STABILITY-INDICATING RP-UHPLC METHOD FOR DETERMINATION OF TELMISARTAN IN DRUG SUBSTANCE AND MARKETED FORMULATION
HTML Full TextSTABILITY-INDICATING RP-UHPLC METHOD FOR DETERMINATION OF TELMISARTAN IN DRUG SUBSTANCE AND MARKETED FORMULATION
Biswa Ranjan Patra*, S. Mohan and Nagaraj Gowda
Department of Pharmaceutical Analysis, PES college of Pharmacy, Hanumantha Nagar, B.S.K first stage, Bengaluru, Karnataka, India
ABSTRACT: A simple, rapid and precise stability-indicating ultra high performance liquid chromatography (UHPLC) has been developed and validated for the estimation of telmisartan in drug substance and pharmaceutical dosage form. The chromatographic separation was achieved with a Poroshell 120EC-C18 column (4.6 x 50mm, 2.7µm) by using mobile phase acetonitrile: 50mM ammonium acetate buffer in the ratio of (45: 55 v/v), pH adjusted to 4.5 with acetic acid. The instrumental settings were flow rate of 0.5ml min-1, column temperature at 25 0C and detector wavelength of 290 nm using a photodiode array detector. Telmisartan was exposed to thermolytic, photolytic, acid, base, hydrolytic and oxidative stress conditions and the stressed samples were analyzed by the proposed method. The developed method shows excellent linearity over a range of 100-300 µg ml-1 for telmisartan. The recovery of telmisartan was above 96%. The proposed method was found to be suitable and accurate for quantitative determination and stability study of telmisartan pharmaceutical preparations
Key words: |
Telmisartan, RP-Ultra High Performance Liquid Chromatography
Degradation, Method validation
INTRODUCTION: Telmisartan, chemically described as 2-(4-{[4-methyl-6-(1-methyl-1H-1, 3-benzoimidiazol – 2 - yl) - 2-propyl-1H-1, 3-benzoimidazole-1-yl] methyl} phenyl) benzoic acid.Molecular formula C33 H30 N4 O2 Molecular weight 514.62 1. Telmisartan is an angiotensin II receptor antagonist which helps to lower blood pressure by blocking rennin-angiotensin-aldosterone system. Telmisartan is orally active and due to their specificity of action provide good conditions for patient compliance as well as high effectiveness 2. It is available as tablets for oral administration containing 40 mg of telmisartan
Literature survey revealed that several RP-HPLC methods 3-18, HPTLC method 19-23, UV spectroscopy method 24-25 and several bioanalytical methods 26-30, were reported for estimation of telmisartan. The objective of this work is to develop a simple accurate stability indicating isocratic reverse phase UHPLC method for estimation of telmisartan in drug substance and tablet dosage form.
MATERIAL AND METHODS:
Telmisartan standard drug was obtained as a gift sample from Micro lab, Bengaluru, India, ammonium acetate from Merck, India, acetic acid from Spectrochem, sodium dihydrogen orthophosphate dehydrate from Fisher Scientific, hydrochloric acid, sodium hydroxide, hydrogen peroxide, (AR grade) from Sd fine-chem. limited, acetonitrile (HPLC grade)was obtained from Spectrochem and methanol (HPLC grade) was obtained from Finar, India. Mill Q HPLC water was used for all purposes. Telmisartan tablet label claim 40 mg from Telvas tablet (40mg) Aristo Pharmaceuticals Pvt. Ltd. India was purchased from the market.
Instrumentation and chromatographic conditions:
UHPLC system(Agilent 1260 UHPLC System) consists of 1260 quaternary pump, standard auto sampler, poroshell 120EC-C18 column (4.6 x 50mm, 2.7µm), Diode array detector with chemstation Software, Shimadzu AUX220 Weighing Balance, Elico India LI 127 pH meter and Grant Sub-aqua 12 Water bath were used in the analysis. Mobile phase was a mixture of acetonitrile and buffer in the ratio of (45: 55 v/v). The buffer used in the mobile phase contains 50mM ammonium acetate in Mill Q water and pH was adjusted to 6 with acetic acid filtered through a 0.45µm nylon filter and degassed in an ultrasonic bath prior to use. The flow rate was kept at 0.5ml/minute and Column temperature was maintained at 250 C, the injection volume was 10µl and the detection was monitored at a wavelength of 290nm.
