RP-HPLC METHOD QUANTITATIVE ESTIMATION OF TAPENTADOL HYDROCHLORIDE AND ITS TABLETS
HTML Full TextRP-HPLC METHOD QUANTITATIVE ESTIMATION OF TAPENTADOL HYDROCHLORIDE AND ITS TABLETS
Mamta Bishnoi 1, Yash Paul Singla 2, Ankit Jain 3, * 4 and Birendra Shrivastava 1
Department of Pharmaceutics 1, School of Pharmaceutical Sciences, Jaipur National University, Jaipur - 302017, Rajasthan, India.
Department of Pharmaceutical Sciences 2, Guru Jambheshwar University of Science and Technology, Hisar - 125001, Haryana, India.
Pharmaceutics Research Projects Laboratory 3, Department of Pharmaceutical Sciences, Dr. Harisingh Gour Central University, Sagar - 470003, Madhya Pradesh, India.
Department of Materials Engineering 4, Indian Institute of Science, Bangalore - 560012, Karnataka, India.
ABSTRACT: The objective was to develop and validate an easy, economical, fast, reliable, reproducible, precise and accurate reversed-phase high-performance liquid chromatography (RP-HPLC) method for the estimation of tapentadol hydrochloride (TAP, a mu opioid-receptor agonist, and noradrenaline reuptake inhibitor) in the bulk and pharmaceutical dosage form. The chromatographic separation was achieved by using the HPLC system equipped with a C-18(2) column (250 × 4.6 mm, Particle size 5µm) at wavelength 217 nm. The mobile phase consisting of Potassium phosphate buffer: Acetonitrile (50:50 v/v) was used in isocratic mode. The flow rate was fixed at 1.0 mL/min with a continuous run up to 5 min, while the retention time was located near about 2.5 min. In the concentration range of 60-160 µgmL-1, the detector response was found linear with linear regressed equation Y=22791X+120122. In the assay of TAP, 99.65% of the drug was recovered. This method proved a satisfactory validation for all the parameters such as accuracy, linearity, specificity, precision, range, ruggedness, robustness, reproducibility, and peak purity assessment as per ICH guidelines. The results of the study evidenced that it is useful for the routine determination of tapentadol in the bulk and pharmaceutical dosage forms like a tablet.
Keywords: |
Tapentadol, Pharmaceutical dosage forms, Analytical validation, ICH Guidelines, HPLC
INTRODUCTION: Tapentadol, 3-[(1R,2R)-3-(dimethylamino)-1-ethyl-2-methyl]-propylphenol hydrochloride (TAP), the first US FDA-approved centrally acting analgesic for the treatment of moderate to severe acute pain 1, 2.
It is a schedule-II controlled substance, which has a dual mechanism of action, μ-opioid receptor agonist and noradrenaline (norepinephrine) reuptake inhibitor with little serotonin reuptake inhibition 3. Traditional opioids react through a single mechanism (µ opioid receptor binding), while TAP has a non-opioid mechanism also. Relative to a classical µ-opioid at equal analgesia doses, TAP has ≤ a 40%µ load, which contributes to its lower adverse effect profile, i.e., greater gastrointestinal tolerability 4. Tapentadol HCl has molecular formula C14H23NO.HCl with molecular weight 257.80 g.
The partition coefficient log P value n-octanol: water is 2.87 5. TAP has low bioavailability (32%) due to first-pass metabolism and less plasma protein-bound (only 2%) with a half-life of 4.9 h 6. After oral administration, ˃ 95% of the dose is excreted in the urine as glucuronide and sulfate conjugated metabolites within 24 h 7. So far, various methods are reported for the estimation of TAP in bulk and in pharmaceutical with different mobile phases, retention time, linearity range conc., and absorption maxima. In the present study, a trial has been made to develop a rapid, convenient and simple reverse phase HPLC method for the determination of TAP in the pharmaceutical dosage form i.e., Lucynta tablet 50 mg (Lupin Laboratories Ltd.).
