SIMULTANEOUS ESTIMATION OF VALSARTAN AND HYDROCHLOROTHIAZIDE IN FIXED DOSE COMBINATION IN UV SPECTROPHOTOMETRY
HTML Full TextSIMULTANEOUS ESTIMATION OF VALSARTAN AND HYDROCHLOROTHIAZIDE IN FIXED DOSE COMBINATION IN UV SPECTROPHOTOMETRY
M. M. Deshpande*, M P. Mahajan and S. D. Sawant
Department of Pharmaceutical Chemistry, Smt. Kashibai Navale College of Pharmacy, Kondhwa, Pune, Maharashtra, India
ABSTRACT:Valsartan (VAL) and Hydrochlorothiazide (HTZ) are used in combination in treatment of Hypertension. Two simple, accurate, precise, economical and reproducible UV spectrophotometric methods have been developed for the estimation of Valsartan and Hydrochlorothiazide in Pharmaceutical formulation. Method I- Absorption ratio method (Q-analysis) using two wavelengths, 265nm (isobestic point at which both the drugs exhibit absorbance) 249nm (λmax of Valsartan) and Method II- Area under Curve method. For the second method Area under the Curve in the range of 249 -259nm and 261-281nm was selected for the analysis of Valsartan and Hydrochlorothiazide respectively. Linearity for detector response was observed in the concentration range of 2-24mg/ml & 2-14mg/ml for Valsartan and Hydrochlorothiazide respectively. The results of analysis have been validated statistically and by recovery studies the value of standard deviation was satisfactory and recovery studies ranging from 99.54 - 99.97 % for Valsartan and 99.75 - 101.04 % for Hydrochlorothiazide were indicative of the accuracy and precision of the proposed method The proposed methods were successfully applied for the determination of Valsartan and Hydrochlorothiazide in commercial pharmaceutical preparation. All two methods were validated statistically as per ICH guidelines.
Keywords:
Valsartan, Hydrochlorothiazide, Absorption ratio method, Area under Curve |
INTRODUCTION: Valsartan (VAL) is chemically (S)-N-(1-Oxopentyl)-N-[[2’-(1H-tetrazol-5-yl) [1, 1’-biphenyl]-4-yl]methyl]-L-valine, is an orally active specific angiotensin II receptor blocker effective in lowering blood pressure in hypertensive patients 1.
Hydrochlorothiazide (HTZ) is a diuretic of the class of benzothiadiazines widely used in antihypertensive pharmaceutical formulations, alone or in combination with other drugs, which decreases active sodium reabsorption and reduces peripheral vascular resistance 2. It is chemically 6-chloro-3, 4-dihydro-2H-1, 2, 4-benzothiadiazine-7-sulfonamide-1, 1-dioxide.
VAL and HTZ are official in USP-NF2007 6, 8 and IP 2007 3, BP 2004 5, respectively. Extensive literature survey revealed that few UV spectrophotometric methods 9-11, RP-HPLC methods 12-17 and HPTLCmethods 18-21 has been reported. Single drug estimation of VAL by RP-HPLC method has been reported 22-30.
Since, no spectrophotometric method is reported for simultaneous estimation of VAL and HTZ in combination by these two methods therefore, the present work describes successful attempt to estimate both these drugs simultaneously by two simple UV spectrophotometric methods (Q-Analysis method and Area under curve method). The proposed methods were optimized and validated as per ICH guidelines 7.
FIG. 1: STRUCTURE OF VAL
FIG. 2: STRUCTURE OF HTZ
MATERIAL AND METHODS: Instrumentation:For the present study JASCO double beam UV/Visible spectrophotometer (Model V-530) was used with slit width fixed at 2nm, equipped with spectra manager software (Version 1.5). Pair of 1-cm matched quartz cells were used to measure the absorbance of solution. The samples were weighed on electronic analytical balance (Contech Model CB-50) (Table 1).
TABLE 1: INSTRUMENTS
Name | Model | Manufacturer |
UV-Visible spectrophotometer | V-530 | Jasco |
Electronic analytical balance | CB-50 | Contech |
Materials: VAL and HTZ drug samples were kindly supplied as gift samples by Lupin Pharmaceuticals, J & K India. The Pharmaceutical dosage form used in the present study was Valent –H tablets (Lupin Pharmaceuticals Ltd).Assay was carried out on tablets containing label claim, each film coated tablet contains 12.5mg of HTZ and 80 mg of VAL.