Preparation of Standard stock solutions:
Standard Solution was prepared by dissolving the drug in methanol and diluting it to the desired concentration.
Telmisartan:
A 25mg of standard telmisartan (99.26%) was accurately weighed, transferred in to 25ml volumetric flask and dissolved with methanol (1000 µg/ml).
Standard solution:
A standard solution was prepared by pipetting out 2 ml of telmisartan from stock solution (1000 µg/ml), transferred to 10ml volumetric flask and the volume was made up to 10ml by using methanol. This solution contained 200µg/ml of telmisartan.
Calibration curve solution:
The calibration curve solutions of concentration contain 100 to 300 µg ml -1 for telmisartan.
Preparation of Sample solution:
20 tablets of telmisartan (40 mg) were weighed accurately. A quantity of the powder equivalent to one tablet containing 40 mg of telmisartan was transferred in to a 100 volumetric flask. To this flask, 50 ml of methanol was added and solution was sonicated for 30 minutes. The solution was cooled to ambient temperature. Then the volume was made up with methanol, filtered through whatman filter paper. Pipetting out 5 ml from the above prepared solution (400 µg/ml), transferred to 10ml volumetric flask and the volume was made up to 10ml by using methanol. This solution contained 200 µg/ml of telmisartan. Prior injecting the solution in chromatographic system, it was filtered through 0.45 µm HPLC syringe filter. Sample analysis was performed for three replicates at 290nm.
Procedure for forced degradation studies of Drug substance:
Forced degradation of drug substance and drug product was carried out under thermolytic, photolytic, acid, base hydrolytic water hydrolysis and oxidative stress conditions.
Acid degradation:
Pipetting out 2 ml of telmisartan standard solution from the stock solution (1000 µg ml-1) and transferring the solution in to a 10ml volumetric flask add 2ml of 1N HCl and diluted up to 10ml with methanol. The solution was kept at 60 0C for 4 hours. The solution was allowed to attend ambient temperature. Repeated the same with 0.1N HCl.
Alkali degradation:
Pipetting out 2 ml of telmisartan standard solution from the stock solution (1000 µg ml-1) and transferring the solution in to a 10ml volumetric flask add 2ml of 1N NaOH and diluted up to 10ml with methanol. The solution was kept at 60 0C for 4 hours. The solution was allowed to attend ambient temperature. Repeated the same with 0.1N NaOH.
Degradation under neutral hydrolytic condition:
Pipetting out 2 ml of telmisartan standard solution from the stock solution (1000 µg ml-1) and transferring the solution in to a 10ml volumetric flask, add 2ml of distilled water and diluted up to 10ml with methanol. The solution was kept at 60 0C for 8 hours.
Degradation under oxidative condition:
Pipetting out 2 ml of telmisartan standard solution from the stock solution (1000 µg ml-1) and transferring the solution in to a 10ml volumetric flask, add 2ml of 30% v/v H2O2 and diluted up to 10ml with methanol. The solution was kept at 60 0C for 1 hour.
Degradation under dry Heat:
Dry heat study was performed by keeping drug sample in an oven at 80 0C for 24 hours. Samples were withdrawn, allowed to attend ambient temperature and dissolved in methanol to prepare sample solution to get concentrations of 1000 mcg/ml. Pipetting out 2 ml of telmisartan sample solution and transferring the solution in to a 10ml volumetric flask and diluted up to 10ml with methanol.
Sunlight degradation studies:
Sunlight study was performed by exposing the drug sample directly to sunlight for 8 hours. Samples were withdrawn, dissolved in methanol to prepare sample solution to get concentrations of 1000 mcg/ml. Pipetting out 2 ml of telmisartan sample solution and transferring the solution in to a 10ml volumetric flask and diluted up to 10ml with methanol.