FIG. 1: CHEMICAL STRUCTURE OF TAPENTADOL HYDROCHLORIDE
EXPERIMENTAL SECTION:
Chemicals and Reagents: Tapentadol hydrochloride (TAP) was procured as a gift sample from Innova Captab, Baddi (H.P.). Lucynta tablet (Lupin Laboratories Ltd.) was obtained from the market. Cellulose nitrate membrane filters (0.45 mm and 0.22 mm), acetonitrile, and Dipotassium Phosphate buffer of HPLC grade were brought from Sigma Aldrich, New Delhi, India. All analytical grade chemicals were used unless otherwise stated. All solutions were prepared in purified HPLC grade water produced by the Milli-Q Millipore Water System (Milford, MA, USA).
Preparation of Standard Stock Solutions: A stock solution of TAP (50mg/50mL) was prepared in the volumetric flask by adding a mobile phase with the help of bath sonication (up to 15 min), and the final volume was adjusted. Appropriate aliquots (60, 80, 100, 120, 140, and 160 µg/mL) of drug solutions were prepared from 100 µg/mL solution with the mobile phase. For validation of various parameters, other dilution samples (i.e., 1, 10, and 50 µg/mL) were also prepared in appropriate concentrations and stored at -20 °C 8. All the resultant solutions before use were filtered through Whatman filter paper number 41 9.
Preparation of Sample Solution: A working solution of the tablet was prepared with twenty tablets of TAP. All the tablets were crushed to a fine powder in the mortar and pestle. On the basis of the average weight of the tablets, fine powder equivalent to one tablet was accurately weighed and transferred to 25 mL volumetric flask. After the addition of 10 mL of the mobile phase, the volumetric flask was sonicated for 20 min in the bath sonicator to obtain a clear solution. The final volume was adjusted with the mobile phase, and the whole solution was filtered through a 0.45 µm nylon filter. Suitable aliquots of TAP solutions were prepared in the range of linearity, as previously described in the standard solution.
Instrumentation and Chromatographic Conditions: The chromatographic separation was carried out by using the HPLC system included a Shimadzu model LC-10 ADVP(Shimadzu Co., Kyoto, Japan), having a Phenomenex Luna C-18(2) column (250 × 4.6 mm, dp = 5 mm), SPD-M20A Prominence Diode array detector, (holding the deuterium lamp with a sensitivity of 0.005 AUFs), a Shimadzu model C-R5A chromatograph integrator module (chart speed at 10 mm/min), a Shimadzu model SIL-6A auto-injector, and a Shimadzu module SCL-6A system. The mobile phase used was a mixture of 0.1 M potassium dihydrogen phosphate buffer (pH adjusted to 7 with triethanolamine) and acetonitrile in the ratio of 50:50% v/v, degassed and filtered through the 0.45μm cellulose acetate membrane filter. Injection volume was fixed 20 µL with a flow rate of 1 mL/min under the isocratic mode of the mobile phase. The column was maintained at 25 °C temperature, and Photodiode array detector wavelength was fixed at 217 nm. Lab Solutions software was used for the acquisition and processing of chromatographic data.
Method Validation: Various preliminary trials were conducted by changing various chromatographic parameters to develop a new method for TAP analysis in bulk and finished pharmaceutical product i.e., tablet. The developed method was validated as per ICH guidelines for linearity, accuracy, precision, robustness, the limit of detection (LOD), the limit of quantification (LOQ), and stability of the stock solution, using the serial dilution method 2.
Peak Purity Assessment: To judge peak purity, the evaluation was carried out after the development and validation of the appropriate analytical conditions in the TAP analysis. Based on the degree of similarity of UV spectra (Peak range 190-800 nm), peak purity was assessed by class VP software (Shimadzu HPLC system). If the peak purity index was greater than the single point (i.e., a positive value of the minimum peak purity index), then it was classified as pure. To evidence accuracy, six replicate analyses of a TAP at a concentration of 60 µg/mL were performed. The acceptance limit was set at ±2% for the percent coefficient of variation (% CV) to the peak area, plate number, tailing factor, the peak asymmetry of samples, and the retention time of TAP, as shown in Table 1.
