: RESULTS OF RUGGEDNESS FOR IMPURITY G AND IMPURITY S
Injection | Impurity G Concentration obtained in sample | Impurity S Concentration obtained in sample |
1(Precision) | 467.18 | 461.52 |
2(Precision) | 465.93 | 462.86 |
3(Precision) | 489.98 | 461.82 |
4(Precision) | 478.62 | 475.45 |
5(Precision) | 474.84 | 486.09 |
6(Precision) | 478.76 | 472.33 |
1(Ruguddness) | 496.26 | 517.10 |
2(Ruguddness) | 497.90 | 526.81 |
3(Ruguddness) | 522.21 | 518.63 |
4(Ruguddness) | 491.53 | 520.05 |
5(Ruguddness) | 515.95 | 528.40 |
6(Ruguddness) | 502.98 | 515.28 |
Mean | 490.18 | 495.53 |
S.D. | 18.01 | 27.70 |
% R.S.D. | 3.67 | 5.59 |
Intentional changes in flow rate, mobile phase pH checked the method's robustness. The flow rate of the eluent in the method of analysis was 1.0mL, which was changed by 10% (0.90 to 1.00 mL/min). The mobile phase pH effect on the analysis was explored at 2.8 pH and 3.2 pH (Mobile phase pH changed by +0.2 units). All the changes in the above-mentioned parameters did not show any considerable changes in separation of impurity G and impurity S from the Valaciclovir Hydrochloride Hydrate and on chromatographic performance.
To prove the stability of impurity G and impurity S solutions, specification level impurities solution spiked in the sample solution and kept at room temperature (25oC) for 48 hrs. Solution stability was evaluated by calculating the percent relative standard deviation of area of impurities G and S solution between 0 hrs and 39 hrs. The percent relative standard deviation of the area of impurities G and S solution should be not more than 20.0 %. The data presented in Table 10 revealed that the solution of impurities G and S was steady up to 39.30 hrs at room temperature.
TABLE 10: SOLUTION STABILITY DATA OF IMPURITY G AND IMPURITY S
Conditions | Impurity G Area in PPM | Impurity S Area in PPM |
At 0 hrs | 8715433 | 664232 |
At RT for 39.30 hrs | 7505262 | 583583 |
% R.S.D. | 15.6 | 12.6 |
CONCLUSION: Accurate, sensitive, selective, specific analytical method developed for the quantification of impurities G and S in Valciclovir Hydrochloride, Hydrate API at 0.05% with respect to Valaciclovir Hydrochloride, Hydrate API sample concentartion using liquid chromatograph mass spectrometer. Electrospray ionization source/probe was used in positive mode of ionization. Also verified that LC-MS method is more sensitive and effective than TLC method for quantifying impurities G and S. Specificity, precision, linearity, accuracy, and solution stability studies were performed to validate the analytical method. The method specificity was proved by the acceptable resolution of impurities with the Valaciclovir Hydrochloride Hydrate API. This method linearity covered 207.20 ppm to 777.00 ppm concerning Valaciclovir Hydrochloride Hydrate API sample for impurity G, 216.00 ppm to 810.00 ppm concerning Valaciclovir Hydrochloride Hydrate API sample for impurity S with a coefficient correlation of 0.9984 & 0.9976 respectively. The recovery values confirmed the method accuracy in the range of 96.0 % to 98.3 % for impurity G, 98.02 % to 103.4 % for impurity S with % RSD 1.53 & 2.96, respectively. This developed method is sensitive with a lower limit of quantification of 207.20 for impurity G and 216.00 ppm for impurity S.
ACKNOWLEDGMENT: We thank Dr. Virupaksha A. Bastikar, Associate Professor from Amity Institute of Biotechnology, Amity University, for providing technical support during the research.
CONFLICTS OF INTEREST: The authors have no conflicts of interest regarding this investigation.
REFERENCES:
- Ranavkumar Shah and Bhavin Dhadhuk: Related Impurities High-performance Liquid Chromatography Method Development and Validation for drug combinations: Olmesartan Medoxomil, Chlorthalidone and Cilnidipine. Int J of Pharma Scien and Drug Research 2020; 12: 1-10.