Photo degradation studies:
Photolytic studies were carried out by exposing the drugs to UV short 254nm and UV long light 366nm for 24 hours. Samples were withdrawn, dissolved in methanol to prepare sample solutions to get concentrations of 1000 mcg/ml. Pipetting out 2 ml of telmisartan sample solution and transferring these solutions in to a 10ml volumetric flask and diluted up to 10ml with methanol.
Procedure for forced degradation studies of Drug products:
A forced degradation study of drug products in acidic, basic, water hydrolysis, oxidative conditions was carried out using filtered solution (as described in sample preparation) to achieve 200µg/ml of telmisartan. For thermolytic and photolytic degradation, a quantity of powder equivalent to on tablet containing 40 mg of telmisartan was exposed. Then the solutions were prepared as described in preparation of sample.
RESULTS:
Optimization of the Chromatographic Conditions:
UHPLC procedure was optimized with a view to develop a stability-indicating method. Depending on the nature of the sample (ionic or neutral molecule), its molecular weight and solubility; chromatographic method is to be selected. Our drug is polar in nature and so reverse phase chromatographic technique was selected for the present study. To develop stability indicating method different stationary phases like C18 C8, different mobile phases containing buffers like ammonium acetate, phosphate etc were used. The main objective of the chromatographic method development was to achieve a peak tailing factor less than 2, retention time in between 1 to 5 minutes due to short column length. The chromatographic separation was achieved using poroshell 120EC-C18 column (4.6 x 50mm, 2.7µm) by changing the composition of mobile phase, wavelength, chromatographic method was optimized.
During development studies, mobile phase containing 50mM ammonium acetate in water, pH adjusted to 4.5 with acetic acid and acetonitrile in the ratio of 55:45 (v/v) with flow rate of 0.5 ml min-1 and a column temperature of 25 0C was selected. In these selected optimized conditions, the drug telmisartan (in Fig. 1 and 2) had adequate retention, peak shape with less tailing and the chromatographic analysis was less than 5 minutes, which also separates the degradants from telmisartan.
Degradation observed:
Blessy and Ruchi 31, in their article on stress testing suggested a target degradation of 5-20% has been accepted as reasonable for validation of chromatography assay. Similarly Singh and Bakshi 32, in their article on stress testing suggested a target degradation of 20-80% for establishing stability-indicating studies, and also intermediate degradation products should not interfere with any stage of drug analysis. Though conditions used for forced degradation were attenuated to achieve degradation in the range of 5-80% but this could not achieve in case of water hydrolysis, thermal and photolytic degradation even exposure for prolonged duration. Extensive degradation of telmisartan was observed on acidic, alkali hydrolysis and oxidative condition (in Fig. 3a to 3c). Table 1 showed the extent of degradation of telmisartan under various stress condition.
CHROMATOGRAM OF TELMISARTAN STANDARD (RT: 3.909 MIN)
CHROMATOGRAM OF TELMISARTAN IN TELVAS TABLET (RT: 3.909 MIN)
RESULTS OF ANALYSIS OF FORCED DEGRADATION STUDY SAMPLES
Conditions(Stress induced) | Peak Purity
(P) |
% Telmisartan
degradation |
Acidic | 999.857 | 1.6 (0.1N HCl) |
12.6 (1N HCl) | ||
Alkali | 999.456 | 11.8 (0.1N NaOH) |
71.9 (1N NaOH)) | ||
Water | 999.984 | No degradation |
Oxidative | 999.676 | 26.4 |
Thermal | 999.915 | 0.7 |
UV light (254 nm for 24 hours) | 999.903 | 0.5 |
UV light (326nm for 24 hours) | 999.915 | 0.7 |
Sun light for 8 hours | 999.967 | No degradation |
1N HCL Time in( minutes)
1N NaOH Time in( minutes)
Oxidation Time in( minutes)
Chromatograms of forced degradation of acid & alkali hydrolysis, oxidative degradation for telmisartan.
Method Validation:
The method was validated according to the ICH guideline 33. The following validation parameters were considered for the newly developed method such as system suitability, linearity, precision/reproducibility, accuracy, specificity, robustness, stability of analytical solution.