Preparation of Stock Solutions: Accurately weighed 10 mg of VAL & 10 mg of HTZ was transferred to 100ml of volumetric flask and was dissolved separately in 0.1 M NaOH, sonicated for 3 mins, to give the standard stock solution of 100µg/ml respectively. Aliquots were prepared by using 0.1 M NaOH in the increasing concentration range.
Experimental methods:
Method I -
Absorption ratio/Q-analysis method: In this method absorbances are measured at two wavelengths, one being the λmax of Valsartan and other being a wavelength of absorptivity of the Hydrochlorothiazide. Then absorbance of both drugs was recorded on selected wavelengths. Concentrations of Valsartan & Hydrochlorothiazide were calculated by using following equations.
CVAL = QM -QY x A1 ------------ (1)
QX -QY aX1
CHTZ = QM –QX X A2 ------------ (2)
QY –QX aX1
Where, QM is ratio of absorbances A1 and A2 of mixture at λ1 and λ2 (isobestic point wavelength) QX is ratio of absorptivities ax1 and ax2 at λ1 and λ2. QY is ratio of absorptivities ay1 and ay2 at λ1 and λ2. CVAL and CHTZ are concentrations of Valsartan & Hydrochloro-thiazide respectively.
FIG. 3: OVERLAIN SPECTRA OF VAL & HTZ SHOWING ISOBESTIC POINT
METHOD II -
Simultaneous equation using area under the curve method: For the selection of analytical wavelength standard solutions of VAL & HTZ were prepared and series of dilutions of standard solutions of VAL and HTZ were prepared by using 0.1N NaOH and were scanned from 400 to 200 nm.
From the spectra of drug obtained after scanning of standard solution of VAL and HTZ, area under the curve in the range of 249-259nm and 261-281nm was selected for the analysis respectively. After optimization, the concentration of 12.8 µg/ml of VAL as standard 1 and 2 µg/ml of HTZ as standard 2 were found to give best results. The absorbance of each resulting solution was measured at 249-259 nm and 261-281 nm. The absorptivity values calculated. The calibration curve was plotted with concentration v/s area under the curve and regression equation was calculated.
Determination of Absorptivity values: (Table 2)
Absorbance
Absorptivity =
Concentration of that component in gm/l
FIG. 4: OVERLAIN SPECTRA OF VAL & HTZ SHOWING AREA UNDER THE CURVE
Concentration of VAL and HTZ was calculated using following formula;
A2ay1 - A1ay2
CVAL =
ax2ay1 - ax1ay2
A1ax2 - A2ax1
CHTZ =
ax2ay1 - ax1ay2
Where,
CVAL = Concentrations of VAL,
CHTZ = Concentrations of HTZ,
A1 = Area at 239-249 nm,
A2 = Area at 261-281 nm,
ax1 = Absorptivity value of HTZ at 239-249 nm,
ax2 = Absorptivity value of HTZ at 261-281 nm,
ay1 = Absorptivity value of VAL at 239-249 nm,
ay2 = Absorptivity value of VAL at 261-281 nm.
TABLE 2: ABSORPTIVITY VALUES OF VAL AND HTZ
Drug | Absorptivity at 249-259nm | Absorptivity at 261-281nm |
VAL | 0.043(ax1) | 0.036(ax2) |
HTZ | 0.080(ay1) | 0.0510(ay2) |
Analysis of Tablet Formulation: For the estimation of drugs in the commercial formulations, twenty tablets containing 80 mg of VAL and 12.5 mg of HTZ were weighed and average weight was calculated. The tablets were crushed and powdered in glass mortar. For the analysis of drugs, quantity of powder equivalent to 1.28 mg of VAL and 0.2 mg of HTZ was transferred to 100 ml volumetric flasks and dissolved in sufficient quantity of 0.1N NaOH. It was sonicated for 30mins and volume was made up to obtain a stock solution of 12.8μg/ml of VAL (maintaining 2 μg/ml of HTZ).
This solution was then filtered through Whatmann filter paper # 42. Further dilutions were made from this stock solution to get required concentration. The concentration of both VAL and HTZ was determined by measuring absorbances of sample solutions in wavelength range of 249nm-259nm (for VAL) and 261nm-281nm (for HTZ) using equation 3 and 4 Results of tablet analysis are shown in Table 3. The assay procedure was repeated six times (n=6).