Linearity: Linearity was evaluated by calculating the correlation coefficient 10. A calibration curve of TAP with six different concentrations from 60 to 160 µg/mL range was plotted with peak area versus concentrations, while the samples were injected into the HPLC column in triplicate. Linear regression analysis was carried out for slope, intercept, and linear correlation coefficient (r2) calculation 11. The data was given in Table 1.
TABLE I: LINEARITY DATA OF TAPENTADOL HCl
Sample name | Sample ID | Ret. Time (min.) | Area | Theoretical Plate | Tailing Factor |
TAP | 60 PPM | 2.500 | 1576187 | 1924.156 | 1.643 |
60 PPM | 2.507 | 1572020 | 1932.659 | 1.604 | |
60 PPM | 2.508 | 1572629 | 1861.918 | 1.600 | |
(Average) | 2.505 | 1573612 | 1906.244 | 1.615 | |
80 PPM | 2.519 | 2036280 | 1954.655 | 1.635 | |
80 PPM | 2.527 | 2040222 | 1887.808 | 1.616 | |
80 PPM | 2.533 | 2043011 | 1936.561 | 1.661 | |
(Average) | 2.526 | 2039838 | 1926.341 | 1.637 | |
100 PPM | 2.538 | 2447755 | 1945.700 | 1.641 | |
100 PPM | 2.539 | 2450911 | 1943.186 | 1.633 | |
100 PPM | 2.539 | 2448901 | 1947.336 | 1.637 | |
(Average) | 2.539 | 2448925 | 1945.407 | 1.637 | |
120 PPM | 2.544 | 2879409 | 1859.908 | 1.645 | |
120 PPM | 2.549 | 2879775 | 1913.709 | 1.681 | |
120 PPM | 2.557 | 2879996 | 1935.066 | 1.662 | |
(Average) | 2.549 | 2879727 | 1902.894 | 1.663 | |
140 PPM | 2.561 | 3262297 | 1840.661 | 1.671 | |
140 PPM | 2.565 | 3263778 | 1896.287 | 1.696 | |
140 PPM | 2.579 | 3262569 | 1867.982 | 1.687 | |
(Average) | 2.568 | 3262881 | 1867.310 | 1.685 | |
160 PPM | 2.575 | 3678424 | 1898.627 | 1.685 | |
160 PPM | 2.574 | 3679069 | 1897.496 | 1.690 | |
160 PPM | 2.576 | 3679967 | 1824.393 | 1.684 | |
(Average) | 2.575 | 3679153 | 1873.505 | 1.687 |
Accuracy and Precision: The accuracy and precision of this HPLC method were determined by analyzing three different concentrations within the calibration range in triplicate (n=3) 12.
Accuracy: The accuracy of this method was determined by calculating recoveries of TAP from the standard solutions of known concentrations at 3 different levels, i.e., 70, 90, and 110 µg/mL. The average recovery of TAP was calculated from each concentration. The detailed results were given in Table 2.
Precision: The precision (% RSD) of the analytical procedure was evaluated by determining the intra- and inter-day coefficients of variation. Standards solutions were prepared, which were different from those used in the preparation of calibration curves.
These samples were evaluated 6 times by assaying, i.e., 3 times in a day and 3 more times on the next consecutive days. The results were given in Table 2.
TABLE 2: ACCURACY AND PRECISION STUDIES FOR TAP
Parameter | Inter-day (Drug Conc. µg/mL) | Intra-day (Drug Conc. µg/mL) | ||||
110 | 90 | 70 | 110 | 90 | 70 | |
Precision (% RSD) | 1.29 | 1.22 | 1.98 | 1.06 | 0.98 | 0.99 |
Accuracy (% recovery) | 99.81 ± 0.35 | 98.99 ± 0.82 | 100.23 ± 1.06 | 99.68 ± 1.71 | 99.79 ± 186 | 100.46 ± 1.00 |
Robustness: Robustness can be evaluated by varying the various developed TAP analysis parameters e.g., the composition of the mobile phase, mobile phase flow rate, detector wavelength, changing the retention time, relative retention time (RRT), resolution and number of plates, and temperature. The flow rate of the mobile phase was changed to ± 20%, and the temperature was changed to ± 5 °C. The results are disclosed in Table 3.