System suitability:
The system suitability of the method was tested by Injected blank (1 injection), standard preparation 200 µg ml -1 (5 injection) and then system suitability parameters like theoretical plates, tailing factor, % RSD were studied and found that all the system suitability parameters are within acceptance criteria.
Calibration & Linearity:
Linearity of the method was tested by preparing five standard solutions from 50% to 150% of telmisartan and injected in triplicate. The standard solutions contain the concentration range from 100 to 300 µg ml -1 of telmisartan. The calibration graph was obtained by plotting peak area against the concentration of the drug (Fig.4). The equation of the calibration curve for telmisartan obtained was 54.776x+270.12. The calibration graph was found to be linear with regression coefficient and correlation coefficients were found to be 0.9992 and 0.9996.
LINEARITY RANGE OF TELMISARTAN
Precision (Repeatability):
The precision of the analytical method is determined by using telmisartan standard preparation to ensure that the analytical system is working properly. The retention time and area of six determinations of standard solution of concentration 200 µg ml -1 was measured and % RSD was calculated. Similarly intermediate precision of the method was determined by analyzing the telmisartan standard solution of concentration 200 µg ml -1 for six times on different days, by different analysts. The precision of analytical method was usually expressed as standard deviation or relative standard deviation of series of measurements. The results of the precision study indicate that the method is precise. (RSD % <2)
Accuracy (Recovery Test):
Accuracy of the method was studied by recovery experiments. The accuracy of an analytical method was performed in three different levels, with each level in triplicate for telmisartan standard drug (nine determinations). Spiked known quantity of telmisartan standard drugs at 50%, 100% and 150% levels in to the telmisartan tablet sample solutions containing 40 µg ml-1 of telmisartan. The prepared solutions were then analyzed and percentage recoveries were calculated. The recovery value of telmisartan ranges from 99.4 to 100.4 %. The average recovery of three levels (9 determinations) for telmisartan was 99.7%.
Specificity:
Photodiode array detection was used as an evidence of the specificity of the method and to evaluate the homogeneity of the drug peak. The peak purity values are more than 997 for drug substance as well as drug product at 290 nm which shows that the peaks of analyte were pure and also the formulation excipients and degradants were not interfering with the analyte peaks.
Robustness:
The robustness of an analytical method was measure of its capacity to remain unaffected by small but deliberate variations in method parameters. To determine robustness of the method, experimental conditions were purposely altered chromatogram of telmisartan was studied. The flow rate of the mobile phase was 0.5ml/minute; it was changes to ±0.2 units from 0.5 to 0.3 ml/minute and 0.7 ml/minute. The column temperature was 25 0 C, changes to ±50 C from 25 0 C to 20 0C and 30 0C while the other mobile phase components were held constant. Similarly change in mobile phase composition was studied at buffer: acetonitrile (57:43v/v) and 53:47 (v/v). At all conditions area % RSD was within acceptance criteria. Hence it is concluded that the method is robust.
Stability of analytical solution:
Stability of standard and sample solutions of telmisartan was evaluated injecting sample and standard at regular intervals up to 48 hours and calculated the area % difference. The area % difference value indicate that the solutions were stable for 48 hours at ambient temperature as there is no formation of unknown peaks and solution remained stable.