TABLE 3: RESULT OF MARKETED FORMULATION ANALYSIS
Method | Drug | Amt of drug taken (μg/ml) | % of drug estimated* | %RSD (±SD) |
Q- analysis | VAL | 12.8 | 99.89±0.75 | 0.840 |
Area under curve | VAL | 12.8 | 99.44±0.55 | 0.155 |
Q- analysis | HTZ | 2 | 99.94±0.66 | 0.531 |
Area under curve | HTZ | 2 | 101.40±0.87 | 0.991 |
*Average of six determinations
Validation: The method was validated according to ICH guidelines to study linearity, accuracy, precision, LOD and LOQ.
Linearity: The measurement of linearity was evaluated by analyzing different concentrations of the standard solution of VAL and HTZ. For both the methods, the Beer law was obeyed in the concentration range 2-24 μg/ml and 2-14 μg/ml for VAL and HTZ respectively. The Absorbance was plotted against the corresponding concentrations to obtain the calibration graphs.
Accuracy (Recovery studies): To ascertain the accuracy of proposed methods, recovery studies were carried out by standard addition method at three different levels (80%, 100% and 120%). Percent recovery was calculated for VAL and HTZ, by both the methods (Table 4).
TABLE 4: RESULTS OF RECOVERY STUDIES
Recovery level | Drug | Conc. of drug Level taken in µg/m | Conc. of drug Level taken in µg/ml | Method I | Method II | ||||
% Recovery | SD | %RSD | % Recovery | SD | %RSD | ||||
80 | VAL | 9.6 | 7.68 | 99.54 | 0.054 | 0.624 | 99.23 | 0.066 | 0.720 |
100 | VAL | 9.6 | 9.6 | 99.84 | 0.069 | 0.753 | 100.02 | 0.390 | 0.570 |
120 | VAL | 9.6 | 11.52 | 99.97 | 0.089 | 0.982 | 100.08 | 0.240 | 0.333 |
80 | HTZ | 1.5 | 1.2 | 99.75 | 0.042 | 0.568 | 99.84 | 0.026 | 0.321 |
100 | HTZ | 1.5 | 1.5 | 99.97 | 0.075 | 0.510 | 99.92 | 0.042 | 0.630 |
120 | HTZ | 1.5 | 1.8 | 101.04 | 0.083 | 0.960 | 100.33 | 0.681 | 0.951 |
Precision: The reproducibility of the proposed methods was determined by performing tablet assay at different time intervals on same day (Intra-day precision) and on three different days (Inter-day precision) (table 5).
Limit of Detection and Limit of Quantitation: The LOD and LOQ were separately determined based on calibration curve. The residual standard deviation of a regression line or the standard deviation of y- intercepts of regression lines were used to calculate the LOD and LOQ (table 5).
The detection limit (LOD) may be expressed as:
LOD = 3.3 σ/S
The quantitation limit (LOQ) may be expressed as:
LOQ =10 σ/S
Where,
σ = the standard deviation of the response
S = the slope of the calibration curve
TABLE 5: RESULTS OF VALIDATION PARAMETERS
Parameter | Method I | Method II | ||
VAL | HTZ | VAL | HTZ | |
Linearity (μg/ml) | 2-24 | 2-14 | 2-24 | 2-14 |
Correlation coefficient (r2) | 0.9992 | 0.9998 | 0.9990 | 0.9995 |
Interday precision | 0.63 | 0.75 | 0.19 | 1.48 |
Intraday precision | 0.42 | 0.39 | 0.33 | 1.59 |
LOD (µg/ml) | 0.0024 | 0.033 | 0.0013 | 0.040 |
LOQ (µg/ml | 0.0063 | 0.039 | 0.0054 | 0.0164 |
RESULTS AND DISCUSSION: The present work provides an accurate, reproducible, sensitive method for the simultaneous analysis of VAL & HTZ in bulk and tablet formulation. Linear relationships between drug concentrations were obtained over the range of at 2-24 μg/ml & 2-14 μg/ml for VAL and HTZ respectively. Under experimental conditions described assay of tablet, linearity, accuracy studies and precision, LOD and LOQ were estimated. Correlation coefficient was found to be > 0.998.