TABLE 3: ROBUSTNESS TESTING FOR THE METHOD
Parameter | Modification | TAP (% Recovery) |
Mobile phase composition ACN: 0.1 M potassium dihydrogen phosphate buffer (pH adjusted to 7 with Triethanolamine) | 40:60
50:50 60:40 |
99.6 ± 0.34
98.9 ± 0.27 99.7 ± 0.32 |
Flow Rate (mL/min.) | 1.08
1.20 1.32 |
96.6 ± 0.43
99.7 ± 0.29 99.7 ± 0.26 |
Injection Volume (mL) | 18
20 22 |
98.8 ± 0.31
99.7 ± 0.18 98.7 ± 0.19 |
Column Temp. (°C) | 23
25 27 |
98.9 ± 0.27
98.5 ± 0.29 99.3 ± 0.10 |
Detector Wavelength (nm) | 215
217 220 |
98.6 ± 0.43
99.1 ± 0.23 98.7 ± 0.45 |
Quantification Limits (LOD and LOQ): LOD (Lowest detectable concentration of the analyte) and LOQ (Lowest amount of the analyte) were used to assess the sensitivity of the method by the standard deviation of the response and slope. LOD and LOQ were quantitatively determined by the formula 3.3×(σ/s) and 10×(σ/s) respectively where σ is the standard deviation of intercept and s is the mean of the slope. Slope S was calculated from the calibration curve of the analyte, and the standard deviation was estimated by running six blank samples.
LOQ limits were expressed as the concentration of the analyte (parts per million). After LOQ determination, LOD was taken as the one-third of LOQ for their simultaneous analysis.
Determination of the TAP in Marketed Tablet: Three working solutions of TAP was prepared from the formulation (Tablet) and analyzed with the similar HPLC parameter settings and which revealed the drug content within the specified limits. The results were shown in Table 4.
TABLE 4: DETERMINATION OF DRUG TAP IN TABLET
Actual concentration (µg/mL) | Calculated concentration (µg/mL) | % Recovery |
80 | 79.85 ± 2.620 (1.27) | 99.9421 ± 0.22 (0.65) |
100 | 100.56 ± 2.398 (0.98) | 100.0601 ± 0.16 (0.17) |
120 | 119.69 ± 0.664 (0.19) | 99.9236 ± 0.12 (0.26) |
Values represent mean ± SD (n = 6)
Stability of Stock Solutions: A stock solution of TAP was evaluated for the stability and extent of degradation, if any, on storage. From the standard stock solution (1mg/mL), three working solutions with different concentrations (80, 100, and 120 µg/mL) were prepared and filtered through a 0.22µm filter and kept at room temperature.
HPLC run was fixed for each sample in triplicate at regular time intervals (i.e., after 6 h, 3 days and on 7th day) for a period of 7 days. During the period of 7 days, the stability was judged by comparing the peak areas of the drug at different time points. Table 5 represents stability data of the stock solution containing the drug.
TABLE 5: STABILITY OF THE DRUG’S STOCK SOLUTION
Drug Concentration
(µg/mL) |
% Recovery | ||
6 h | Day 3 | Day 7 | |
120 | 99.9867 ± 0.05 (0.57) | 99.379 ± 0.23 (0.32) | 99.7654 ± 0.21 (0.18) |
100 | 100.018 ± 0.02 (0.12) | 99.126 ± 0.42 (0.11) | 100.1298 ± 0.11 (0.14) |
80 | 99.7672 ± 0.14 (0.46) | 100.0238 ± 0.21 (0.18) | 99.8667 ± 0.30 (0.22) |
Values represent mean ± SD (%RSD, n = 6)
Short Term Stock Stability: A Stock solution of TAP was kept at room temperature for 6 h and chromatograms obtained by running three concentrations on the same day after 6 h.
Long Term Stock Stability: The chromatogram was developed on the 3rd and the 7th days from the preparation of the stock solution and compared with those obtained initially. Values were denoted as the peak area ± SD (%RSD) and compared with which obtained initially.