- Mabbu IP, Sumathi G & Devanna N: New LC-MS Method Development and Validation for Quantitative Estimation of 2-Acetoxy Ethyl Acetoxy Methyl Ether in Acyclovir. Research Journal of Pharmacy and Technology, 2021; 14(11): 5881-5885.
- Revathi G, Elavarasi S & Saravanan K: Phytochemical screening and GC-MS analysis of polyherbal formulation for diabetes. Int J Adv Res 2018; 6(5): 1439-43.
- Mabbu IP, Sumathi G & Devanna N: A novel LC-MS method development and validation for the determination of phenyl vinyl sulfone in eletriptan hydrobromide. Future Journal of Pharmaceutical Sciences 2021; 7(1): 1-6.
- Kłysik K, Pietraszek A, Karewicz A & Nowakowska M: Acyclovir in the treatment of herpes viruses–a review. Current Medicinal Chemistry 2020; 27(24): 4118-4137.
- Shi J, Hu Y, Smith DE & Zhu HJ: A sensitive liquid chromatography-tandem mass spectrometry method for the quantification of valacyclovir and its metabolite acyclovir in mouse and human plasma. Journal of Chromatography B 2018; 1092: 447-452.
- Reçber T, Haznedaroğlu İC & Çelebier M: Review on characteristics and analytical methods of rivaroxaban. Critical Reviews in Analytic Chem 2022; 52(4): 865-877.
- Irakli M, Skendi A, Bouloumpasi E, Chatzopoulou P, & Biliaderis CG: LC-MS identification and quantification of phenolic compounds in solid residues from the essential oil industry. Antioxidants 2016; 10(12).
- Chidella KS, Dasari VB & Anireddy J: Ultra-Sensitive LC-MS/MS Method for the Trace Level Quantification of Six Potential Genotoxic Nitrosamine Impurities in Telmisartan. American Journal of Analytical Chemistry, 2021: 12(6): 227-240.
- Ozkan SA & Bakirhan NK: Pharmaceutical Analysis: Current Status and Future Perspectives. Current Pharmaceutical Analysis 2021; 17(3): 301-302.
- Sowjanya G and Mohana K: Quantification and stability aspects of luliconazole in bulk and pharmaceutical dosage forms by UV spectroscopy. Journal of Drug Delivery and Therapeutics 2019; 9(2): 300-06.
- Tomal M, Razibul H, Pritam R, Ratan P and Nazmul H: Method development and validation of RP-HPLC method for estimation of luliconazole in marketed formulation. The Pharma Innovation Journal 2019; 8(5): 103-08.
- Tripathi KD: Essentials of Medical Pharmacology. Jaypee Publications 2019; 8: 842-43.
- Eswarudu MM, Sakheena P, Lahari K and Srinivasa Babu PM: Chinna Eswaraiah, Validated RP-HPLC Method for Simultaneous Estimation of Atenolol, Hydrochlorothiazide and Losartan Potassium in Bulk and Pharmaceutical Dosage Form. As J of Phar Res and Dev 2021; 9: 60-66.
- Kalshetti MS and Kankure SS: Development and validation of RP HPLC method for simultaneous estimation of metoprolol, telmisartan and cilnidipine in tablet. IJPSR 2021; 12: 1651-1657.
- Akshay Paliwal, Neha Tiwari and Pragnesh Patani: RP-HPLC method development & validation for simultaneous estimation of chlorthalidone, metoprolol succinate and telmisartan in tablets, Journal of Emerging Technologies and Innovative Research (JETIR) 2021; 8: 804-811.
- Kondeti H. Pavani and Dannana G. Sankar: Development and Validation of Stability-indicating Method for the Estimation of Cilnidipine, Olmesartan Medoxomil and Chlorthalidone by Reverse Phase High Performance Liquid Chromatography. International Journal of Pharmaceutical Quality Assurancen 2021; 2: 53-59.