The results of commercial tablet formulation are presented in Table 3. Results of accuracy studies are presented in Table 4. The % assay was found to be 99.44 - 99.89% for VAL and 99.94-101.40 for HTZ, S.D. and R.S.D. for six determinations of tablet sample, by both the methods, was found to be less than 2.0 indicating the precision of both the methods. No interference was observed from the pharmaceutical adjuvants.
CONCLUSION: The two spectrophotometric methods were developed and validated as per ICH guidelines. The standard deviation and % RSD calculated for the proposed methods are within limits, indicating high degree of precision of the methods. The results of the recovery studies performed indicate the methods to be accurate. Hence, it can be concluded that the developed spectrophotometric methods are accurate, precise and can be employed successfully for the estimation of Valsartan and Hydrochlorothiazide in bulk and formulation.
ACKNOWLEDGEMENT: The authors are thankful to Lupin Pharmaceuticals, J & K, India, for providing standard drug samples. The authors are grateful to the Principal and the Management of STES’s Smt. Kashibai Navale College of Pharmacy, Kondhwa (Bk) Pune for providing excellent infrastructure facility to carry out this research work.
REFERENCES:
- Markham A, Goa KL. Valsartan, A review of its pharmacology and therapeutic use in essential hypertension. Drugs.1997; 54:299–311.
- Carlucci G, Palumbo G, Mazzeo P, Quaglia MG. Simultaneous determination of Losartan and hydrochlorothiazide in tablets by high-performance liquid chromatography. J Pharm Biomed Anal.2000: 23:185–189.]
- Indian Pharmacopoeia Controller of Publication, Govt. of India, Ministry of Health and Family Welfare, New Delhi, 2007: Vol (2), 521, 1194-1195.
- Merck Index, Merck Research Laboratories: USA, 14th Edition, 2002:1705.
- British Pharmacopoeia, Her Majesty’s Stationary Office, London, Vol.1, 2004:979-980.
- United States Pharmacopoeia, USP 29/NF 24, The Official Compendia of standard Asian edition, USP Convention, Rockville, MD, 2005: 955-956.
- ICH, Q2 (R1), Harmonized Tripartite Guideline, Validation of analytical Procedure Methodology, IFPMA, Proceedings of the International Conference on Harmonization Geneva, November 2005.
- http://www.USP.org/PDF/EN/USPNF/VOL32,no1.pdf. last accessed in June 2011.
- Chaudhary A, Patel R, Chaudhary S, Gadhvi. Estimation of valsartan and Hydrochlorothiazide in pharmaceutical dosage forms by absorption ratio method. Journal of International Applied Biology and Pharmaceutical Technology2010;I (2): 455-464.
- Tatar S, Saglik S, Comparison of UV- and second derivative-spectrophotometric and LC methods for the determination of Valsartan in pharmaceutical formulation. Journal of Pharmaceutical and Biomedical Analysis.2002; 30: 371–375.
- Erk.N, Spectrophotometric analysis of Valsartan and Hydrochlorothiazide, Analytical letters.2002;35(2):283-302.
- Mohammad Y, Karnaker Reddy, Y. Ravindra Reddy, Md. Fasiuddin Arif, RP- HPLC Method Development and Validation for Simultaneous estimation of Amlodipine besylate, Valsartan and Hydrochlorothiazide in Tablet Dosage Form. Journal ofPharmacy Research.2010: 3(11): 2647- 2650.
- Chitlange S.S, Bagri K.S, and Sakarkar D.M, Stability Indicating RP-HPLC method for Simultaneous estimation of Valsartan & Amlodipine in Capsule Formulation.Asian J. Research Chem. 2008; 1(1): 15-18.
- Tian D.F, Tian X.L, Wang Z.Y and Mo F.K, Simultaneous determination of Valsartan & Hydrochlorothiazide in Tablets by RP-HPLC. Indian Journal of Pharmaceutical Sciences 2008; 70(3): 372-374.
- Prasad Ch, S. Kumari, and Ramulu S, A stability indicating RP-HPLC method with shorter runtime for simultaneous estimation of Valsartan and Amlodipine from their combination drug product. International Journal of Research and Chemistry 2010; 1(2): 102-110.
- Sheikh R.J, Puranik M.P, Bavadkar D.N, Mali P.R, and Yeole P.G, RP-HPLC method for simultaneous estimation of Valsartan and Hydrochlorothiazide in solid Dosage Form International Research Journal of Pharmacy.2011; 2(3):162-164.