RESULTS:
Chromatographic Separation: The isocratic mode was employed for the elution of the drug, and TAP showed a retention time of 2.508 min. The run was further continued up to 5 min for the removal of traces of drugs and to re-equilibrate the system to the initial conditions. Fig. 2 illustrates the complete chromatogram produced over 5 min, which was showing a peak at the retention times. The linearly regressed standard curve of the TAP was obtained in the range 60-160 µg/mL, shown in Fig. 3.
FIG. 2: CHROMATOGRAMS OF TAP (RT = 2.508 min)
FIG. 3: LINEARLY REGRESSED CALIBRATION CURVE OF TAP IN DISTILLED WATER (λmax = 217 nm)
Validation of the Method: Validation parameters have been highlighted in Table 6 of TAP analysis. The method was validated with respect to parameters including linearity, precision and accuracy, the limit of quantitation (LOQ), and limit of detection (LOD), and stability of the solution. A standard curve was found linear. Linear regression analysis confirmed that the r2 value was found to be 0.9951, confirming the linear relationship between the concentration of the drug and the area under the curve. Accuracy and precision of recovery data ranged from 99 to 101%.
The results of robustness showed no significant statistical differences between various altered parameters with respect to those which were received initially, and finally, the method was found to be robust. Both intra- and inter-day precision (% RSD) of working standards were less than 2% over the selected range of the drug. The calculated LOD and LOQ concentrations proved the sensitivity of the method.
The stability of the drug solution was assured by comparing the percentage recovery of concentration of TAP from the sample after 7 days with the same day percentage recovery analysis, which had shown the negligible difference. Hence, the method validated and proved for quantitative analysis of TAP in pharmaceutical formulations.
TABLE 6: VALIDATION PARAMETERS OF THE HPLC METHOD FOR TAP
Parameters | Value |
Analytical wavelength (nm) | 217 |
Linearity (µg/mL) | 60-160 |
Slope | Y=22791X+120122 |
% RSD of slope (%) | 1.37 |
Intercept | 120122 |
Correlation coefficient (r2) | 0.9951 |
LOD (µg/mL) | 3.33 |
LOQ (µg/mL) | 10 |
Values represent mean ± SD (% RSD, n = 6)
DISCUSSION: Many mobile phase solvents/ combinations have been reported in various ratios, in literature for the separation of TAP. Here, TAP was separated using the mobile phase consisting of ACN: Potassium phosphate buffer containing triethanolamine for pH adjustments in the ratio 50:50. The rapidity of the method was proved as under the isocratic mode, the drug’s retention time was observed at 2.5 min. There was no interference of the impurities as well as additives of the tablet because method selectivity was proved by comparing the obtained chromatogram with the standard tapentadol drug chromatogram.
Linearity was confirmed over the TAP concentration range of 60-160 µg/mL with a correlation coefficient of 0.9951, as shown in Fig. 3. The accuracy and precision were studied in the drug concentration range of 70, 90, and 110 µg/mL, and TAP recovery was found in the range of 99.68-100.46%, as presented in Table 2.
For the robustness various chromatography parameters were varied viz. mobile phase ratio (40:60, 50:50, 60:40), flow rate (1.08, 1.20, 1.32 mL/min.), injection volume (18, 20, 22 µL), column temp (23 °C, 25 °C, and 27 °C) and the detector wavelength (215, 217, and 220 nm).
All the results were summarized in table III proved the method robust. LOD and LOQ were calculated at 3.33 and 10 µg/mL. Stability was calculated by taking percent ratio of the determined 80, 100, and 120 µg/mL concentrations after 6 h, 3 days, and 7days period which was found satisfactory as shown in Table 5 2, 13.
CONCLUSION: The HPLC method was developed and validated for the estimation of TAP in bulk, pharmaceutical formulations as well as for the detection of impurities and other related substances in stored formulations of TAP. The developed method was found simple, reproducible, rapid, and reliable for analysis of the drug. The validation report confirmed this method with good linearity, accuracy, precision, and adequate specificity.