TABLE 2: RESULTS OF METHOD VALIDATION FOR TELMISARTAN
Validation parameters | Observed Value | Acceptance criteria | |
System Suitability
|
Tailing factor | 0.86% | RSD NMT 2.0% |
Theoretical Plate count | 8305 | RSD NLT 2000 | |
The % RSD for the areas of 5 replicate injections of telmisartan peak | 0.1% | RSD NMT 2.0% | |
Linearity
|
Regression coefficient | 0.9992 | RSD NLT 0.998 |
Correlation coefficient | 0.9996 | RSD NLT 0.998 | |
slope | 54.776 | ---- | |
Intercept | 270.12 | ----- | |
Precision | System precision | 0.04% | RSD NMT 2.0% |
Method precision | 0.05% | RSD NMT 2.0% | |
Intermediate precision | 0.06% | RSD NMT 2.0% | |
Accuracy | 50%, 100% and 150% Level | 99.7 % | Mean Recovery 95% and 105%. |
Specificity | Peak purity | 999.948 | More than 997 |
Robustness | Change in column temperature ±50 C | 0.06%(200C),
0.1%(300C) |
RSD NMT 2.0% |
Change in flow rate ±0.2mL/min | 0.93 (-0.2)
0.3(+0.2) |
RSD NMT 2.0% | |
Change in organic phase ratio ±2% | 0.34 (-2 %)
0.5 (+2 %) |
RSD NMT 2.0% | |
Solution stability | Area % difference for telmisartan standard solution | 0.1% | NMT 2.0% |
Area % difference for telmisartan sample solution | 0.2% |
DISCUSSION: Telmisartan drug substance and drug product when it is treated with acid, alkali, neutral, hydrogen peroxide, dry heat, sunlight and UV light, showed well separated peak of pure telmisartan as well as some additional degradants. The peaks of the degraded products were well resolved from the telmisartan drug peak. The chromatogram of the 1 N acid degraded sample for telmisartan showed 01 additional peak of retention time 0.887 minute. The chromatogram of the 1 N NaOH degraded sample for telmisartan showed 01 additional peak of retention time 1.233 minutes. The chromatogram of the oxidation degraded sample for telmisartan showed 2 additional peaks of retention time 1.022 and 4.227 respectively. No additional peaks were developed in water, dry heat & UV light and sunlight degradation studies. Major degradation of telmisartan was observed under acidic, alkali, oxidation. No degradation was observed under hydrolytic, thermal and photolytic conditions.
The response for the drug was found to be linear in the concentration range 100-300 µg ml -1 for telmisartan with respect to the peak area. The % RSD value for repeatability of standard application for precision study was found to be less than 2% for telmisartan, this conform that the method is precision. The accuracy of the method was determined and the mean recovery of telmisartan was 99.7%. The specificity of the method was ascertained by peak purity profiling studies and the developed method was specific. The low values of % R.S.D obtained after introducing small deliberate changes in the developed UHPLC method indicated the robustness of the method.
CONCLUSION: The developed isocratic RP-UHPLC method for the analysis of telmisartan is precise, accurate and with a short run time for the estimation of telmisartan in bulk and pharmaceutical formulation. The method was fully validated, that shows satisfactory data for the validation parameters according to ICH guideline. As the developed method separates the drug from its degradation products, it can be employed as a stability indicating one. Hence this method can be used in quality control laboratories, research institutions for the analysis of telmisartan.
CONFLICT OF INTEREST STATEMENT: We declare that we have no conflict of interest.
ACKNOWLEDGEMENTS: The authors are grateful to the Micro lab, Bengaluru, India for gift sample telmisartan and to the Department of Pharmaceutical Analysis, PES College of Pharmacy, Bengaluru for providing laboratory facilities for their research work.
REFERENCES:
- Merck & Co: The Merck index USA, 14th edition 2006.
- Bergovac M, Plavec D, knezevic A, Trkulja V: Telmisartan in daily clinical practice: Factors affecting efficacy in treatment of primary arterial hypertension. Journal of Postgradaduate Medicine 2009; 55:27-32.
- Bhavna patel A, Anand kumari D, Captain: RP-HPLC method for simultaneous estimation of telmisartan and hydrochlorothiazide in API and dosage form. Indo American Journal of Pharmaceutical Research 2014; 4:3031-3038.
- E, Syama Sundhar B: New analytical method development and validation for the simultaneous estimation of telmisartan and hydrochlorothiazide in bulk and tablet dosage form using RP-HPLC. Caribbean Journal of Science and Technology 2014; 2:519-529.
- Bhavani V, Siva Rao T, Raju SVN, Madhusudan B and Jamelunnisa B: Stabililty indicating UPLC method for the estimation of telmisartan related substancs in tablets formulation. International Journal of Scientific & Research Pubublications. 2013; 3(2):01-08.
- Saroj Bala, Mahatma OP, Md Sabir Azim: RP-HPLC method development and validation of telmisartan and hydrochlorothiazide in combination dosage form (Tablets). American Journal of Pharmacy and Health Research 2013; 1:60-77.