- shenzhenausa pharmaceutical co ltd [cn]; henzhenausa pharmaceutical co., ltd.; shenzhenausa pharmed co., ltd, bai jie [cnchen ping [cn]; tian minqing [cn]; xu xiping [cn]; yu duo [cn], bai, jie; pharmaceutical composition containing amlodipine, chlorthalidone, and amiloride. WO2020207355A1•2020.
- weightechhunan pharmaceutical co ltd, he liang, high performance liquidchromatography analysis method and detection method for impurities in telmisartan medicine. CN107966519A•2018.
- jiangxixinglinbaima pharmaceutical co ltd, li baolin; liu yiping; luoyunhua; zhu qin, analysis method for impurity detection of telmisartan tablet and telmisartan capsule, CN110836943A•2020.
- Jinan kanghe medical tech co ltd,lipeng; liuwanlu; xu chunhai; yang jiangyong; zhangying; zhangzuoping, Atenolol injection and preparation method thereof CN108888593A•2018.
- Mei xuefeng [cn]; shanghai instmateria medica cas [cn]; wang jianrong [cn]; yuqihui [cn], meixuefeng [cn]; wang jianrong [cn]; yuqihui [cn], cocrystal of telmisartan and hydrochlorothiazide, US2020181121A1•2020.
- Suzhou chunghwa chemical & pharmaceutical ind co ltd,li dan, telmisartan hydrochlorothiazide tablet and preparation method thereof,CN112641745A•2021.
- Chengdu hengruipharmaceutical, englu; hanpengqin; li xuemei; liu hui; liujiajun; ma yiming; shenhua; wei wen; xiao shiyuan; zhangmao; zhangyu; zhudeqi, losartan potassium oral preparation and preparation method thereof.CN112516105A•2021.
- Sichuan sunrise biopharmltd,dai ping; liuqiuhong; pengxianfeng; ren zhijun; tang yan; ye zizheng, preparation method of amlodipine losartan potassium compound composition, CN112274490A•2021.
- Alphacaithangzhou ai bio tech co ltd, li bin, application of losartan potassium and dacarbazine combined medicine for treating intestinal cancer. CN112438986A•2021.
- Zhuhai rundu pharmaceutical co ltd, guan doing; li jinqing; shenjilan; xie bin, olmesartanmedoxomil tablet and preparation method thereof, CN111956624A•2020.
- Changsha chenchen pharmaceutical tech co ltd,zhang gen, preparation and separation method of olmesartanmedoxomil dimer impurity, CN111004223A•2020.
- Changzhou pharmaceutical factory co ltd, liuruichao; ren jinfeng; tang lijuan, valsartan oral solid preparation.CN112315926A•2021.
- Zhejiang huahai pharm co ltd [cn]; zhejianghuahaitiancheng pharmaceutical co ltd [cn]; zhejianghuahaizhicheng pharmaceutical co ltd [cn], wang peng [cn]; wang yifeng [cn]; yan fengfeng [cn]; zhuxiaoren [cn]; zhuyuanxun [cn], valsartan refinement method. WO2021022516A1•2021.
- Wang T, Yang H, Yang J, Guo N, Wu G, Xu X & An M: Quantitative Determination of Four Potential Genotoxic Impurities in the Active Pharmaceutical Ingredients in TSD-1 Using UPLC-MS/MS. Molecules 2022; 27(13): 4129.
- Rao GS, Ramadevi D, Rao BM, Rajana N & Basavaiah K: Novel stability indicating LC-MS method for N-Nitroso dimethyl amine genotoxic impurity quantification in ranitidine drug substance and drug product. Journal of Applied Pharmaceutical Science 2022; 12(7): 106-114.
How to cite this article:
Baluni SM and Bastikar VA: Quantitative evaluation of impurities G and S in valaciclovir hydrochloride hydrate active pharmaceutical ingredient using liquid chromatography-mass spectrometery analytical method. Int J Pharm Sci & Res 2023; 14(1): 330-39. doi: 10.13040/IJPSR.0975-8232.14(1).330-39.
All © 2023 are reserved by International Journal of Pharmaceutical Sciences and Research. This Journal licensed under a Creative Commons Attribution-NonCommercial-ShareAlike 3.0 Unported License.
Article Information