ACKNOWLEDGEMENT: We thank Mr. Rajiv Mahajan, Production Manager Innova-Captab, Dehradun, Uttrakhand, India, for rendering gift samples of Tapentadol hydrochloride for analytical work.
CONFLICTS OF INTEREST: All authors report no conflict of interest directly or indirectly in the publication of this manuscript.
REFERENCES:
- Rinonapoli G, Coaccioli S and Panella L: Tapentadol in the treatment of osteoarthritis: pharmacological rationale and clinical evidence. J Pain Res 2019; 12: 1529-36.
- Bishnoi M, Jain A, Singla Y and Shrivastava B: Quick and precise RP-HPLC method development for Tapentadol hydrochloride. Asian Journal of Pharmacy and Pharmacology 2019; 5(5): 984-90.
- Zajączkowska R, Przewłocka B, Kocot-Kępska M, Mika J, Leppert W and Wordliczek J: Tapentadol-A representative of a new class of MOR-NRI analgesics. Pharmacol Rep 2018; 70(4): 812-20.
- Lange C and Rombout F: Titration of Tapentadol. Google Patents: 2009.
- Ishaq BM, Prakash KV and Mohan GK: Development and validation of RP-HPLC method for simultaneous estimation of tapentadol and paracetamol in bulk drug and its pharmaceutical dosage form. RJPT 2014, 7(2): 8.
- Basu A, Mahadevan J, Ithal D, Selvaraj S, Chand PK and Murthy P: Is tapentadol a potential Trojan horse in the postdextropropoxyphene era in India? Indian J Pharmacol 2018; 50(1): 44.
- Barbosa J, Faria J, Queirós O, Moreira R, Carvalho F and Dinis-Oliveira RJ: Comparative metabolism of tramadol and tapentadol: a toxicological perspective. Drug Metab Rev 2016; 48(4): 577-92.
- Swartz ME and Krull IS: Method Development Optimization and Validation Approaches. In Analytical method development and validation, CRC Press; 2018, 25-35.
- Jones GR and Handy RP: Quantitation of Tapentadol by Liquid Chromatography: Tandem Mass Spectrometry. In LC-MS in Drug Analysis, Springer 2019; 61-65.
- Jain A, Gulbake A, Jain A, Shilpi S, Hurkat P, Kashaw S and Jain SK: Development and validation of the HPLC method for simultaneous estimation of Paclitaxel and topotecan. J Chromatogr Sci 2014; 52(7): 697-03.
- Shrivastava A and Gupta VB: HPLC: Isocratic or Gradient Elution and Assessment of Linearity In Analytical Methods. J Adv Sci Res 2012: 3(2): 12-20.
- Pati NB, Mayasa V, Velivela S and VRM G: Development and validation of RP-HPLC method for estimation of Tapentadol hydrochloride. J Phar Res 2016; 10(6): 479-83.
- Verma A, Jain A, Tiwari A and Jain SK: Preformulation considerations of Natamycin and development of Natamycin loaded niosomal formulation. Asian Journal of Pharmacy and Pharmacology 2019; 5(5): 1022-30.
How to cite this article:
Bishnoi M, Singla YP, Jain A and Shrivastava B: RP-HPLC method quantitative estimation of tapentadol hydrochloride and its tablets. Int J Pharm Sci & Res 2020; 11(9): 4495-00. doi: 10.13040/IJPSR.0975-8232.11(9).4495-00.
All © 2013 are reserved by the International Journal of Pharmaceutical Sciences and Research. This Journal licensed under a Creative Commons Attribution-NonCommercial-ShareAlike 3.0 Unported License.
Article Information
46
4495-4500
644
775
English
IJPSR
M. Bishnoi, Y. P. Singla, A. Jain * and B. Shrivastava
DBT-Postdoctoral Research Associate, Department of Materials Engineering, Indian Institute of Science, Bangalore, Karnataka, India.
ankitjainsagar@gmail.com
26 September 2019
20 February 2020
11 March 2020
10.13040/IJPSR.0975-8232.11(9).4495-00
01 September 2020