- Dhanalakshmi K, Tatineni Vasuki, Nagarjuna Reddy, Jotheiswari D: Analytical method development and validation of telmisartan and hydrochlorothiazide in dissolution by RP-HPLC. International Journal of Biological and Pharmaceutical Research 2013; 4:200-211.
- Shubhangi Shinde, Niranjan Chivate, Pramodinee Kulkarni, Nilofer Naikwade: Analytical method development and validation of telmisartan and hydrochlorothiazide in tablets using orthogonalpolynomial function method. International Journal of Pharmacy and Pharmaceutical Science 2013; 5:73-75.
- Rama mohana reddy M, Praveen kumar A, Krishna reddy V, Wasimul haque S: Stability-indicating HPLC method for simultaneous estimation of low level impurities of telmisartan and hydrochlorothiazide in tablet dosage forms. International Journal of Pharmacy and Pharmaceutical Science 2012; 4:497-504.
- Psrchnp varma D, A. Lakshmana rao A, and Dinda SC: Stability indicating RP-HPLC method for simultaneous determination of telmisartan and hydrochlorothiazide in pharmaceutical dosage form. International Journal of Pharmaceutical, chemical and biological sciences 2012; 2:382-391.
- Jayesh panchal G, Ravindra patel V, Bhoomika mistry R, Shobhana Menon K: Development and validation of reversed-Phase LC method for simultaneous determination of telmisartan, amlodipine and their degradation products in fixed dose combination tablets. Eurasian Journal of Analytical Chemistry 2012; 7:28-42.
- Kapendra S, Shikha S, Mohammad S and Anees A Siddiqui: Comparatives study of forced degradation behavior of Telmisartan by UPLC and HPLC and development of validated stability indicating assay method according to ICH Guidelines. Journal of Chromatography Separation techniques 2012; 3:01-06.
- Narendra kumar reddy K, Devala rao G & Hema pratyusha P: Isocratic RP-HPLC method validation of telmisartan in pharmaceutical formulation with stress test stability evaluation of drug substance. Journal of Chemical & Pharmaceutical Sciences 2012; 5:16-21.
- Manoj S. Charde Gupta A, Chakole R D: Stability-indicating RP-HPLC method for analysis of telmisartan in the dosage form. International Journal of Advances in Pharmaceutical Analysis 2012; 2:01-05.
- Sujana k, Gowri sankar D, Bala souri O & Swathi rani G: Stability indicating RP-HPLC method for the determination of telmisartan in pure and pharmaceutical formulation. International Journal of Pharmacy & Pharmaceutical Sciences 2011; 3:164-167.
- Vijay Kumar G, Murthy TEGK and Sambasiva Rao KRS: Validated RP-HPLC method for the estimation of telmisartan in serum samples. International Journal of Research in Pharmacy & Chemistry 2011; 1:703-706.
- Londhe S V, Kaul N, Agrawal H and Mahadik K R: Stability-indicating RP-HPLC method for analysis of telmisartan in the bulk drug and in formulations. Acta Chromatographica 2010; 22:539-548.
- Kiran R, Vipul P, Jaiprakash N, Devand B: Stability indicating liquid chromatography method for the simultaneous determination of telmisartan and ramipril in dosage form. Journal of Chromatography 2008; 67:575-582.
- Radhasri T, Subba reddy M, Nagaraj G: Validated specific densitometric method for simultaneous estimation of telmisartan and atorvastatin in presence of degradation products formed under ICH-recommended stress conditions. Journal of Planner Chrom.2013; 26:445-451.
- Smita Londhe V: Stability-indicating HPTLC method for telmisartan in the presence of degradation products, its process related impurity and identification of acid degradation product. Pharm Analysis & Quality Assurance 2011; 2011:194-211.
- Patil U P, Gandhi S V, Senga M R, Rajmane V S: A validated densitometry method for analysis of telmisartan and atorvastatin calcium in fixed dose combination. Journal of Chilean Chemical Society 2010; 55:94-96.