- Jothieswari. D, Anand kumar. K, Vijayasanthi. D, Vijayakumar. B ,Priya.D,Validated, RP-HPLC Method for the Simultaneous determination of Amlodipine besylate, Valsartan and Hydrochlorothiazide in bulk and Pharmaceutical formulation.Journal of Pharmaceutical& Biomedical Sciences, 2010; 5(12): 1-7.
- Kumbhar S.T, Chougule G.K, Tegeli V.S, Gajeli G.B,A Validated HPTLC Method for Simultaneous Quantification of Nebivolol and Hydrochlorothiazide in Bulk and Tablet FormulationInternational Journal of Pharmaceutical Sciences & Drug Research, 2011; 3(1):62-66.
- Shah N.J, Suhagia B.N, Shah R.R, and Patel N.M, Development and validation of a simultaneous HPTLC method for the estimation of olmesartan medoxomil and hydrochlorothiazide in tablet dosage form. Indian Journal of Pharmaceutical Sciences,2007; 69(6):834-836.
- Kadam B.R, Bari S.B, Quantitative analysis of Valsartan and Hydrochlorothiazide in Tablet by High Performance Thin Layer Chromatography with ultraviolet absorption densitometry .Acta Chromatographica, 2007; 18:260-269.
- Shah N.J, Suhagia B.N, Shah R.R, and Patel N.M, HPTLC method for the simultaneous estimation of valsartan and hydrochlorothiazide in tablet dosage form.Indian Journal ofPharmaceutical Sciences, 2009;71(1):72-74.
- Ch. Krishnaiah. B, A. Raghupathi Reddy, Ramesh Kumar and K. Mukkanti, Stability- indicating UPLC method for determination of Valsartan and their degradation products in API & Pharmaceutical Formulation. Journal of Pharmaceutical & Biomedical Analysis,2010; 53(2):483-489.
- Pachauri S, Development & validation of HPLC Method for Analysis of Some Antihypertensive agents in their Pharmaceutical Dosage Forms. Journal of Pharmaceutical Sciences and Research, 2010;1-10.
- Rao K.S, Jena .N, Rao .M,Development and validation of a specific stability indicating high performance liquid chromatographic method for Valsartan.J Young Pharmacists, 2010; 2(2):183-189.
- Agrahari V, Kabra V, Gupta S, Nema R and Nagar MDetermination of Inherent Stability of Valsartan by Stress Degradation and Its Validation by HPLC. International Journal of Pharmaceutical and Clinical Research, 2009; 1(2):77-81.
- Bhatia S.M, Kokil U.S,Determination and validation of valsartan and its Degradation products by isocratic HPLC. Journal of chemical metrology,2009; 3(1):1-12.
- Thanusha G, Jose C, Babu G, Channa Basavaraj K, Reddy V.P, Sharadha C, A Validated RP- HPLC Method for the Quantitative Estimation of Valsartan in Bulk and Pharmaceutical Dosage Forms. International Journal of ChemTech Research, 2010; 2, (2):1194- 1198.
- Vinzuda D, Sailor G, Sheth N, RP-HPLC Method for Determination of Valsartan in Tablet Dosage Form. International Journal of Chemtech Research, 2010;2(3): 1461-1467.
- Patro S.K, Kanungo S.K, Patro V.J, and Choudhury N.S, Stability Indicating RP-HPLC Method for Determination of Valsartan in Pure and Pharmaceutical Formulation. E- Journal of Chemistry, 2010; 7(1): 246-252.
- Zarghi A, Shafaati A, Seyed M. F, Hooman M,Rapid Quantification of Valsartan in Human Plasma by Liquid Chromatography using a Monolithic Column and a Fluorescence Detection: Application for Pharmacokinetic Studies. scipharm2008: 439- 450.
Article Information
36
236-240
665KB
1872
English
IJPSR
M. M. Deshpande*, M P. Mahajan and S. D. Sawant
STES’s Smt Kashibai Navale College of Pharmacy, Kondhwa, Pune - 411048, Maharashtra, India
24 August, 2011
26 December, 2011
29 December, 2011
http://dx.doi.org/10.13040/IJPSR.0975-8232.3(1).236-40
1-January-2012