- Patel VA, Patel P G, Chaudhary B G, Rajgor N B, Rathi S G: Development and validation of HPTLC method for the simultaneous estimation of ramipril and telmisartan in combined dosage form. International Journal of Pharmaceutical & biological research 2010; 01:18-
- Chitra Prabhu, Ganesh S, Subramaniam, Arumugam K, Suvarnakini, Mallayasamy S R: Determination of telmisartan by HPTLC- A stability indicating assay. Journal of Planner Chromatography.2007; 20:477-481.
- Komal Patel, Komal Dhudasia, Amit Patel, Jayant Dave, and Chaganbhai Patel: Stress degradation studies on telmisartan and development of a validated method by UV spectrophotometry in bulk and pharmaceutical dosage forms. Pharmaceutical Methods 2011; 2:253-259.
- Bankey S, Tapadiya G G, Sahoo S S, Bindaiya S, Deepti Jain, Khadbadi S S: Simultaneous determination of ramipril, hydrochlorothiazide and telmisartan by spectrophotometry. International Journal of Chem Tech Research 2009; 1: 183-188.
- Maria Del Rosario Brunetto, Yaritza Contreras, Sabrina Clavijo, Dina Torres, Yelitza Delgado,Carlos Ayala et al: Determination of losartan, telmisartan and valsartan by direct injection of human urine in to a column-switching liquid chromatographic system with fluorescence. Journal of Pharmaceutical and Biomedical Analysis 2009; 50:194-199.
- Sanjay Bais,Chakravarti Singh, Indrajeet Singhvi, Yash Joshi, Bhaumik Patel, Anil Chandewar: Novel method for quantitative estimation of telmisartan from tablet formulation by colorimetric method Asian Journal of Pharmaceutical Analysis 2014; 04:54-56.
- He Zhang, He Zhang, Yunyun Jiang, Jun Wen, Tingting Zhou, Guorong Fan et al: Rapid determination of telmisartan in human plasma by HPLC using a monolithic column with fluorescence detection and its application to a bioequivalence study. Journal of Chromatography B 2009; 877:3729-3733.
- Pengfei Li,Yingwu Wang, Yan Wang, Yunbiao Tang, Yimin Cui, Jingkai Gu et al: Determination of telmisartan in human plasma by liquid chromatography-tandem mass spectrometry. Journal of Chromatography B 2005; 828126-129.
- Cagigal E, Gonzalez L, Alonso R M, Jimenez RM: pKa determination of angiotensin II receptor antagonists (ARA II) by spectrofluorimetry. Journal of Pharmaceutical and Biomedical Analysis 2001; 26:477-86.
- Blessy M, Ruchi Patel D, Prajesh Prajapati N, Agrawal Y K: Development of forced degradation and stability indicating studies of drugs-A review. Journal of Pharmaceutical Analysis 2014; 4:159-165.
- Monika Bakshi, Saranjit Singh: Development of validated stability-indicating assay methods-critical review. Journal of Pharmaceutical and Biomedical Analysis 2002; 28(6):1011-40.
- Validation of analytical procedure: Text & methodology, In ICH Harmonized Tripartite Guidelines, Q2 (R1) November 2005.
How to cite this article:
Patra BR, Mohan S and Gowda N: Stability-Indicating RP-UHPLC Method for Determination of Telmisartan in Drug Substance and Marketed Formulation. Int J Pharm Sci Res 2016; 7(5): 2031-39.doi: 10.13040/IJPSR.0975-8232.7(5).2031-39.
All © 2013 are reserved by International Journal of Pharmaceutical Sciences and Research. This Journal licensed under a Creative Commons Attribution-NonCommercial-ShareAlike 3.0 Unported License.
Article Information
25
2031-39
416
1456
English
IJPSR
Biswa Ranjan Patra*, S. Mohan and Nagaraj Gowda
Department of Pharmaceutical Analysis, PES college of Pharmacy, Bengaluru, Karnataka, India.
biswaipt@gmail.com
30 November, 2015
09 February, 2016
19 March, 2016
10.13040/IJPSR.0975-8232.7(5).2031-39
01 May 2016